TY - JOUR A1 - Kittner, Maria A1 - Eisentraut, Paul A1 - Dittmann, Daniel A1 - Braun, Ulrike T1 - Decomposability versus detectability: First validation of TED‐GC/MS for microplastic detection in different environmental matrices N2 - A fast method for microplastic detection is thermal extraction desorption‐gas chromatography/mass spectrometry (TED‐GC/MS), which uses polymer‐specific thermal decomposition products as marker compounds to determine polymer mass contents in environmental samples. So far, matrix impacts of different environmental matrices on TED‐GC/MS performance had not yet been assessed systematically. Therefore, three solid freshwater matrices representing different aquatic bodies with varying organic matter contents were spiked with a total of eight polymers. Additionally, for the first time, the two biodegradable polymers polybutylene adipate terephthalate (PBAT) and polylactide (PLA) were analysed using TED‐GC/MS. The methodological focus of this work was on detectability, quality of signal formation as well as realisation of quantification procedures and determination of the limit of detection (LOD) values. Overall, TED‐GC/MS allowed the unambiguous detection of the environmentally most relevant polymers analysed, even at low mass contents: 0.02 wt% for polystyrene (PS), 0.04 wt% for the tyre component styrene butadiene rubber (SBR) and 0.2 wt% for polypropylene (PP), polyethylene (PE) and PBAT. Further, all obtained LOD values were increased in all matrices compared to the neat polymer without matrix. The LOD of the standard polymers were increased similarly (PS: 0.21–0.34 μg, SBR: 0.27–0.38 μg, PP: 0.32–0.36 μg, PMMA: 0.64–1.30 μg, PET: 0.90–1.37 μg, PE: 3.80–6.99 μg) and their decompositions by radical scission processes were not significantly influenced by the matrices. In contrast, matrixspecific LOD increases of both biodegradable polymers PBAT (LOD: 1.41–7.18 μg) and PLA (0.84–20.46 μg) were observed, probably due to their hetero‐functional character and interactions with the matrices. In conclusion, the TED‐GC/MS performance is not solely determined by the type of the polymers but also by the composition of the matrix. KW - TED-GC/MS KW - Microplastic KW - Freshwater matrices KW - Gas chromatography KW - Mass spectrometry PY - 2023 DO - https://doi.org/10.1002/appl.202200089 SN - 2702-4288 VL - 2 IS - 3 SP - 1 EP - 12 PB - Wiley-VCH CY - Weinheim AN - OPUS4-57419 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Wiesner, Yosri A1 - Bednarz, Marius A1 - Braun, Ulrike A1 - Bannick, Claus Gerhard A1 - Ricking, Mathias A1 - Altmann, Korinna T1 - A promising approach to monitor microplastic masses in composts N2 - Inputs of plastic impurities into the environment via the application of fertilizers are regulated in Germany and the EU by means of ordinances. Robust and fast analytical methods are the basis of legal regulations. Currently, only macro- and large microplastic contents (>1 mm) are measured. Microplastics (1–1,000 µm), are not yet monitored. Thermal analytical methods are suitable for this purpose, which can determine the mass content and can also be operated fully automatically in routine mode. Thermal extraction desorption-gas chromatography/mass spectrometry (TED-GC/MS) allows the identification of polymers and the determination of mass contents in solid samples from natural environments. In accordance with the German or European Commission (EC) Fertiliser Ordinance, composting plants should be monitored for microplastic particles with this method in the future. In this context a compost plant was sampled. At the end of the rotting process, the compost was sieved and separated in a coarse (>1 mm) and a fine fraction (<1 mm). The fine fraction was processed using density separation comparing NaCl and NaI as possible salt alternative and screened for microplastic masses by TED-GC/MS with additional validation and quality assurance experiments. With TED-GC/MS total microplastics mass contents of 1.1–3.0 μg/mg in finished compost could be detected with polyethylene mainly. What differs much to the total mass of plastics in the coarse fraction with up to 60 μg/mg, which were visually searched, identified via ATR-FTIR and gravimetrically weighted. KW - Microplastics KW - TED-GC/MS KW - Compost KW - Monitoring KW - Soil PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-586688 DO - https://doi.org/10.3389/fenvc.2023.1281558 SN - 2673-4486 VL - 4 SP - 1 EP - 12 PB - Frontiers Media CY - Lausanne AN - OPUS4-58668 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Klöckner, P. A1 - Seiwert, B. A1 - Eisentraut, Paul A1 - Braun, Ulrike A1 - Reemtsma, T. A1 - Wagner, S. T1 - Characterization of tire and road wear particles from road runoffindicates highly dynamic particle properties N2 - Tire and road wear particles (TRWPs) are heteroagglomerates of tire rubber and other particles deposited on the road surface and one of the main contributors to non-exhaust emissions of automobile traffic. In this study, samples from road environments were analyzed for their TRWP contents and concentra- tions of eight organic tire constituents. TRWP concentrations were determined by quantifying Zn in the density fraction < 1.9 g/cm ³and by thermal extraction desorption-gas chromatography-mass spectrometry (TED-GC/MS) and the concentrations ranged from 3.7 to 480 mg TRWP/g. Strong and statistically signif- icant correlations with TRWPs were found for 2-hydroxybenzothiazole and 2-aminobenzothiazole, indi- cating that these substances may be suitable markers of TRWPs. The mass distribution of TRWPs in road dust suggests that the main mass fraction formed on roads consists of coarse particles ( > 100 μm). Data for a sedimentation basin indicate that the fine fraction ( < 50 μm) is preferentially transported by road runoffinto receiving waters. The size distribution and density data of TRWP gathered by three different quantitation approaches also suggest that aging of TRWPs leads to changes in their particle density. An improved understanding of the dynamics of TRWP properties is essential to assess the distribution and dissipation of this contaminant of emerging concern in the environment. KW - Tire Wear KW - Density separation KW - Microplastic KW - Urban PY - 2020 DO - https://doi.org/10.1016/j.watres.2020.116262 VL - 185 SP - 116262 PB - Elsevier Ltd. AN - OPUS4-51256 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Dittmann, Daniel A1 - Lucke, T. A1 - Ruhl, A. S. A1 - Winzenbacher, R. A1 - Jekel, M. A1 - Braun, Ulrike T1 - Untersuchungen zu biologischen, organischen und anorganischen Veränderungen granulierter Aktivkohlen während der Trinkwasseraufbereitung N2 - Untersuchungen zu biologischen, organischen und anorganischen Veränderungen granulierter Aktivkohlen während der Trinkwasseraufbereitung des Landeswasserversorgers Langenau. KW - Thermogravimetrie KW - Trinkwasser KW - Wasseraufbereitung KW - Aktivkohle PY - 2020 SN - 0083-6915 VL - 118 IS - 2 SP - 47 EP - 50 PB - Wiley-VCH Verlag GmbH & Co. KGaA AN - OPUS4-51434 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Becker, Roland A1 - Altmann, Korinna A1 - Sommerfeld, Thomas A1 - Braun, Ulrike T1 - Quantification of microplastics in a freshwater suspended organic matter using different thermoanalytical methods – outcome of an interlaboratory comparison N2 - A sedimented freshwater suspended organic matter fortified with particles of polyethylene (PE), polypropylene (PP), polystyrene (PS), and polyethylene terephthalate (PET) was employed in an interlaboratory comparison of thermoanalytical methods for microplastics identification and quantification. Three laboratories performed pyrolysis gas chromatography-mass spectrometry (Py-GC-MS), three others provided results using thermal extraction desorption followed by gas chromatography coupled to mass spectrometry (TED-GC-MS). One participant performed thermogravimetry-infrared spectroscopy (TGA-FTIR) and two participants used thermogravimetry coupled to mass spectrometry (TGA-MS). Further participants used differential scanning microscopy (DSC), a procedure based on micro combustion calorimetry (MCC) and a procedure based on elemental analysis. Each participant employed a different combination of sample treatment, calibration and instrumental Settings for polymer identification and quantification. Though there is obviously room for improvements regarding the between-laboratory reproducibility and the harmonization of procedures it was seen that the participants Performing Py-GC-MS, TED-GC-MS, and TGA-FTIR were able to correctly identify all polymers and to report reasonable quantification results in the investigated concentration range (PE: 20.0 μg/mg, PP: 5.70 μg/mg; PS: 2.20 μg/mg, PET: 18.0 μg/mg). Although for the other methods limitations exists regarding the detection of specific polymers, they showed potential as alternative approaches for polymer quantification in solid environmental matrices. KW - Interlaboratory comparison KW - Microplastics KW - Suspended organic matter KW - Pyrolysis PY - 2020 DO - https://doi.org/10.1016/j.jaap.2020.104829 VL - 148 SP - 1 EP - 6 PB - Elsevier B.V. AN - OPUS4-50977 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Wander, Lukas A1 - Vianello, A. A1 - Vollertsen, J. A1 - Westad, F. A1 - Braun, Ulrike A1 - Paul, Andrea T1 - Exploratory analysis of hyperspectral FTIR data obtained from environmental microplastics samples N2 - Hyperspectral imaging of environmental samples with infrared microscopes is one of the preferred methods to find and characterize microplastics. Particles can be quantified in terms of number, size and size distribution. Their shape can be studied and the substances can be identified. Interpretation of the collected spectra is a typical problem encountered during the analysis. The image datasets are large and contain spectra of countless particles of natural and synthetic origin. To supplement existing Analysis pipelines, exploratory multivariate data analysis was tested on two independent datasets. Dimensionality reduction with principal component analysis (PCA) and uniform manifold approximation and projection (UMAP) was used as a core concept. It allowed for improved visual accessibility of the data and created a chemical two-dimensional image of the sample. Spectra belonging to particles could be separated from blank spectra, reducing the amount of data significantly. Selected spectra were further studied, also applying PCA and UMAP. Groups of similar spectra were identified by cluster analysis using k-means, density based, and interactive manual clustering. Most clusters could be assigned to chemical species based on reference spectra. While the results support findings obtained with a ‘targeted analysis’ based on automated library search, exploratory analysis points the attention towards the group of unidientified spectra that remained and are otherwise easily overlooked. KW - Microplastics KW - FTIR KW - Exploratory analysis PY - 2020 DO - https://doi.org/10.1039/c9ay02483b VL - 12 IS - 6 SP - 781 EP - 791 PB - Royal Society of Chemistry AN - OPUS4-50396 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Paul, Andrea A1 - Wander, Lukas A1 - Becker, Roland A1 - Goedecke, Caroline A1 - Braun, Ulrike T1 - High-throughput NIR spectroscopic (NIRS) detection of microplastics in soil N2 - The increasing pollution of terrestrial and aquatic ecosystems with plastic debris leads to the accumulation of microscopic plastic particles of still unknown amount. To monitor the degree of contamination analytical methods are urgently needed, which help to quantify microplastics (MP). Currently, time-costly purified materials enriched on filters are investigated both by micro-infrared spectroscopy and/or micro-Raman. Although yielding precise results, these techniques are time consuming, and are restricted to the analysis of a small part of the sample in the order of few micrograms. To overcome these problems, here we tested a macroscopic dimensioned NIR process-spectroscopic method in combination with chemometrics. For calibration, artificial MP/soil mixtures containing defined ratios of polyethylene, polyethylene terephthalate, polypropylene, and polystyrene with diameters < 125 µm were prepared and measured by a process FT-NIR spectrometer equipped with a fiber optic reflection probe. The resulting spectra were processed by chemometric models including support vector machine regression (SVR), and partial least squares discriminant analysis (PLS-DA). Validation of models by MP mixtures, MP-free soils and real-world samples, e.g. and fermenter residue, suggest a reliable detection and a possible classification of MP at levels above 0.5 to 1.0 mass% depending on the polymer. The benefit of the combined NIRS chemometric approach lies in the rapid assessment whether soil contains MP, without any chemical pre-treatment. The method can be used with larger sample volumes and even allows for an online prediction and thus meets the demand of a high-throughput method. KW - Microplastics KW - Soil KW - Chemometrics KW - PLS-DA KW - Support vector machines KW - Near Infrared Spectroscopy PY - 2018 DO - https://doi.org/10.1007/s11356-018-2180-2 SN - 1614-7499 SN - 0944-1344 VL - 26 IS - 8 SP - 7364 EP - 7374 PB - Springer AN - OPUS4-45405 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike T1 - Messen ist nicht alles, aber ohne Messmethoden ist alles nichts! N2 - Es wir der Wissenstand zur Analytik von Mikroplastik dargestellt, sowie die Notwendigkeit der Hamonisierung der Verfahren. T2 - Jahrestagung der deutschen Labordienstleistungsbranche CY - Gießen, Berlin DA - 05.09.2019 KW - TED-GC-MS KW - Mikroplastik KW - Analytik PY - 2019 AN - OPUS4-49994 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike A1 - Bannick, C.G. T1 - Analytik von Mikroplastik - Verfahren-Ergebnisse-Harmonisierung N2 - Die zunehmenden Nachweise von Kunststoffen in verschiedenen Umweltmedien erfahren in den letzten Jahren weltweit ein zunehmendes wissenschaftliches, gesellschaftliches und politisches Interesse. Besondere Aufmerksamkeit erfahren Kunststoffpartikel von 1-1.000 µm sogenanntes Mikroplastik. Es wurde mittlerweile in Wasser, Böden, Luft und Biota nachgewiesen. Obwohl die Wirkungen auf die natürlichen Umweltfunktionen noch unklar sind und kontrovers diskutiert werden, ist allein durch die ubiquitäre Präsenz eine systematische Erfassung von Eintragsquellen und Verbleib der Partikel über alle Umweltmedien notwendig, um Maßnahmen zur Reduktion der Einträge zu veranlassen. Für eine zuverlässige Erfassung sind medienübergreifende Verfahren zu entwickeln, welche die drei analytischen Schritte, Probennahme, Probenaufbereitung und Detektion beinhalten. Die Verfahren adressieren unterschiedliche Zielgruppen, welche verschiedene analytische Informationen fordern. Für die akademische Bewertung werden Verfahren mit hohem analytischen Aufwand favorisiert, welche möglichst detaillierte Information über Partikelsorte, -größe, -zahl und -oberflächenbeschaffenheit geben. Im Gegensatz dazu, fordern Anwender aus der Praxis schnelle und einfache Monitoring Verfahren, welche die Bestimmung eines Massengehaltes, bzw. eines Grenzwertes zulassen. Allen ist gemeinsam, dass ein hoher interdisziplinärer Anspruch besteht, der sowohl technische Voraussetzungen zur repräsentativen Probennahme und die Limitationen von Detektionsverfahren berücksichtigt, als auch die Fachkompetenz von Biologen, Umweltwissenschaften, Analytiker und Polymerwissenschaftler einschließt. Verfahren zum Mikroplastiknachweis müssen auch verschieden Proben mit sehr unterschiedlichen Gehalten an natürlicher Matrix gerecht werden. So müssen allein für Wasser Verfahren entwickelt werden, die Mikroplastik verschiedener Form (sphärische Partikel, komplexe Bruchstücke, Fasern) aus Medien mit sehr variablen Partikelanteil (u.a. Trinkwasser, Oberflächenwasser, Abwasser) repräsentativ erfassen sowie schnell und kostengünstig sind – einschließlich der notwendigen Probenaufbereitung. Die jeweiligen Detektionsverfahren müssen z.B. einzelne Partikel oder integrale Gehalte über drei Dekadenklassen für thermoplastische und insbesondere auch elastomere Partikelsorten erfassen können. T2 - GDCh Wissenschaftsforum Chemie CY - Aachen, Germany DA - 15.09.2019 KW - TED-GC-MS KW - Mikroplastik KW - Analytik PY - 2019 AN - OPUS4-49995 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike T1 - Analytical opportunities & Challenges in (ocean) microplastics N2 - An overview about the analytical opportunities and challenges in microplastic detection is given. T2 - ISO TC61/ Annual bioplastics & ocean microplastics symposium CY - Chengdu, People's Republic of China DA - 29.09.2019 KW - Harmonisation KW - Microplastics KW - TED-GC-MS PY - 2019 AN - OPUS4-50000 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike T1 - Mikroplastik als analytische Herausforderung N2 - Der Wissenstand zum Thema Mikroplastik und Harmonisierung wird dargestellt. T2 - Laborleitertreffen Futtermittelmonitoring CY - Bonn, Germany DA - 21.11.2019 KW - Hamonisierung KW - Mikroplastik KW - TED-GC-MS PY - 2019 AN - OPUS4-50001 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike T1 - Analytic of microplastics in environmental samples N2 - An overview about the existing possibilities for mircoplastic analysis is given. T2 - 2nd International Symposium on Microplastics CY - Seoul, Korea DA - 02.12.2019 KW - Harmonisation KW - Microplastics KW - TED-GC-MS PY - 2019 AN - OPUS4-50003 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike A1 - Bannick, C.G. T1 - Analytik von Mikroplastik in komplexen Umweltproben N2 - Der Sachstand zur Analytik von Mikroplastik in Umweltproben wird dargestellt. T2 - Fachausstellung Kunststoffanalytik CY - Lüdenscheid, Germany DA - 04.12.2019 KW - TED-GC-MS KW - Mikroplastik KW - Analytik PY - 2019 AN - OPUS4-50005 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Wander, Lukas A1 - Lommel, Lukas A1 - Meyer, Klas A1 - Braun, Ulrike A1 - Paul, Andrea T1 - Development of a low-cost method for quantifying microplastics in soils and compost using near-infrared spectroscopy N2 - Near-infrared (NIR) spectroscopy is a promising candidate for low-cost, nondestructive, and high-throughput mass quantification of micro¬plastics in environmental samples. Widespread application of the technique is currently hampered mainly by the low sensitivity of NIR spectroscopy compared to thermo-analytical approaches commonly used for this type of analysis. This study shows how the application of NIR spectroscopy for mass quantification of microplastics can be extended to smaller analyte levels by combining it with a simple and rapid microplastic enrichment protocol. For this purpose, the widely used flotation of microplastics in a NaCl solution, accelerated by centrifugation, was chosen which allowed to remove up to 99 % of the matrix at recovery rates of 83–104 %. The spectroscopic measurements took place directly on the stainless-steel filters used to collect the extracted particles to reduce sample handling to a minimum. Partial least squares regression (PLSR) models were used to identify and quantify the extracted microplastics in the mass range of 1–10 mg. The simple and fast extraction procedure was systematically optimized to meet the requirements for the quantification of microplastics from common PE-, PP-, and PS-based packaging materials with a particle size < 1 mm found in compost or soils with high natural organic matter content (> 10 % determined by loss on ignition). Microplastics could be detected in model samples at a mass fraction of 1 mg g-1. The detectable microplastic mass fraction is about an order of magnitude lower compared to previous studies using NIR spectroscopy without additional enrichment. To emphasize the cost-effectiveness of the method, it was implemented using some of the cheapest and most compact NIR spectrometers available. KW - NIR KW - Soil KW - compost KW - PLSR PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-546405 DO - https://doi.org/10.1088/1361-6501/ac5e5f SN - 0957-0233 VL - 33 IS - 7 SP - 075801 EP - 075814 PB - IOP Publishing Ltd. CY - UK AN - OPUS4-54640 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Wander, Lukas A1 - Lommel, Lukas A1 - Braun, Ulrike A1 - Meyer, Klas A1 - Paul, Andrea T1 - Development of a Low-Cost Method for Quantifying Microplastics in Soils and Compost Using Near-Infrared Spectroscopy N2 - Near-infrared (NIR) spectroscopy is a promising candidate for low-cost, nondestructive, and highthroughput mass quantification of microplastics in environmental samples. Widespread application of the technique is currently hampered mainly by the low sensitivity of NIR spectroscopy compared to thermoanalytical approaches commonly used for this type of analysis. This study shows how the application of NIR spectroscopy for mass quantification of microplastics can be extended to smaller analyte levels by combining it with a simple and rapid microplastic enrichment protocol. For this purpose, the widely used flotation of microplastics in a NaCl solution, accelerated by centrifugation, was chosen which allowed to remove up to 99 % of the matrix at recovery rates of 83–104 %. The spectroscopic measurements took place directly on the stainless-steel filters used to collect the extracted particles to reduce sample handling to a minimum. Partial least squares regression (PLSR) models were used to identify and quantify the extracted microplastics in the mass range of 1–10 mg. The simple and fast extraction procedure was systematically optimized to meet the requirements for the quantification of microplastics from common PE-, PP-, and PS-based packaging materials with a particle size < 1 mm found in compost or soils with high natural organic matter content (> 10 % determined by loss on ignition). Microplastics could be detected in model samples at a mass fraction of 1 mg g-1. The detectable microplastic mass fraction is about an order of magnitude lower compared to previous studies using NIR spectroscopy without additional enrichment. To emphasize the cost-effectiveness of the method, it was implemented using some of the cheapest and most compact NIR spectrometers available. KW - Mikroplastik KW - NIR KW - Sensor KW - Kompost KW - Multivariat PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-552605 DO - https://doi.org/10.1088/1361-6501/ac5e5f VL - 33 IS - 7 SP - 1 EP - 13 PB - IOP Publishing Ltd. CY - Bristol AN - OPUS4-55260 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Müller, Axel A1 - Goedecke, Caroline A1 - Eisentraut, Paul A1 - Piechotta, Christian A1 - Braun, Ulrike T1 - Microplastic analysis using chemical extraction followed by LC‑UV analysis: a straightforward approach to determine PET content in environmental samples N2 - Background: The ubiquitous occurrence of microplastic particles in marine and aquatic ecosystems was intensively investigated in the past decade. However, we know less about the presence, fate, and input paths of microplastic in terrestrial ecosystems. A possible entry path for microplastic into terrestrial ecosystems is the agricultural application of sewage sludge and solid bio-waste as fertilizers. Microplastic contained in sewage sludge also includes Polyethylene terephthalate (PET), which could originate as fiber from textile products or as a fragment from packaging products (foils, bottles, etc.). Information about microplastic content in such environmental samples is limited yet, as most of the used analytical methods are very time-consuming, regarding sample preparation and detection, require sophisticated analytical tools and eventually need high user knowledge. Results: Here, we present a simple, specific tool for the analysis of PET microplastic particles based on alkaline extraction of PET from the environmental matrix and subsequent determination of the monomers, terephthalic acid, using liquid chromatography with UV detection (LC-UV). The applicability of the method is shown for different types of PET in several soil-related, terrestrial environmental samples, e.g., soil, sediment, compost, fermentation residues, but also sewage sludge, suspended particles from urban water management systems, and indoor dust. Recoveries for model samples are between 94.5 and 107.1%. Limit of determination and limit of quantification are absolute masses of 0.031 and 0.121 mg PET, respectively. In order to verify the measured mass contents of the environmental samples, a method comparison with thermal extraction-desorption-gas chromatography–mass spectrometry (TED-GC/MS) was conducted. Both methods deliver similar results and corroborated each other. PET mass contents in environmental samples range from values below LOQ in agriculture soil up to 57,000 mg kg−1 in dust samples. Conclusions: We demonstrate the potential of an integral method based on chemical extraction for the Determination of PET mass contents in solid environmental samples. The method was successfully applied to various matrices and may serve as an analytical tool for further investigations of PET-based microplastic in terrestrial ecosystems. KW - Soil KW - Analysis KW - Microplastic KW - PET KW - LC-UV PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-509060 DO - https://doi.org/10.1186/s12302-020-00358-x IS - 32 SP - 85 PB - Springer Open CY - Berlin AN - OPUS4-50906 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike A1 - Altmann, Korinna A1 - Eisentraut, Paul A1 - Ricking, M. A1 - Reiche, M. A1 - Barthel, A.-K. A1 - Bannick, C. G. T1 - New sampling strategies for optimised microplastic analysis in surface waters N2 - The occurrence of microplastic particles (MP) in nearly each environmental compartment but especially in water is well documented in an increasing amount of publication within the last few years. However, a comprehensive understanding of MP occurrence, sources and pathways in the environment as well as effects of MP on the environment is still unclear. Many of the studies are not compatible, because they used different sampling, sample preparation and detection techniques, resulting in a very broad variation and very different dimension of archived results. Hence a useful avoiding strategy for reduction the MP entry in the environment is not possible. Furthermore, the existing methods are often very time consuming, need user knowledge and labour force, which is often not acceptable for the routine monitoring of an action or sanction. The call for harmonised methods become loud and louder and promising efforts are done in this area. However, the scientific discussion is often focus on advantages and disadvantages of detection tools and sample preparation, less initiative is given to the sampling strategy. That is a pity, because every environmental scientist knows, a good analytical result depend on a good sampling strategy. The sampling of water seems on the first view easy because the sampling process include already an up concentration of the MP by means of sieves and filters. However, this filtering of water is often limited by filter cake on the sieves, consisting of high amount of natural particular matter. Hence often a small water volume is filtered, which will not own a representative amount of trait carriers or complex sieves cascades are used, which need a high technical equipment. Recently we presented a fast thermoanalytical method for the detection of polymers fractions in water samples. This method, thermo-extraction desorption gas chromatography mass spectrometry (TED-GC-MS) can analyse MP in water filtrates without sample preparations within 2-3 hours. Advantageous is the additional analysis of rubber particles from tires beside the MP. In order to come even faster we now start to develop new sampling strategies, which should be use for different water systems. It is intended to reduce the effort in the field and to elongate times for sampling to obtain a representatives result without random effects. In the present presentation we will show first result using a sedimentation box for sampling of MP from surface water. Advantages and disadvantages will be discussed and comparison to data from conventional fractionated filtration will be given. T2 - SETAC CY - Helsinki, Finland DA - 26.05.2019 KW - Microplastics KW - Sampling KW - Sampling techniques KW - Water PY - 2019 AN - OPUS4-48118 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike T1 - Detection of microplastics using thermoanalytical methods  N2 - The presentation summarized the use of thermal analytic methods for the detection of microplastics. T2 - Workshop State of teh art of anylatical methodes for reliable detection of micro- and nanoplastics CY - Ispra, Italy DA - 13.05.2019 KW - microplastics, sampling, sampling techniques, water PY - 2019 AN - OPUS4-48119 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Braun, Ulrike A1 - Bannick, C.G. A1 - Szewzyk, R. A1 - Ricking, M. A1 - Schniegler, S. A1 - Obermaier, N. A1 - Barthel, A. K. A1 - Altmann, Korinna A1 - Eisentraut, Paul T1 - Development and testing of a fractionated filtration for sampling of microplastics in water N2 - A harmonization of sampling, sample preparation and detection is pivotal in order to obtain comparable data on microplastics (MP) in the environment. This paper develops and proposes a suitable sampling concept for waterbodies that considers different plastic specific properties and influencing factors in the environment. Both artificial water including defined MP fractions and the discharge of a wastewater treatment plant were used to verify the derived sampling procedure, sample preparation and the subsequent analysis of MP using thermal extraction-desorption gas chromatography - mass spectrometry (TED-GC-MS). A major finding of this paper is that an application of various particle size classes greatly improves the practical handling of the sampling equipment. Size classes also enable the TED-GC-MS to provide any data on the MP size distribution, a substantial sampling property affecting both the necessary sampling volume and the optimal sampling depth. In the artificial water with defined MP fractions, the recovery rates ranged from 80 to 110%, depending on the different MP types and MP size classes. In the treated wastewater, we found both Polyethylene and polystyrene in different size classes and quantities. KW - Microplastics KW - Sampling KW - Sampling techniques KW - Water PY - 2019 DO - https://doi.org/10.1016/j.watres.2018.10.045 SN - 0043-1354 VL - 149 SP - 650 EP - 658 PB - Elsevier AN - OPUS4-47200 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike T1 - Microplastics: Degradation properties of polymers in the environment N2 - The accumulation of plastics leaked into the environment achieve a high perception in the last years. It is uncontroversial that often an uncontrolled, high consumption of plastic products and a poor waste management resulted in plastic findings all over in the environment. However, less is known about the relevant pathways and fate of plastic in the environment. Especially plastic fragments smaller 5 mmm, so called microplastics, achieve a high attention, because their amount and risks arising from these fragments are still unknown. The basic mechanisms of plastic or more precise, of different polymer degradation processes are often not considered, resulting in unrealistic statements. For generation of fundamental understanding of this problem we developed in the last years a fast thermoanalytical method for identification and mass quantification of microplastics in complex samples. The presentation will also give a short introduction about polymer degradation. T2 - Environmental Geosciences Seminar CY - Vienna, Austria DA - 19.11.2018 KW - Microplastic KW - Analysis KW - Standardisation PY - 2018 AN - OPUS4-47182 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -