TY - JOUR A1 - Scala-Benuzzi, M. L. A1 - Takara, E. A. A1 - Alderete, M. A1 - Soler-Illia, G. J. A. A. A1 - Schneider, Rudolf A1 - Raba, J. A1 - Messina, G. A. T1 - Ethinylestradiol quantification in drinking water sources using a fluorescent paper based immunosensor N2 - In this work we report a novel paper-based analytical device read-out via LED-induced fluorescence detection (FPAD) for the quantification of the emerging pollutant ethinylestradiol (EE2) in river water samples. The PAD was used as a reaction platform for a competitive enzyme immunoassay. For the PAD development, microzones of filter paper, printed by a wax printing method, were modified with amino-functionalized SBA-15 and subsequently, anti-EE2 specific antibodies were covalently immobilized. The determination of EE2 in water was carried out by adding a fixed concentration of EE2 conjugated with the enzyme horseradish peroxidase (HRP) to samples and standards. Then, the FPAD were added and incubated for 10 min. Finally, the detection was performed by the reaction of 10-acetyl-3,7-dihydroxyphenoxazine (ADHP) whose oxidation is catalyzed by HRP in the presence of H2O2, obtaining the highly fluorescent resorufin (R). Resorufin was detected by LED excitation at 550 nm, observing emission at 585 nm. The EE2 concentration in the samples was inversely proportional to the relative fluorescence obtained from the enzymatic reaction products. The FPAD assay showed a detection Limit (LOD) of 0.05 ng L−1 and coefficients of variation (CV) below 4.5% within-assay and below 6.5% between-assay, respectively. The results obtained show the potential suitability of our FPAD for the selective and sensitive quantification of EE2 in river water samples. In addition, it has the PADs advantages of being disposable, easy to apply and inexpensive. KW - Immunoassay KW - Arzneimittel KW - Antikörper KW - LED-induzierte Fluoreszenz KW - Paper based devices KW - Mesoporöses Silica KW - Partikel PY - 2018 DO - https://doi.org/10.1016/j.microc.2018.05.038 SN - 0026-265X VL - 141 SP - 287 EP - 293 PB - Elsevier B.V. CY - Amsterdam, NL AN - OPUS4-45269 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Scala-Benuzzi, M. L. A1 - Raba, J. A1 - Soler-Illia, G. J. A. A. A1 - Schneider, Rudolf A1 - Messina, G. A. T1 - Novel electrochemical paper-based immunocapture assay for the quantitative determination of ethinylestradiol in water samples N2 - We report a novel and innovative electrochemical paper-based immunocapture assay (EPIA) to address the need for ultrasensitive detection of emerging pollutants without regulatory status and whose effects on environment and human health are not completely yet understood. In particular, we present the application of this system toward highly sensitive detection of the emerging pollutant ethinyl estradiol (EE2). The EPIA approach is based on the use of paper microzones modified with silica nanoparticles (SNs) and anti-EE2 specific antibodies for capture and preconcentration of EE2 from river water samples. After the preconcentration procedure, the paper microzones are placed onto a screen-printed carbon electrode modified with electrochemically reduced graphene (RG). The bound EE2 is subsequently desorbed adding a diluted solution of sulfuric acid on the paper microzones. Finally, recovered EE2 is electrochemically detected by OSWV. The proposed novel methodology showed an appropriate LOD and linear range for the quantification of EE2 for water samples with different origins. The nonsophisticated equipment required, the adequate recovery values obtained (from 97% to 104%, with a RSD less than 4.9%), and the appropriate LOD and linear range value (0.1 ng L−1 and 0.5−120 ng L−1, respectively) achieved by our immunocapture sensor present significant analytical figures of merit, particularly when the routine quantification of EE2 is considered. In addition, our System was based on electrochemical paper-based technology, which allows obtainment of portable, easy-to-use, inexpensive, and disposable devices. The EPIA can also serve as a general-purpose immunoassay platform applicable to quantitation of other drugs and emerging pollutants in environmental samples. KW - Immunoassay KW - EE2 KW - Estrogens KW - Screen-printed electrode KW - Square-wave voltammetry PY - 2018 DO - https://doi.org/10.1021/acs.analchem.8b00028 SN - 0003-2700 SN - 1520-6882 VL - 90 IS - 6 SP - 4104 EP - 4111 PB - American Chemical Society CY - Washington, D.C., USA AN - OPUS4-44690 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Carvalho, J.J.D. A1 - Bahlmann, Arnold A1 - Schneider, Rudolf T1 - ELISA goes Berlin - Bestimmung von Carbamazepin, Cetrizin und Coffein in Berliner Gewässern N2 - Sensitive und selektive Immunoassays unterstützen Untersuchungen im Vollzug der EU Wasserrahmenrichtlinie. Sie erlauben, eine hohe Anzahl von Proben ohne Anreicherungsschritte zu untersuchen und ermöglichen daher den Aufbau eines zeitlich und räumlich engmaschigen Messnetzes. Mit Hilfe von ELISA (enzymelinked immunosorbent assay) wurde ein Gewässerscreening in Berlin durchgeführt. KW - ELISA KW - Koffein KW - Immunoassay KW - Oberflächenwasser KW - Wasseranalytik PY - 2012 SN - 0016-3538 VL - 4 SP - 278 EP - 281 PB - GIT-Verlag CY - Darmstadt AN - OPUS4-27518 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ecke, Alexander A1 - Westphalen, Tanja A1 - Hornung, J. A1 - Voetz, M. A1 - Schneider, Rudolf T1 - A rapid magnetic bead-based immunoassay for sensitive determination of diclofenac N2 - Increasing contamination of environmental waters with pharmaceuticals represents an emerging threat for the drinking water quality and safety. In this regard, fast and reliable analytical methods are required to allow quick countermeasures in case of contamination. Here, we report the development of a magnetic bead-based immunoassay (MBBA) for the fast and cost-effective determination of the analgesic diclofenac (DCF) in water samples, based on diclofenac-coupled magnetic beads and a robust monoclonal anti-DCF antibody. A novel synthetic strategy for preparation of the beads resulted in an assay that enabled for the determination of diclofenac with a significantly lower limit of detection (400 ng/L) than the respective enzyme-linked immunosorbent assay (ELISA). With shorter incubation times and only one manual washing step required, the assay demands for remarkably shorter time to result (< 45 min) and less equipment than ELISA. Evaluation of assay precision and accuracy with a series of spiked water samples yielded results with low to moderate intra- and inter-assay variations and in good agreement with LC–MS/MS reference analysis. The assay principle can be transferred to other, e.g., microfluidic, formats, as well as applied to other analytes and may replace ELISA as the standard immunochemical method. KW - Immunoassay KW - Magnetic beads KW - Diclofenac KW - Water analysis KW - LC-MS/MS PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-542346 DO - https://doi.org/10.1007/s00216-021-03778-7 SN - 1618-2650 VL - 414 SP - 1563 EP - 1573 PB - Springer CY - Heidelberg AN - OPUS4-54234 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Martínez, N.A. A1 - Pereira, S.V. A1 - Bertolino, F.A. A1 - Schneider, Rudolf A1 - Messina, G.A. A1 - Raba, J. T1 - Electrochemical detection of a powerful estrogenic endocrine disruptor: ethylestradiol in water samples through bioseparation procedure N2 - The synthetic estrogen ethinylestradiol (EE2) is an active component of oral contraceptives (OCs), considered as an endocrine disrupting compound (EDC). It is excreted from humans and released via sewage treatment plant effluents into aquatic environments. EDCs are any environmental pollutant chemical that, once incorporated into an organism, affects the hormonal balance of various species including humans. Its presence in the environment is becoming of great importance in water quality. This paper describes the development of an accurate, sensitive and selective method for capture, preconcentration and determination of EE2 present in water samples using: magnetic particles (MPs) as bioaffinity support for the capture and preconcentration of EE2 and a glassy carbon electrode modified with multi-walled carbon nanotubes (MWCNTs/GCE) as detection system. The capture procedure was based on the principle of immunoaffinity, the EE2 being extracted from the sample using the anti-EE2 antibodies (anti-EE2 Ab) which were previously immobilized on MPs. Subsequently the analyte desorption was done employing a sulfuric acid solution and the determination of the EE2 in the pre-concentrated solution was carried out by square wave voltammetry (SWV). This method can be used to determine EE2 in the range of 0.035–70 ng L-1 with a detection limit (LOD) of 0.01 ng L-1 and R.S.D. < 4.20%. The proposed method has been successfully applied to the determination of EE2 in water samples and it has promising analytical applications for the direct determination of EE2 at trace levels. KW - Magnetic particles KW - Bioseparation KW - Ethinylestradiol KW - Electrochemistry KW - Immunoassay KW - Elektrochemischer Sensor KW - Endokrine Disruptoren KW - Hormone KW - EE2 KW - Östrogene PY - 2012 DO - https://doi.org/10.1016/j.aca.2012.02.033 SN - 0003-2670 SN - 1873-4324 SN - 0378-4304 VL - 723 SP - 27 EP - 32 PB - Elsevier CY - Amsterdam AN - OPUS4-25656 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Riedel, Soraya A1 - Hülagü, Deniz A1 - Bennet, Francesca A1 - Carl, Peter A1 - Flemig, Sabine A1 - Schmid, Thomas A1 - Schenk, J. A. A1 - Hodoroaba, Vasile-Dan A1 - Schneider, Rudolf T1 - Electrochemical Immunomagnetic Ochratoxin A Sensing: Steps Forward in the Application of 3,3’,5,5’- Tetramethylbenzidine in Amperometric Assays N2 - Electrochemical methods offer great promise in meeting the demand for user-friendly on-site devices for Monitoring important parameters. The food industry often runs own lab procedures, for example, for mycotoxin analysis, but it is a major goal to simplify analysis, linking analytical methods with smart technologies. Enzyme-linked immunosorbent assays, with photometric detection of 3,3’,5,5’-tetramethylbenzidine (TMB),form a good basis for sensitive detection. To provide a straightforward approach for the miniaturization of the detectionstep, we have studied the pitfalls of the electrochemical TMB detection. By cyclic voltammetry it was found that the TMB electrochemistry is strongly dependent on the pH and the electrode material. A stable electrode response to TMB could be achieved at pH 1 on gold electrodes. We created a smartphonebased, electrochemical, immunomagnetic assay for the detection of ochratoxin A in real samples, providing a solid basis forsensing of further analytes. KW - Ochratoxin A KW - Amperometry KW - Cyclic voltammetry KW - Electrochemistry KW - Immunoassay PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-530421 DO - https://doi.org/10.1002/celc.202100446 N1 - Geburtsname von Riedel, Soraya: Höfs, S. - Birth name of Riedel, Soraya: Höfs, S. VL - 8 IS - 13 SP - 2597 EP - 2606 AN - OPUS4-53042 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Carl, Peter A1 - Sarma, Dominik A1 - Gregório, Bruno J. R. A1 - Hoffmann, Kristin A1 - Lehmann, Andreas A1 - Rurack, Knut A1 - Schneider, Rudolf T1 - Wash-Free Multiplexed Mix-and-Read Suspension Array Fluorescence Immunoassay for Anthropogenic Markers in Wastewater N2 - Pharmaceuticals, certain food ingredients, and mammalian endogenous metabolic products in wastewater are mostly of human origin. They are anthropogenic markers. Proper knowledge of their levels in wastewater helps to track sources of pollutants in natural waters and allows for calculation of removal efficiencies in wastewater Treatment plants. Here, we describe the development and application of an indirect competitive, multiplexing suspension Array fluorescence immunoassay (SAFIA) for the detection of carbamazepine (CBZ), diclofenac (DCF), caffeine (CAF), and isolithocholic acid (ILA) in wastewater, covering those classes of anthropogenic markers. The assay consists of haptens covalently conjugated to fluorescence-encoded polystyrene core/silica shell microparticles to create a site for competitive binding of the antibodies (Abs). Bound Abs are then stained with fluorophore-labeled Abs. Encoding and signaling fluorescence of the particles are determined by an automated flow cytometer. For compatibility of the immunoassay with the 96-well microtiter plate format, a stop reagent, containing formaldehyde, is used. This enables a wash-free procedure while decreasing time-to-result. Detection limits of 140 ± 40 ng/L for CBZ, 180 ± 110 ng/L for CAF, 4 ± 3 ng/L for DCF, and 310 ± 70 ng/L for ILA are achieved, which meet the sensitivity criteria of wastewater analysis. We demonstrate the applicability of SAFIA to real wastewater samples from three different wastewater Treatment plants, finding the results in good agreement with LC-MS/MS. Moreover, the accuracy in general exceeded that from classical ELISAs. We therefore propose SAFIA as a quick and reliable approach for wastewater analysis meeting the requirements for process analytical technology. KW - Suspension Array KW - Immunoassay KW - Carbamazepine KW - Caffeine KW - Diclofenac KW - SAFIA PY - 2019 DO - https://doi.org/10.1021/acs.analchem.9b03040 SN - 0003-2700 VL - 91 IS - 20 SP - 12988 EP - 12966 PB - ACS Publications AN - OPUS4-49305 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Devi, Sarita A1 - Goel, S. A1 - Rani, S. A1 - Schneider, Rudolf J. A1 - Rohilla, R. A1 - Kumari, Surbhi A1 - Prabhakar, N. T1 - MOF-metal nanohybrid-assisted charge transfer amplification for electrochemical biosensing of the MUC1 cancer biomarker N2 - Cancer is a severe disease characterized by high mortality and complex pathophysiology; however, its early and accurate diagnosis remains inadequate. Conventional diagnostic approaches often fall short, particularly for dense tissues, and are frequently invasive, costly, and of limited availability. This reinforces the need for a compact, economical, and ultrasensitive assay that is operationally simple and interpretable. We present an efficient electrochemical detection platform for the cancer biomarker mucin 1 (MUC1). A fluorine-doped tin oxide (FTO) surface was modified with an iron-based metal–organic framework (FeMOF) intercalated with palladium nanorods (PdNR). FeMOF was prepared using Fe3+/Fe2+ precursors at a 1.2/1 mmol ratio and dual ligands, i.e. tetrahydroxy-1,4-benzoquinone and 2-aminobenzene-1,4-dicarboxylic acid. AntiMUC1 antibodies were immobilized on a modified electrode via p-phenylenediamine (PDA) (FTO/FeMOF@PdNR/PDA/antiMUC1Ab) and evaluated using electrochemical impedance spectroscopy (EIS) and voltammetry. The designed sensor demonstrated an excellent binding affinity for the MUC1 antigen. Among these techniques, the EIS method stands out for its technical performance, as evidenced by the high sensitivity (detection limit 0.074 fg mL−1), quantification limit 0.24 fg mL−1, and high analytical sensitivity (1.39 × 103 Ω fg−1 mL−1 cm−2). The negligible cross-reactivity with interferent biomolecules, rapid response (10-minute equilibrium), regenerability up to 5 cycles, high reproducibility (RSD ∼1–3%), and long-term stability (up to 35 days) further validate the suitability of the proposed MUC1 immunosensor. This study presents an ultrasensitive biosensor that is compact, cost-effective, and easy for individuals at home to use after further development into a kit-based end product. Moreover, its excellent functionality for spiked serum samples shows promise for next-generation clinical diagnostics. KW - Antibodies KW - Metal-organic framework KW - Immunoassay PY - 2026 DO - https://doi.org/10.1039/D6AN00018E SP - 1 EP - 11 PB - Royal Society of Chemistry CY - London AN - OPUS4-65726 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -