TY - JOUR A1 - Yang, L. A1 - Vogl, Jochen A1 - Mann, J. A1 - Kraft, R. A1 - Vocke, R. A1 - Pramann, A. A1 - Eberhardt, J. A1 - Rienitz, O. A1 - Lee, K.-S. A1 - Lim, J. S. A1 - Sobina, E. A1 - Song, P. A1 - Wang, J. A1 - Mester, Z. A1 - Meija, J. T1 - Copper isotope delta measurements in high purity materials: CCQM-P213 pilot study N2 - Accurate and precise isotope ratio measurements of heavy elements are playing an increasinglyimportant role in modern analytical sciences and have numerous applications. Today, isotope ratio measurements are typically performed with two principal techniques: thermal ionization mass spectrometry (TIMS) and multiple collector-inductively coupled plasma mass spectrometry (MC-ICP-MS). To obtain accurate results by mass spectrometry, isotopic certified reference materials (iCRMs) are needed for mass bias correction and for the validation of the method used for analysis.Thus, it is of paramount importance to achieve measurement comparability of all data reported, and to assess measurement capability of each CRM producer/National Metrology Institute (NMI). Therefore, the international comparison (CCQM-P213) was performed to assess the analytical capabilities of NMIs for the accurate determination of copper isotope ratio delta values in high purity materials. The study was proposed by the coordinating laboratories, National Research Council Canada (NRC), National Institute of Standards and Technology (NIST), Bundesanstalt für Materialforschung und -prüfung (BAM) and Physikalisch-Technische Bundesanstalt (PTB), as an activity of the Isotope Ratio Working Group (IRWG) of the Consultative Committee for Amount of Substance - Metrology in Chemistry and Biology (CCQM). Participants included six NMIs and one designated institute (DI) from the six countries. Although no measurement method was prescribed by the coordinating laboratories, MC-ICP-MS with either standard-sample bracketing (SSB) or combined SSB with internal normalization (C-SSBIN) models for mass bias correction were recommended. Results obtained from the six NMIs and one DI were in good agreement. KW - Comparability KW - Traceability KW - Metrology KW - Isotope delta KW - Copper PY - 2023 DO - https://doi.org/10.1088/0026-1394/60/1A/08019 VL - 60 IS - 1A SP - 1 EP - 23 PB - IOP Science AN - OPUS4-58040 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - RPRT A1 - Vogl, Jochen A1 - Becker, Dorit A1 - Koenig, Maren A1 - Yim, Y.-H. A1 - Lee, K.-S. A1 - Goenaga-Infante, H. A1 - Malinowskiy, D. A1 - Hill, S. A1 - Ren, T. A1 - Wang, J. A1 - Vocke, R. D. A1 - Murphy, K. A1 - Nonose, N. A1 - Rienitz, O. A1 - Noordmann, J. T1 - Certification Report for the Isotopic Reference Materials ERM-AE142 and ERM-EB400 N2 - Lead (Pb) isotope amount ratios are commonly used in applications ranging from archaeology and forensic sciences to terrestrial and extra-terrestrial geochemistry. Despite their utility and frequency of use, only three certified isotope amount ratio reference materials are currently available for Pb: NIST SRMs 981, 982 and 983. Because SRM 981 has a natural Pb isotopic composition, it is mainly used for correcting instrumental mass discrimination or fractionation. This means that, at present, there are no other certified isotope reference materials with natural Pb isotopic composition that could be used for validating or verifying an analytical procedure involving the measurement of Pb isotope amount ratios. To fill this gap, two new reference materials, both certified for their Pb isotopic composition, have been produced together with a complete uncertainty assessment. These new reference materials offer SI traceability and an independent means of validating or verifying analytical procedures used to produce Pb isotope amount ratio measurements. ERM-EB400 is a bronze material containing a nominal Pb mass fraction of 45 mg/kg. ERM-AE142 is a high purity solution of Pb with a nominal mass fraction of 100 mg/kg. Both materials have been specifically produced to assist analysts in verifying or validating their analytical procedures. Note that while one of these reference materials requires the chemical separation of Pb from its matrix (ERM-EB400), the other does not (ERM-AE142). Details on the certification of these isotope reference materials are provided in this report. KW - Lead isotopic composition KW - Isotope ratio KW - Reference material KW - Mass spectrometry PY - 2016 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-392060 SP - 1 EP - 16 PB - Bundesanstalt für Materialforschung und -prüfung (BAM) CY - Berlin AN - OPUS4-39206 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Semkova, V. A1 - Reimer, P. A1 - Altzitzoglou, T. A1 - Plompen, A.J.M. A1 - Quétel, C. A1 - Sudár, S. A1 - Vogl, Jochen A1 - Koning, A.J. A1 - Qaim, S.M. A1 - Smith, D.L. T1 - Neutron activation cross sections on lead isotopes KW - Cross sections KW - Neutron activation KW - Lead isotopes PY - 2009 DO - https://doi.org/10.1103/PhysRevC.80.024610 SN - 0556-2813 SN - 1089-490X VL - 80 IS - 2 SP - 024610-1 - 024610-12 PB - Inst. CY - Woodbury, NY, USA AN - OPUS4-20698 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen A1 - Yim, Y.-H. A1 - Lee, K.-S. A1 - Goenaga-Infante, H A1 - Malinovskiy, D. A1 - Hill, S. A1 - Ren, T. A1 - Wang, J. A1 - Vocke, R. D. A1 - Murphy, K. E. A1 - Nonose, N. A1 - Rienitz, O. A1 - Noordmann, J. T1 - Certification of ERM-EB400, the first matrix reference material for lead isotope amount ratios, and ERM-AE142, a lead solution providing a lead isotopic composition at the edge of natural variation N2 - Lead isotope amount ratios are commonly used in diverse fields such as archaeometry, geochemistry and forensic science. Currently, five reference materials with certified lead isotope amount ratios are available, namely NIST SRM 981, 982 and 983, GBW-04442 and NMIJ 3681-a. Only NIST SRM 981 and NMIJ 3681-a have approximately natural isotopic compositions, and NIST SRM 981 is predominantly used for correcting mass discrimination/mass fractionation in the applied mass spectrometric procedures. Consequently, there is no other certified reference material available to be used for validation and/or quality control of the analytical procedures applied to lead isotope amount ratio measurements. To fill this gap, two new reference materials have been produced and certified for their lead isotope amount ratios. For both certified reference materials, complete uncertainty budgets have been calculated and SI traceability has been established. This provides the users with independent means for validating and verifying their analytical procedures and for conducting quality control measures. ERM-EB400 is a bronze material with a nominal lead mass fraction of 45 mg kg-1 and certified lead isotope amount ratios of n(206Pb)/n(204Pb) = 18.072(17) mol mol-1, n(207Pb)/n(204Pb) = 15.578(18) mol mol-1 and n(208Pb)/n(204Pb) = 38.075(46) mol mol-1 with the associated expanded uncertainties (k = 2) given in brackets. ERM-AE142 is a high-purity solution of lead in 2% nitric acid with a nominal mass fraction of 100 mg kg-1 and certified Pb isotope amount ratios of n(206Pb)/n(204Pb) = 21.114(17) mol mol-1, n(207Pb)/n(204Pb) = 15.944(17) mol mol-1 and n(208Pb)/n(204Pb) = 39.850(44) mol mol-1 with the associated expanded uncertainties (k = 2) given in brackets. Both materials are specifically designed to fall within the natural lead isotopic variation and to assist users with the validation and verification of their analytical procedures. Note that while one of these reference materials requires the chemical separation of Pb from its matrix (ERM-EB400), the other does not (ERM-AE142). As additional information, δ208/206PbNIST SRM981 values are provided for both materials. For ERM-AE142, a delta value of δ208/206PbNIST SRM981 = -28.21(30) ‰ was obtained, and for ERM-EB400, a delta value of δ208/206PbNIST SRM981 = -129.47(38) ‰ was obtained, with the associated expanded uncertainties (k = 2) given in brackets. KW - Lead isotope variations KW - Radiogenic isotopes KW - Isotope reference material KW - Metrology in chemistry KW - Measurement uncertainty PY - 2019 DO - https://doi.org/10.1111/ggr.12253 SN - 1751-908X SN - 1639-4488 VL - 43 IS - 1 SP - 23 EP - 37 PB - John Wiley & Sons AN - OPUS4-47383 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - RPRT A1 - Andrzejuk, W. A1 - Bau, A. A1 - Charoud-Got, J. A1 - de Vos, P. A1 - Emteborg, H. A1 - Lamberty, A. A1 - Linsinger, T. A1 - Oostra, A. A1 - Quetel, C. A1 - Roebben, G. A1 - Tresl, I. A1 - Hearn, R. A1 - Wood, S. A1 - Pritzkow, W. A1 - Vogl, Jochen T1 - Certification of the sulfur mass fraction in three commercial petrol materials - Certified reference materials ERM-EF211, ERM-EF212 and ERM-EF213 N2 - Three petrol reference materials were certified for their total sulfur content in support of Directive 2003/17/EC of the European Parliament and of the European Council, which stipulates that petrol with a maximum S content of 10 mg/kg must be available in all member states by 2009. Commercially available petrol was obtained and filled into borosilicate ampoules without further treatment. Homogeneity of the materials was tested and no heterogeneity was detected for two of the materials, whereas minor heterogeneity was observed for the third material. Stability of the materials was tested for 8-12 months at 60 °C and no degradation was observed. Characterisation was based on isotope-dilution mass spectrometry (IDMS) applied as primary method of measurement by three European metrology institutes and certified values were assigned using all results. The certified uncertainties include contributions of (potential) heterogeneity, potential degradation as well as characterisation. The final assigned values are: ERM-EF211: 48.8 ± 1.7 mg/kg ERM-EF212: 20.2 ± 1.1 mg/kg ERM-EF213: 9.1 ± 0.8 mg/kg The materials are available from IRMM (ERM-EF211), LGC (ERM-EF212) and BAM (ERMEF213) KW - Sulfur KW - Petrol KW - IDMS PY - 2007 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-355804 SN - 978-92-79-05370-2 SN - 1018-5593 SP - 1 EP - 37 PB - European Communities CY - Luxembourg AN - OPUS4-35580 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Wunderli, S. A1 - Fortunato, G. A1 - Vogl, Jochen A1 - Pritzkow, Wolfgang T1 - Meterology of determination for a new cadmium atomic weight and metrological application to cadmium in rice using MC-ICP-MS KW - Atomic weight KW - Isotope KW - MC-ICP-MS KW - Cadmium KW - Determination of metrology of new cadmium atomic weight KW - Mass spectrometry KW - Review KW - Multicollector inductively coupled plasma mass spectrometry PY - 2003 SN - 1004-2997 VL - 24 IS - 4 SP - 467 EP - 470 CY - Beijing AN - OPUS4-31076 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Geilert, Sonja A1 - Vogl, Jochen A1 - Rosner, M. A1 - Voerkelius, S. A1 - Eichert, T. T1 - Boron isotope fractionation in bell pepper N2 - Various plant compartments of a single bell pepper plant were studied to verify the variability of boron isotope composition in plants and to identify possible intra-plant isotope fractionation. Boron mass fractions varied from 9.8 mg/kg in the fruits to 70.0 mg/kg in the leaves. Boron (B) isotope ratios reported as δ11B ranged from -11.0‰ to +16.0‰ (U ≤ 1.9‰, k=2) and showed a distinct trend to heavier δ11B values the higher the plant compartments were located in the plant. A fractionation of Δ11Bleaf-roots = 27‰ existed in the studied bell pepper plant, which represents about about 1/3 of the overall natural boron isotope variation (ca. 80‰). Two simultaneous operating processes are a possible explanation for the observed systematic intra-plant δ11B variation: 1) B is fixed in cell walls in its tetrahedral form (borate), which preferentially incorporates the light B isotope and the remaining xylem sap gets enriched in the heavy B isotope and 2) certain transporter preferentially transport the trigonal 11B-enriched boric acid molecule and thereby the heavy 11B towards young plant compartments which were situated distal of the roots and typically high in the plant. Consequently, an enrichment of the heavy 11B isotope in the upper young plant parts located at the top of the plant could explain the observed isotope systematic. The identification and understanding of the processes generating systematic intra-plant δ11B variations will potentially enable the use of B isotope for plant metabolism studies. KW - Boron isotopes KW - Bell pepper KW - Boron transport KW - Boron isotope fractionation KW - Intra-plant isotope variability KW - Plants KW - Bell KW - Pepper KW - Isotope fractionation PY - 2015 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-352289 DO - https://doi.org/10.4172/mso.1000101 VL - 1 IS - 1 SP - Article 1000101, 1 EP - 6 AN - OPUS4-35228 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Prohaska, T. A1 - Irrgeher, J. A1 - Benefield, J. A1 - Böhlke, J. K. A1 - Chesson, L. A. A1 - Coplen, T. B. A1 - Ding, T. A1 - Dunn, P. J. H. A1 - Gröning, M. A1 - Holden, N. E. A1 - Meijer, H. A. J. A1 - Moossen, H. A1 - Possolo, A. A1 - Takahashi, Y. A1 - Vogl, Jochen A1 - Walczyk, T. A1 - Wang, J. A1 - Wieser, M. E. A1 - Yoneda, S. A1 - Zhu, X.-K. A1 - Meija, J. T1 - Standard atomic weights of the elements 2021 (IUPAC Technical Report) N2 - Following the reviews of atomic-weight determinations and other cognate data in 2015, 2017, 2019 and 2021, the IUPAC (International Union of Pure and Applied Chemistry) Commission on Isotopic Abundances and Atomic Weights (CIAAW) reports changes of standard atomic weights. The symbol Ar(E) was selected for standard atomic weight of an element to distinguish it from the atomic weight of an element E in a specific substance P, designated Ar(E, P). The CIAAW has changed the values of the standard atomic weights of five elements based on recent determinations of terrestrial isotopic abundances: Ar (argon): from 39.948 ± 0.001 to [39.792, 39.963] Hf (hafnium): from 178.49 ± 0.02 to 178.486 ± 0.006 Ir (iridium): from 192.217 ± 0.003 to 192.217 ± 0.002 Pb (lead): from 207.2 ± 0.1 to [206.14, 207.94] Yb (ytterbium): from 173.054 ± 0.005 to 173.045 ± 0.010 The standard atomic weight of argon and lead have changed to an interval to reflect that the natural variation in isotopic composition exceeds the measurement uncertainty of Ar(Ar) and Ar(Pb) in a specific substance. The standard atomic weights and/or the uncertainties of fourteen elements have been changed based on the Atomic Mass Evaluations 2016 and 2020 accomplished under the auspices of the International Union of Pure and Applied Physics (IUPAP). Ar of Ho, Tb, Tm and Y were changed in 2017 and again updated in 2021: Al (aluminium), 2017: from 26.981 5385 ± 0.000 0007 to 26.981 5384 ± 0.000 0003 Au (gold), 2017: from 196.966 569 ± 0.000 005 to 196.966 570 ± 0.000 004 Co (cobalt), 2017: from 58.933 194 ± 0.000 004 to 58.933 194 ± 0.000 003 F (fluorine), 2021: from 18.998 403 163 ± 0.000 000 006 to 18.998 403 162 ± 0.000 000 005 (Ho (holmium), 2017: from 164.930 33 ± 0.000 02 to 164.930 328 ± 0.000 007) Ho (holmium), 2021: from 164.930 328 ± 0.000 007 to 164.930 329 ± 0.000 005 Mn (manganese), 2017: from 54.938 044 ± 0.000 003 to 54.938 043 ± 0.000 002 Nb (niobium), 2017: from 92.906 37 ± 0.000 02 to 92.906 37 ± 0.000 01 Pa (protactinium), 2017: from 231.035 88 ± 0.000 02 to 231.035 88 ± 0.000 01 Pr (praseodymium), 2017: from 140.907 66 ± 0.000 02 to 140.907 66 ± 0.000 01 Rh (rhodium), 2017: from 102.905 50 ± 0.000 02 to 102.905 49 ± 0.000 02 Sc (scandium), 2021: from 44.955 908 ± 0.000 005 to 44.955 907 ± 0.000 004 (Tb (terbium), 2017: from 158.925 35 ± 0.000 02 to 158.925 354 ± 0.000 008) Tb (terbium), 2021: from 158.925 354 ± 0.000 008 to 158.925 354 ± 0.000 007 (Tm (thulium), 2017: from 168.934 22 ± 0.000 02 to 168.934 218 ± 0.000 006) Tm (thulium), 2021: from 168.934 218 ± 0.000 006 to 168.934 219 ± 0.000 005 (Y (yttrium), 2017: from 88.905 84 ± 0.000 02 to 88.905 84 ± 0.000 01) Y (yttrium), 2021: from 88.905 84 ± 0.000 01 to 88.905 838 ± 0.000 002 KW - Argon KW - Ciaaw.org KW - Hafnium KW - Iridium KW - Lead KW - LSVEC KW - Ytterbium PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-548443 DO - https://doi.org/10.1515/pac-2019-0603 SN - 0033-4545 VL - 94 IS - 5 SP - 573 EP - 600 PB - De Gruyter Verlag CY - Berlin AN - OPUS4-54844 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - GEN A1 - Kazlagić, Anera A1 - Rosner, M. A1 - Cipriani, A. A1 - Frick, D. A. A1 - Glodny, J. A1 - Hoffmann, E. J. A1 - Hora, J. M. A1 - Irrgeher, J. A1 - Lugli, F. A1 - Magna, T. A1 - Meisel, T. C. A1 - Meixner, A. A1 - Possolo, A. A1 - Pramann, A. A1 - Pribil, M. J. A1 - Prohaska, T. A1 - Retzmann, A. A1 - Rienitz, O. A1 - Rutherford, D. A1 - Paula-Santos, G. M. A1 - Tatzel, M. A1 - Widhalm, S. A1 - Willbold, M. A1 - Zuliani, T. A1 - Vogl, Jochen T1 - Data of the characterisation of conventional 87Sr/86Sr isotope ratios in cement, limestone and slate reference materials based on an interlaboratory comparison study N2 - This dataset represents the electronic supplementary material (ESM) of the publication entitled "Characterisation of conventional 87Sr/86Sr isotope ratios in cement, limestone and slate reference materials based on an interlaboratory comparison study", which is published in Geostandards and Geoanalytical Research under the DOI: 10.1111/GGR.12517. It consists of four files. 'ESM_Data.xlsx' contains all reported data of the participants, a description of the applied analytical procedures, basic calculations, the consensus values, and part of the uncertainty assessment. 'ESM_Figure-S1' displays a schematic on how measurements, sequences and replicates are treated for the uncertainty calculation carried out by PTB. 'ESM_Technical-protocol.pdf' is the technical protocol of the interlaboratory comparison, which has been provided to all participants together with the samples and which contains bedside others the definition of the measurand and guidelines for data assessment and calculations. 'ESM_Reporting-template.xlsx' is the Excel template which has been submitted to all participants for reporting their results within the interlaboratory comparison. Excel files with names of the the structure 'GeoReM_Material_Sr8786_Date.xlsx' represent the Rcon(87Sr/86Sr) data for a specific reference material downloaded from GeoReM at the specified date, e.g. 'GeoReM_IAPSO_Sr8786_20221115.xlsx' contains all Rcon(87Sr/86Sr) data for the IAPSO seawater standard listed in GeoReM until 15 November 2022. KW - Reference data KW - Strontium isotope ratio KW - Interlaboratory comparison KW - Reference material KW - Cement KW - Geological material KW - Value assignment KW - Measurement uncertainty KW - Conventional method PY - 2023 UR - https://doi.org/10.5281/zenodo.7804445 DO - https://doi.org/10.5281/zenodo.7804444 PB - Zenodo CY - Geneva AN - OPUS4-57809 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Can, S. Z. A1 - Engin, B. A. A1 - İşleyen, A. A1 - Jotanovic, A. A1 - Acosta, O. A1 - Prina, P. A1 - Schvartz, M. A1 - Savić, M. A1 - Stojanović, M. A1 - Ahumada, D. A. A1 - Abella, J. P. A1 - Näykki, T. A1 - Saro-Aho, T. A1 - Vogl, Jochen A1 - Koenig, Maren A1 - Rienitz, O. A1 - Noordmann, J. A1 - Pape, C. A1 - Towara, J. A1 - Kakoulides, E. A1 - Alexopoulos, C. A1 - Ketrin, R. A1 - Mardika, E. A1 - Komalasari, I. A1 - Elishian, C. A1 - Naujalis, E. A1 - Knašienė, B. A1 - Uribe, C. A1 - Carrasco, E. A1 - Zoń, A. A1 - Warzywoda, B. A1 - Stakheev, A. A1 - Dobrovolskiy, V. A1 - Stolboushkina, T. A1 - Glinkova, A. A1 - Sobina, E. A1 - Tabatchikova, T. A1 - Gažević, L. A1 - Paunovic, M. A1 - Jaćimović, R. A1 - Zuliani, T. A1 - Zambra, R. P. A1 - Napoli, R. T1 - Determination of elements in river water N2 - The need for quality assessment of anthropogenic impact on environmental pollution is increasing due to discharge from various industries, the use of chemicals in agriculture and the consumption of fossil fuels. Diminishing resources such as natural waters used for the cultivation of agricultural products, plant and animal habitats are under severe pollution pressure and are at constant risk. Several parameters, such as Pb, Cd, Ni, Hg were listed by Water Framework Directive in Directive(2008/105/EC) in the priority substances. Cadmium and Hg were identified as priority hazardous substances whereas As is an important contaminant for its potential toxicological and carcinogenic effects. An inter-comparison study is organised in EURAMET TC-MC in order to demonstrate the capability participants for measuring five elements in river water. The participants carried out measurements for analytes: Pb, Cd, Ni and As as mandatory elements, and Se as an optional one. Participants were asked to perform the measurements with respect to the protocol provided. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA). KW - Metrology KW - Traceability KW - Toxic elements KW - River water PY - 2023 DO - https://doi.org/10.1088/0026-1394/60/1A/08001 VL - 60 IS - 1a SP - 1 EP - 40 PB - BIPM & IOP Publishing Ltd AN - OPUS4-56786 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Kazlagić, Anera A1 - Rosner, M. A1 - Cipriani, A. A1 - Frick, D. A. A1 - Glodny, J. A1 - Hoffmann, E. J. A1 - Hora, J. M. A1 - Irrgeher, J. A1 - Lugli, F. A1 - Magna, T. A1 - Meisel, T. C. A1 - Meixner, A. A1 - Possolo, A. A1 - Pramann, A. A1 - Pribil, M. J. A1 - Prohaska, T. A1 - Retzmann, Anika A1 - Rienitz, O. A1 - Rutherford, D. A1 - Paula-Santos, G. M. A1 - Tatzel, M. A1 - Widhalm, S. A1 - Willbold, M. A1 - Zuliani, T. A1 - Vogl, Jochen T1 - Characterisation of conventional 87Sr/86Sr isotope ratios in cement, limestone and slate reference materials based on an interlaboratory comparison study N2 - An interlaboratory comparison (ILC)was organised to characterise 87Sr/86Sr isotope ratios in geological and industrial reference materials by applying the so-called conventional method for determining 87Sr/86Sr isotope ratios. Four cements (VDZ 100a,VDZ 200a, VDZ 300a, IAG OPC-1), one limestone (IAG CGL ML-3) and one slate (IAG OU-6) reference materials were selected, covering a wide range of naturally occurring Sr isotopic signatures. Thirteen laboratories received aliquots of these six reference materials together with a detailed technical protocol. The consensus values for the six reference materials and their associated measurement uncertainties were obtained by applying a Gaussian, linear mixed effects model fitted to all the measurement results. By combining the consensus values and their uncertainties with an uncertainty contribution for potential heterogeneity, reference values ranging from 0.708134 mol mol-1 to 0.729778 mol mol-1 were obtained with relative expanded uncertainties of ≤ 0.007 %. This study represents an ILC on conventional 87Sr/86Sr isotope ratios, within which metrological principles were considered and the compatibility of measurement results obtained by MC-ICP-MS and by MC-TIMS is demonstrated. The materials characterised in this study can be used as reference materials for validation and quality control purposes and to estimate measurement uncertainties in conventional 87Sr/86Sr isotope ratio measurement. KW - Sr isotope analysis KW - Isotope ratios KW - Cement KW - Geological material KW - MC-TIMS KW - MC-ICP-MS KW - Interlaboratory comparison KW - Measurement uncertainty KW - Cconventional method PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-579836 DO - https://doi.org/10.1111/ggr.12517 SN - 1639-4488 VL - 47 IS - 4 SP - 821 EP - 840 PB - Wiley online library AN - OPUS4-57983 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Das, P. A1 - Chakraborty, Gouri A1 - Roeser, J. A1 - Vogl, S. A1 - Rabeah, J. A1 - Thomas, A. T1 - Integrating Bifunctionality and Chemical Stability in Covalent Organic Frameworks via One-Pot Multicomponent Reactions for Solar-Driven H2O2 Production N2 - Multicomponent reactions (MCRs) can be used to introduce different functionalities into highly stable covalent organic frameworks (COFs). In this work, the irreversible three-component Doebner reaction is utilized to synthesize four chemically stable quinoline-4-carboxylic acid DMCR-COFs (DMCR-1−3 and DMCR-1NH) equipped with an acid−base bifunctionality. These DMCR-COFs show superior photocatalytic H2O2 evolution (one of the most important industrial oxidants) compared to the imine COF analogue (Imine-1). This is achieved with sacrificial oxidants but also in pure water and under an oxygen or air atmosphere. Furthermore, the DMCR-COFs show high photostability, durability, and recyclability. MCR-COFs thus provide a viable materials’ platform for solar to chemical energy conversion. KW - Reactions KW - Bifunctionality KW - Postsynthetic modification KW - Multicomponent PY - 2022 DO - https://doi.org/10.1021/jacs.2c11454 SN - 0002-7863 VL - 145 IS - 5 SP - 2975 EP - 2984 PB - ACS Publications AN - OPUS4-56922 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Breuninger, Esther S. A1 - Tolu, Julie A1 - Aemisegger, Franziska A1 - Thurnherr, Iris A1 - Bouchet, Sylvain A1 - Mestrot, Adrien A1 - Ossola, Rachele A1 - McNeill, Kristopher A1 - Tukhmetova, Dariya A1 - Vogl, Jochen A1 - Meermann, Björn A1 - Sonke, Jeroen E. A1 - Winkel, Lenny H. E. T1 - Marine and terrestrial contributions to atmospheric deposition fluxes of methylated arsenic species N2 - Arsenic, a toxic element from both anthropogenic and natural sources, reaches surface environments through atmospheric cycling and dry and wet deposition. Biomethylation volatilizes arsenic into the atmosphere and deposition cycles it back to the surface, affecting soil-plant systems. Chemical speciation of deposited arsenic is important for understanding further processing in soils and bioavailability. However, the range of atmospheric transport and source signature of arsenic species remain understudied. Here we report significant levels ofmethylated arsenic in precipitation, cloudwater and aerosols collected under free tropospheric conditions at Pic du Midi Observatory (France) indicating long-range transport, which is crucial for atmospheric budgets. Through chemical analyses andmoisture source diagnostics, we identify terrestrial and marine sources for distinct arsenic species. Estimated atmospheric deposition fluxes of methylated arsenic are similar to reported methylation rates in soils, highlighting atmospheric deposition as a significant, overlooked source of potentially bioavailable methylated arsenic species impacting plant uptake in soils. KW - Arsenic KW - Speciation KW - Fluxes KW - Marine KW - Terrestrial KW - Sulfur isotopes PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-615551 DO - https://doi.org/10.1038/s41467-024-53974-z SN - 2041-1723 VL - 15 IS - 1 SP - 1 EP - 13 PB - Springer Science and Business Media LLC AN - OPUS4-61555 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CHAP A1 - Bürger, S. A1 - Vogl, Jochen A1 - Kloetzli, U. A1 - Nunes, L. A1 - Lavelle, M. ED - Prohaska, T. ED - Irrgeher, J. ED - Zitek, A. ED - Jakubowski, Norbert T1 - Thermal ionisation mass spectrometry N2 - Thermal ionisation, also known as surface ionisation, was one of the first ionisation techniques developed for mass spectrometry, having been invented as early as 1918. The ionization process, described by the empirically derived “fractionation laws”, is widely understood. Isotope ratio measurements can be achieved with high precision and accuracy. Due to this, thermal ionisation has paved the way for great scientific achievements including: the discovery of new isotopes, the determination of radioactive half-lives and atomic weights of the elements, the accurate determination of the age of the earth and investigations on human society in the past such as mobility and trade. TIMS is still regarded as ‘golden standard’ in isotope ratio measurements. Thus the method is a reference technique that remains at the forefront of isotopic analysis particularly in the fields of metrology. A concise overview is given here of the technical background of thermal ionisation as well as the numerous applications of this technique in earth sciences, industry, metrology, and nuclear forensics. PY - 2015 SN - 978-1-84973-392-2 SN - 978-1-84973-540-7 DO - https://doi.org/10.1039/9781849735407-00381 SN - 2044-253X N1 - Serientitel: New developments in mass spectrometry – Series title: New developments in mass spectrometry VL - 3 SP - Chapter 14, 381 EP - 438 CY - Cambridge, UK AN - OPUS4-32557 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen A1 - Kipphardt, Heinrich A1 - del Rocío Arvizu Torres, M. A1 - Manzano, J.V.L. A1 - Rodrigues, J.M. A1 - de Sena, R.C. A1 - Yim, Y.-H. A1 - Heo, S.W. A1 - Zhou, T. A1 - Turk, G.C. A1 - Winchester, M. A1 - Yu, L. L. A1 - Miura, T. A1 - Methven, B. A1 - Sturgeon, R. A1 - Jährling, R. A1 - Rienitz, O. A1 - Tunc, M. A1 - Can, S.Z. T1 - Final report of the key comparison CCQM-K72: Purity of zinc with respect to six defined metallic analytes N2 - KEY COMPARISON High purity elements can serve as a realization of the SI unit amount of substance for the specific element. Solutions prepared from high purity metals by applying gravimetric preparation and the concept of molar mass are used as 'calibration' solutions in many fields of analytical chemistry and provide the metrological basis in elemental analysis. Since ideal purity does not exist for real materials, the actual purity of the high purity material must be known with a specified uncertainty. As required uncertainties around 10-4 relative on the purity statement are not accessible in almost all cases by a direct measurement of the element in itself, the indirect approach is followed, where all elements excepting the matrix element itself are measured and their sum is subtracted from the value for ideal purity, which is 1 kg/kg. It was the aim of this comparison to demonstrate the capability of national metrology institutes and designated institutes to determine the purity of pure elements. In order to limit the effort within this comparison, only six metallic impurities (Ag, Al, Cd, Cr, Ni, Tl) in the low mg/kg range are considered in a zinc matrix. It has to be underlined here that the task was to measure the purity of zinc based on the determination of six analytes. The task is not trace analysis of specific analytes in zinc. This subtle distinction defines different measurands. The sample, pure Zn, was cut in pieces of cubic geometry for wet chemical analysis or of pin geometry for GDMS analysis and was sent to the participants. The comparison was run under the auspices of the Inorganic analysis Working Group (IAWG) of the CCQM and was piloted by the BAM Federal Institute for Materials Research and Testing, Berlin, Germany. The majority of the participants applied ICP-MS techniques and only two participants used additionally atomic absorption spectrometry. GDMS was used only by one participant. The observed spreads for the measurement results reported by the participants were significantly lower than those of the preceding study CCQM-P107 and were well below the target uncertainty of 30% relative. As a consequence, comparability within the participating laboratories is demonstrated to be established. The individual measurement results, mean values and medians derived were in all cases very consistent with the reference values obtained by IDMS and so the accuracy of the measurement results for the participating laboratories is as well demonstrated to be established. Especially with the results of CCQM-P62 and CCQM-P107 in mind, the outcome of CCQM-K72 can be considered as a big step forward in the community. CCQM is aware of the difference between a characterization based on only six analytes and a complete characterization. Therefore, the pilot study CCQM-P149 has been initiated and already started, which focuses on the fit-for-purpose approaches for the purity determination of metals (here: zinc) to be used as primary standards in elemental analysis. Another follow-up in the form of a pilot study on non-metal impurities is mandatory, because non-metal impurities such as oxygen, nitrogen and sulfur often make up the largest contributions. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA). KW - CCQM KW - Metrology KW - Purity PY - 2014 DO - https://doi.org/10.1088/0026-1394/51/1A/08008 SN - 0026-1394 SN - 1681-7575 VL - 51 IS - 1A (Tech.Suppl. 2014) SP - 08008, 1 EP - 40 PB - Inst. of Physics Publ. CY - Bristol AN - OPUS4-30772 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Labarraque, G. A1 - Oster, C. A1 - Fisicaro, P. A1 - Meyer, Christian A1 - Vogl, Jochen A1 - Noordmann, J. A1 - Rienitz, O. A1 - Riccobono, F. A1 - Donet, S. T1 - Reference measurement procedures for the quatification of platinum-group elements (PGEs) from automotive exhaust emissions N2 - The major source of the anthropogenic platinum group element (PGE) emission is attributed to the use of catalytic converters in automobiles. This paper describes the work performed by three National Metrology Institutes (Laboratoire national de métrologie et d’essais, by the Physikalisch-technische bundesanstalt, Bundesanatalt für materialforschung und prûfung), in the framework of the Joint Research Project 'PartEmission' under the European Metrology Research Program. An analytical procedure based on a cationic exchange protocol and the isotope dilution or standard addition using an Inductived Coupled Plasma Mass Spectrometer, ICP-MS, for the quantification of the elements Pt, Pd and Rh from automotive exhaust emissions is described. Results obtained on a road dust certified reference (BCR 723) material showed a good agreement with the certified values, at ng/g levels, and relative expanded uncertainties within the range of 7–10%. Analysis of filters impacted with automotive exhaust particle emissions (from a diesel engine) showed the amount of collected PGE at levels of 10–1000 pg/filter. Their quantifications followed the developed analytical protocol that had been carried out with relative expanded uncertainties in the range of a few per cent up to 20% per filter. Nevertheless, a lack of homogeneity between the filters was observed, making the comparison between the project partners difficult in the sake of the validation of their analytical procedures on real samples. KW - Automotive pollution KW - Catalytic converters KW - Platinum group elements (PGE) KW - ICP-MS KW - Isobaric interferences KW - Cationic exchange KW - Reference measurement procedures KW - Double isotope dilution KW - Palladium KW - Platinum KW - Automotive exhaust emissions KW - IDMS PY - 2015 DO - https://doi.org/10.1080/03067319.2015.1058931 SN - 0306-7319 SN - 1029-0397 SN - 0092-9085 VL - 95 IS - 9 SP - 777 EP - 789 PB - Gordon and Breach CY - New York, NY AN - OPUS4-34015 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Rosner, Martin A1 - Pritzkow, Wolfgang A1 - Vogl, Jochen A1 - Voerkelius, S. T1 - Boron isotope compositions of crop plants: A new tracer for the origin and authenticity of food T2 - TRACE 5th Annual Meeting and Conference CY - Freising, Germany DA - 2009-04-01 KW - Boron isotopes in food KW - Provenance studies KW - Reference materials KW - Analytical uncertainty PY - 2009 IS - Session A / Part I SP - 1 EP - 2 AN - OPUS4-20293 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rosner, Martin A1 - Pritzkow, Wolfgang A1 - Vogl, Jochen A1 - Voerkelius, S. T1 - Development and validation of a method to determine the boron isotopic composition of crop plants N2 - We present a comprehensive chemical and mass spectrometric method to determine boron isotopic compositions of plant tissue. The method including dry ashing, a three-step ion chromatographic boron–matrix separation, and 11B/10B isotope ratio determinations using the Cs2BO2+ graphite technique has been validated using certified reference and quality control materials. The developed method is capable to determine δ11B values in plant tissue down to boron concentrations of 1 mg/kg with an expanded uncertainty of ≤1.7‰ (k = 2). The determined δ11B values reveal an enormous isotopic range of boron in plant tissues covering three-quarters of the natural terrestrial occurring variation in the boron isotopic composition. As the local environment and anthropogenic activity mainly control the boron intake of plants, the boron isotopic composition of plants can be used for food provenance studies. KW - Boron isotope variations KW - Delta-scale KW - Stable isotopes KW - Isotope reference materials KW - Provenance KW - Authenticity PY - 2011 DO - https://doi.org/10.1021/ac102836h SN - 0003-2700 SN - 1520-6882 VL - 83 IS - 7 SP - 2562 EP - 2568 PB - American Chemical Society CY - Washington, DC AN - OPUS4-23495 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Quétel, C. A1 - Vogl, Jochen A1 - Prohaska, T. A1 - Nelms, S. A1 - Taylor, P.D.P. A1 - De Bièvre, P. T1 - Comparative performance study of ICP mass spectrometers by means of U "isotopic measurements" N2 - The performance of four commercially available ICPMS instruments of three different types was compared by means of uranium "isotopic measurements". Examined were two quadrupole sector (different generation, different manufacturer), one single detector double focusing magnetic sector and one multiple collector double focusing magnetic sector instruments. The same samples of the IRMM-072 series were used under routine conditions to measure the 233U/235U and the 233U/238U ratios which, in these samples, vary over almost three orders of magnitude from ~ 1 to ~ 2 · 10-3. Within expanded (k = 2) uncertainties, good agreement was observed between the certified values and the data internally corrected for mass-discrimination effects. The magnitude of the evaluated uncertainties was different for each type of instrument. With the multiple collector instrument, expanded uncertainties varied from - 0.04% to- 0.24% for the 233U/235U ratio, and from - 0.08% to - 0.27% for the 233U/238U ratio. They were ~ 1 to 5 times larger with the single detector magnetic sector instrument, and ~ 10 to 25 times larger with both quadrupole sector instruments. With the multiple collector instrument, repeatability of the measurements seemed to be limited by the difficulty of correcting properly for instrumental background, whereas with the single detector magnetic sector instrument the counting statistics was the only limitation (on smallest ratios). Apparent mass-discrimination was clearly found to be larger but more reproducible (and hence easier to correct for) in the case of magnetic sector instruments than for both quadrupole sector instruments. If space charge effects were the main source of mass-discrimination for all instruments, these results are in contradiction with the hypothesis of the size of mass-discrimination decreasing with the acceleration voltage. With the single detector magnetic sector instrument in particular (when operated by changing the ion energy only), our results pointed at more than only one major source of mass-discrimination, with variable size depending on the ratios measured. KW - ICP-MS KW - Isotope ratio KW - Uranium PY - 2000 DO - https://doi.org/10.1007/s002160000499 SN - 0937-0633 VL - 368 SP - 148 EP - 155 PB - Springer CY - Berlin AN - OPUS4-7218 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Nelms, S. A1 - Quétel, C. A1 - Prohaska, T. A1 - Vogl, Jochen A1 - Taylor, P.D.P. T1 - Evaluation of detector dead time calculation models for ICP-MS KW - ICP-MS KW - Dead time KW - Uncertainty PY - 2001 DO - https://doi.org/10.1039/b007913h SN - 0267-9477 SN - 1364-5544 VL - 16 IS - 4 SP - 333 EP - 338 PB - Royal Society of Chemistry CY - London AN - OPUS4-7219 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Pritzkow, Wolfgang A1 - Wunderli, S. A1 - Vogl, Jochen A1 - Fortunato, G. T1 - The isotope abundances and the atomic weight of cadmium by a metrological approach N2 - Gravimetric synthetic mixtures of seven enriched, purified cadmium isotope materials were used to determine the correction factors for mass fractionation (MC-TIMS) and mass discrimination (MC-ICP-MS). The isotope abundance ratios determined for various natural cadmium materials represents the isotope composition for natural cadmium materials. Correction of the isotope abundance ratios observed yielded isotope abundances of 106Cd 0.012450(8), 108Cd 0.008884(4), 110Cd 0.124846(16), 111Cd 0.127955(14), 112Cd 0.241110(38), 113Cd 0.122254(22), 114Cd 0.287439(60) and 116Cd 0.075183(32). The newly determined atomic weight of natural cadmium based on SI-traceable evaluation of the isotope abundance ratios is 112.41384(18). The cadmium material designated in the paper as Cd-2211 can be used as an isotope reference material with a δ(114Cd/110Cd)-value of 0‰. The results obtained show that the uncertainties for the isotope abundances and the atomic weight given as IUPAC values for cadmium are overestimated. KW - Atomic weight KW - Isotope abundance KW - Isotopic composition KW - Cadmium PY - 2007 SN - 0168-1176 SN - 0020-7381 SN - 1387-3806 SN - 1873-2798 VL - 261 IS - 1 SP - 74 EP - 85 PB - Elsevier CY - Amsterdam AN - OPUS4-14543 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Linsinger, T. A1 - Andrzejuk, W. A1 - Bau, A. A1 - Charoud-Got, J. A1 - De VOs, P. A1 - Emteborg, H. A1 - Hearn, R. A1 - Lamberty, A. A1 - Oostra, A. A1 - Pritzkow, Wolfgang A1 - Quétel, C. A1 - Roebben, G. A1 - Tresl, I. A1 - Vogl, Jochen A1 - Wood, S. T1 - Production of three certified reference materials for the sulfur content in gasoline (petrol) N2 - Directive 2003/17/EC of the European Parliament and the European Council stipulates that petrol (gasoline) with a total sulfur content below 10 mg kg-1 must be available in all European Union member states by 2009. Three certified reference materials were produced in support of this directive in a joint effort of the members of the European Reference Materials Initiative (ERM). Two of the materials were made from commercial petrol, while the third one was prepared from a blend of commercial petrols. Relative between-ampule heterogeneity of the materials was quantified and found to be below 2.5%. Potential degradation during storage and dispatch was quantified, and shelf lives based on these values were set. The three materials were characterized by three institutes using different variants of isotope-dilution mass spectrometry. The results from the three institutes were combined, and the final uncertainties of the respective sulfur mass fractions were estimated including contributions from heterogeneity, stability, and characterization. The following mass fractions were derived: ERM-EF211, 48.8 ± 1.7 mg kg-1; ERM-EF212, 20.2 ± 1.1 mg kg-1; and ERM-EF213, 9.1 ± 0.8 mg kg-1. KW - CRM KW - IDMS KW - ERM PY - 2007 DO - https://doi.org/10.1021/ef070155t SN - 0887-0624 SN - 1520-5029 VL - 21 IS - 4 SP - 2240 EP - 2244 PB - ACS Publ. CY - Washington, DC AN - OPUS4-16371 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - GEN A1 - Paz, B. A1 - Meyer, H.-P. A1 - Varychev, A. A1 - Rosner, M. A1 - Vogl, Jochen A1 - Behrendt, S. ED - Treister, M. ED - Yablonsky, L. T1 - Археометрические исследования стеклянных сосудов из погребений ранних кочевников Южного Приуралья T1 - Archäometrische Untersuchungen an achaimenidischen Glasgefäßen aus den Gräbern der frühen Nomaden des südlichen Urals N2 - Die untersuchten Gläser lassen sich in drei Gruppen einteilen. Die beiden Fragmente aus der Nekropole von Pjatimary deuten auf eine frühe Herstellungszeit hin und lassen sich mit eisenzeitlichen Fragmenten iranischer Herkunft vergleichen. Die Glasfunde aus der Nekropole Filippovka bestehen aus völlig unterschiedlichen Glasmaterialien und müssen einer späteren Herstellungszeit zugeordnet werden. Sowohl die unterschiedlichen Rezepturen als auch der definitive Hinweis auf verschiedene Rohstoffquellen zur Herstellung der untersuchten Gläser, lassen ein breitgefächertes Netz von weitreichenden Handelsbeziehungen erkennen. KW - Archäometrie KW - Isotopenverhältnisse Strontium und Blei KW - Elementanalyse KW - Massenspektrometrie KW - Herkunft PY - 2012 SN - 978-5-903011-85-8 VL - 1 SP - 262 EP - 267 PB - TAUS CY - Moskau AN - OPUS4-26899 LA - rus AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rienitz, O. A1 - Schiel, D. A1 - Görlitz, V. A1 - Jährling, R. A1 - Vogl, Jochen A1 - Lara-Manzano, J.V. A1 - Zon, A. A1 - Fung, W.-H. A1 - Buzoianu, M. A1 - De Sena, R.C. A1 - Dos Reis, L.A. A1 - Valiente, L. A1 - Yim, Y.-H. A1 - Hill, S. A1 - Champion, R. A1 - Fisicaro, P. A1 - Bing, W. A1 - Turk, G.C. A1 - Winchester, M. R. A1 - Saxby, D. A1 - Merrick, J. A1 - Hioki, A. A1 - Miura, T. A1 - Suzuki, T. A1 - Linsky, M. A1 - Barzev, A. A1 - Máriássy, M. A1 - Cankur, O. A1 - Ari, B. A1 - Tunc, M. A1 - Konopelko, L.A. A1 - Kustikov, Y.A. A1 - Bezruchko, M. T1 - Final report on CCQM-K87: Mono-elemental calibration solutions N2 - The aim of this comparison was to demonstrate the capability of national metrology institutes to measure elemental mass fractions at a level of w(E) ≈ 1 g/kg as found in almost all mono-elemental calibration solutions. These calibration solutions represent an important link in traceability systems in inorganic analysis. Virtually all traceable routine measurements are linked to the SI through these calibration solutions. Every participant was provided with three solutions of each of the three selected elements chromium, cobalt and lead. This comparison was a joint activity of the Inorganic Analysis Working Group (IAWG) and the Electrochemical Analysis Working Group (EAWG) of the CCQM and was piloted by the Physikalisch-Technische Bundesanstalt (PTB, Braunschweig, Germany) with the help of the Bundesanstalt für Materialforschung und -prüfung (BAM, Berlin, Germany), the Centro Nacional de Metrología (CENAM, Querétaro, Mexico) and the National Institute of Standards and Technology (NIST, Gaithersburg, USA). A small majority of participants applied inductively coupled plasma optical emission spectrometry (ICP OES) in combination with a variety of calibration strategies (one-point-calibration, bracketing, calibration curve, each with and without an internal standard). But also IDMS techniques were carried out on quadrupole, high resolution and multicollector ICP-MS machines as well as a TIMS machine. Several participants applied titrimetry. FAAS as well as ICP-MS combined with non-IDMS calibration strategies were used by at least one participant. The key comparison reference values (KCRV) were agreed upon during the IAWG/EAWG meeting in November 2011 held in Sydney as the added element content calculated from the gravimetric sample preparation. Accordingly the degrees of equivalence were calculated. Despite the large variety of methods applied no superior method could be identified. The relative deviation of the median of the participants' results from the gravimetric reference value was equal or smaller than 0.1% (with an average of 0.05%) in the case of all three elements. KW - CCQM KW - Metrology KW - IDMS PY - 2012 DO - https://doi.org/10.1088/0026-1394/49/1A/08010 SN - 0026-1394 SN - 1681-7575 VL - 49 IS - 08010, 1A (Technical Supplement 2012) SP - 1 EP - 104 PB - Inst. of Physics Publ. CY - Bristol AN - OPUS4-30458 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Abad Andrade, Carlos Enrique A1 - Florek, S. A1 - Becker-Ross, H. A1 - Huang, M.-D. A1 - Heinrich, Hans-Joachim A1 - Recknagel, Sebastian A1 - Vogl, Jochen A1 - Jakubowski, Norbert A1 - Panne, Ulrich T1 - Determination of boron isotope ratios by high-resolution continuum source molecular absorption spectrometry using graphite furnace vaporizers N2 - Boron isotope amount ratios n(10B)/n(11B) have been determined by monitoring the absorption spectrum of boron monohydride (BH) in a graphite furnace using high-resolution continuum source molecular absorption spectrometry (HR-CS-MAS). Bands (0→0) and (1→1) for the electronic transition X1Σ+ → A1Π were evaluated around wavelengths 433.1 nm and 437.1 nm respectively. Clean and free of memory effect molecular spectra of BH were recorded. In order to eliminate the memory effect of boron, a combination of 2% (v/v) hydrogen gas in argon and 1% trifluoromethane in argon, an acid solution of calcium chloride and mannitol as chemical modifiers was used. Partial least square regression (PLS) for analysis of samples and reference materials were applied. For this, a spectral library with different isotopes ratios for PLS regression was built. Results obtained around the 433.1 nm and 437.1 nm spectral regions are metrologically compatible with those reported by mass spectrometric methods. Moreover, for the evaluated region of 437 nm, an accuracy of 0.15‰ is obtained as the average deviation from the isotope reference materials. Expanded uncertainties with a coverage factor of k = 2 range between 0.15 and 0.44‰. This accuracy and precision are compatible with those obtained by mass spectrometry for boron isotope ratio measurements. KW - Boron isotopes KW - Isotope ratios KW - Boron monohydride KW - Molecular absorption KW - High-resolution continuum source absorption spectrometry KW - Graphite furnace KW - Memory effect KW - HR-CS-MAS PY - 2017 UR - http://www.sciencedirect.com/science/article/pii/S0584854717302537 DO - https://doi.org/10.1016/j.sab.2017.08.012 SN - 0584-8547 VL - 136 SP - 116 EP - 122 PB - Elsevier CY - Amsterdam, The Netherlands AN - OPUS4-42071 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen A1 - Kipphardt, Heinrich A1 - Richter, Silke A1 - Bremser, Wolfram A1 - Arvizu Torres, M. A1 - Lara Manzano, J. A1 - Buzoianu, M. A1 - Hill, S. A1 - Petrov, P. A1 - Goenaga-Infante, H. A1 - Sargent, M. A1 - Fisicaro, P. A1 - Labarraque, G. A1 - Zhou, T. A1 - Turk, G. C. A1 - Winchester, M. A1 - Miura, T. A1 - Methven, B. A1 - Sturgeon, R. A1 - Jährling, R. A1 - Rienitz, O. A1 - Mariassy, M. A1 - Hankova, Z. A1 - Sobina, E. A1 - Krylov, A. I. A1 - Kustikov, Y. A. A1 - Smirnov, V. V. T1 - Establishing comparability and compatibility in the purity assessment of high purity zinc as demonstrated by the CCQM-P149 intercomparison N2 - For the first time, an international comparison was conducted on the determination of the purity of a high purity element. Participants were free to choose any analytical approach appropriate for their institute’s applications and services. The material tested was a high purity zinc, which had earlier been assessed for homogeneity and previously used in CCQM-K72 for the determination of six defined metallic impurities. Either a direct metal assay of the Zn mass fraction was undertaken by EDTA titrimetry, or an indirect approach was used wherein all impurities, or at least the major ones, were determined and their sum subtracted from ideal purity of 100 %, or 1 kg/kg. Impurity assessment techniques included glow discharge mass spectrometry, inductively coupled plasma mass spectrometry and carrier gas hot extraction/combustion analysis. Up to 91 elemental impurities covering metals, non-metals and semi-metals/metalloids were quantified. Due to the lack of internal experience or experimental capabilities, some participants contracted external laboratories for specific analytical tasks, mainly for the analysis of non-metals. The reported purity, expressed as zinc mass fraction in the high purity zinc material, showed excellent agreement for all participants, with a relative standard deviation of 0.011 %. The calculated reference value, w(Zn) = 0.999 873 kg/kg, was assigned an asymmetric combined uncertainty of + 0.000025 kg/kg and – 0.000028 kg/kg. Comparability amongst participating metrology institutes is thus demonstrated for the purity determination of high purity metals which have no particular difficulties with their decomposition / dissolution process when solution-based analytical methods are used, or which do not have specific difficulties when direct analysis approaches are used. Nevertheless, further development is required in terms of uncertainty assessment, quantification of non-metals and the determination of purity of less pure elements and/or for those elements suffering difficulties with the decomposition process. KW - Purity assessment KW - Direct metal assay KW - Impurity assessment KW - Non-metal analysis KW - High-purity elements KW - SI-traceability PY - 2018 DO - https://doi.org/10.1088/1681-7575/aaa677 SN - 1681-7575 SN - 0026-1394 VL - 55 IS - 2 SP - 211 EP - 221 PB - Institute of Physics Publishing CY - Bristol AN - OPUS4-44257 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Winckelmann, Alexander A1 - Nowak, S. A1 - Richter, Silke A1 - Recknagel, Sebastian A1 - Riedel, Jens A1 - Vogl, Jochen A1 - Panne, Ulrich A1 - Abad Andrade, Carlos Enrique T1 - High-Resolution Atomic Absorption Spectrometry Combined With Machine Learning Data Processing for Isotope Amount Ratio Analysis of Lithium N2 - An alternative method for lithium isotope amount ratio analysis based on a combination of high-resolution atomic absorption spectrometry and spectral data analysis by machine learning (ML) is proposed herein. It is based on the well-known isotope shift of approximately 15 pm for the electronic transition 22P←22S at around the wavelength of 670.8 nm, which can be measured by the state-of-the-art high-resolution continuum source graphite furnace atomic absorption spectrometry. For isotope amount ratio analysis, a scalable tree boosting ML algorithm (XGBoost) was employed and calibrated using a set of samples with 6Li isotope amount fractions, ranging from 0.06 to 0.99 mol mol–1, previously determined by a multicollector inductively coupled plasma mass spectrometer (MC-ICP-MS). The calibration ML model was validated with two certified reference materials (LSVEC and IRMM-016). The procedure was applied toward the isotope amount ratio determination of a set of stock chemicals (Li2CO3, LiNO3, LiCl, and LiOH) and a BAM candidate reference material NMC111 (LiNi1/3Mn1/3Co1/3O2), a Li-battery cathode material. The results of these determinations were compared with those obtained by MC-ICP-MS and found to be metrologically comparable and compatible. The residual bias was −1.8‰, and the precision obtained ranged from 1.9 to 6.2‰. This precision was sufficient to resolve naturally occurring variations, as demonstrated for samples ranging from approximately −3 to +15‰. To assess its suitability to technical applications, the NMC111 cathode candidate reference material was analyzed using high-resolution continuum source atomic absorption spectrometry with and without matrix purification. The results obtained were metrologically compatible with each other. KW - Lithium KW - Isotope KW - Machine learning KW - Algorithms KW - Reference material KW - AAS KW - Atomic Absorption Spectrometry PY - 2021 DO - https://doi.org/10.1021/acs.analchem.1c00206 SN - 1520-6882 VL - 93 IS - 29 SP - 10022 EP - 10030 PB - American Chemical Society CY - Washington, DC, USA AN - OPUS4-53028 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - RPRT A1 - Vogl, Jochen A1 - Yim, Y.-H. A1 - Lee, K.-S. A1 - Goenaga-Infante, H. A1 - Malinovskiy, D. A1 - Vocke, R. D. A1 - Murphy, K. A1 - Nonose, N. A1 - Rienitz, O. A1 - Noordmann, J. A1 - Näykki, T. A1 - Sara-Aho, T. A1 - Ari, B. A1 - Cankur, O. T1 - CCQM-P134 Pb isotope amount ratios and delta-values in bronze N2 - Isotope amount ratios (hereafter referred to as simply isotope ratios) are proving useful in an ever increasing array of applications that range from studies unravelling transport processes, to pinpointing the provenance of specific samples as well as trace element quantification by using isotope dilution mass spectrometry (IDMS). These expanding applications encompass fields as diverse as archaeology, food chemistry, forensic science, geochemistry, medicine and metrology. However, to be effective tools, the isotope ratio data must be reliable and traceable to enable the comparability of measurement. The importance of traceability and comparability in isotope ratio analysis has already been recognized by the Inorganic Analysis Working Group (IAWG) within the CCQM. Three pilot studies have focused on the quality of isotope ratio determinations (P48 “U isotope ratios in urine”, P75 “stable isotopes in Methionine”, P105 “87Sr/86Sr in wine”). Moreover, isotope ratio measurements are fundamental to IDMS amount of substance determinations. For example, when Pb quantification using IDMS is undertaken, this requires the measurements of Pb isotope ratios. While the requirements for isotope ratio accuracy and precision in the case of of IDMS are generally quite modest, “absolute” Pb isotope ratio measurements for geochemical age dating and source rock characterization as well as forensic provenance and fingerprinting studies require Pb isotope ratio measurements of the highest quality. To support present and future CMCs on isotope ratio determinations, a Key Comparison was urgently needed. Therefore, it was decided at the IAWG meeting in Paris in April 2011 that a Key Comparison on the determination of Pb isotope ratios in a pure Pb solution and in a bronze sample should be organized and accompanied by a pilot study. Measuring Pb isotope amount ratios in a pure Pb solution, while seemingly straight forward, rigorously tests the ability of analyst to correct for any instrumental effects (such as mass discrimination and blank correction) on the measured ratios. Pb, present in trace amounts in a metal matrix sample (e.g. Pb in bronze), provides a real world test of the whole chemical and instrumental procedure, from chemical separation and sample purification to analysis and subsequent correction of appropriate instrumental effects on the separated samples. A suitable bronze material with a Pb mass fraction between 10 and 100 mg·kg-1 was available at BAM. A high purity solution of Pb with a mass fraction of approximately 100 mg·kg-1 was also available. By comparing the Pb isotope ratio results obtained for the bronze sample with the Pb isotope ratio results from the Pb solution, potential biases arising from the processing of the bronze sample could be effectively identified and separated from the instrumental effects arising from the measurement and data processing protocol. KW - Isotope ratio KW - Delta value KW - Molar mass KW - Measurement uncertainty KW - Traceability PY - 2017 UR - https://www.bipm.org/wg/CCQM/IAWG/Allowed/IAWG_Pilot_Studies/CCQM-P134.pdf SP - 1 EP - 42 PB - BIPM (Bureau International des Poids et Mesures) CY - Paris AN - OPUS4-47709 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen A1 - Yim, Y.-H. A1 - Lee, K.-S. A1 - Goenaga-Infante, H. A1 - Malinowskiy, D. A1 - Ren, T. A1 - Wang, J. A1 - Vocke Jr., R.D. A1 - Murphy, K. A1 - Nonose, N. A1 - Rienitz, O. A1 - Noordmann, J. A1 - Näykki, T. A1 - Sara-Aho, T. A1 - Ari, B. A1 - Cankur, O. T1 - Final report of the key comparison CCQM-K98: Pb isotope amount ratios in bronze KW - CCQM KW - Metrology KW - Isotope amount ratios KW - Lead PY - 2014 UR - http://www.bipm.org/utils/common/pdf/final_reports/QM/K98/CCQM-K98.pdf DO - https://doi.org/10.1088/0026-1394/51/1A/08017 SN - 0026-1394 SN - 1681-7575 VL - 51 IS - 1A Tech. Suppl. SP - 08017-1 EP - 08017-47 PB - Inst. of Physics Publ. CY - Bristol AN - OPUS4-31929 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -