TY - JOUR A1 - Stindt, Arne A1 - Andrade, M. A. B. A1 - Buurman, Merwe A1 - Adamowski, J. C. A1 - Panne, Ulrich A1 - Riedel, Jens T1 - Experimental and numerical characterization of the sound pressure in standing wave acoustic levitators N2 - A novel method for predictions of the sound pressure distribution in acoustic levitators is based on a matrix representation of the Rayleigh integral. This method allows for a fast calculation of the acoustic field within the resonator. To make sure that the underlying assumptions and simplifications are justified, this approach was tested by a direct comparison to experimental data. The experimental sound pressure distributions were recorded by high spatially resolved frequency selective microphone scanning. To emphasize the general applicability of the two approaches, the comparative studies were conducted for four different resonator geometries. In all cases, the results show an excellent agreement, demonstrating the accuracy of the matrix method. KW - Acoustic levitation KW - Matrix calculation KW - Microphony KW - Noncontact ultrasonic transportation KW - Small objects KW - Matrix-method KW - Droplets KW - Air KW - Manipulation KW - Simulation KW - Fields KW - Water PY - 2014 DO - https://doi.org/10.1063/1.4861197 SN - 0034-6748 SN - 1089-7623 N1 - Geburtsname von Buurman, Merwe: Albrecht, M. - Birth name of Buurman, Merwe: Albrecht, M. VL - 85 IS - 015110 SP - 1 EP - 6 PB - American Institute of Physics CY - Melville, NY AN - OPUS4-30405 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Porizka, Pavel A1 - Klessen, Benjamin A1 - Kaiser, J. A1 - Gornushkin, Igor B. A1 - Panne, Ulrich A1 - Riedel, Jens T1 - High repetition rate laser-induced breakdown spectroscopy using acousto-optically gated detection N2 - This contribution introduces a new type of setup for fast sample analysis using laser-induced breakdown spectroscopy (LIBS). The novel design combines a high repetition rate laser (up to 50 kHz) as excitation source and an acousto-optical modulator (AOM) as a fast switch for temporally gating the detection of the emitted light. The plasma radiation is led through the active medium of the AOM where it is diffracted on the transient ultrasonic Bragg grid. The diffracted radiation is detected by a compact Czerny-Turner spectrometer equipped with a CCD line detector. Utilizing the new combination of high repetition rate lasers and AOM gated detection, rapid measurements with total integration times of only 10 ms resulted in a limit of detection (LOD) of 0.13 wt.% for magnesium in aluminum alloys. This short integration time corresponds to 100 analyses/s. Temporal gating of LIP radiation results in improved LODs and consecutively higher sensitivity of the LIBS setup. Therefore, an AOM could be beneficially utilized to temporally detect plasmas induced by high repetition rate lasers. The AOM in combination with miniaturized Czerny-Turner spectrometers equipped with CCD line detectors and small footprint diode pumped solid state lasers results in temporally gateable compact LIBS setups. KW - Laser Induced Breakdown Spectroscopy KW - LIBS KW - Aluminium KW - Charge coupled devices KW - Magnesium KW - Light diffraction PY - 2014 DO - https://doi.org/10.1063/1.4890337 SN - 0034-6748 SN - 1089-7623 VL - 85 IS - 7 SP - 073104-1 EP - 073104-8 PB - American Institute of Physics CY - Melville, NY AN - OPUS4-31358 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Warschat, Carsten A1 - Stindt, Arne A1 - Panne, Ulrich A1 - Riedel, Jens T1 - Mass spectrometry of levitated droplets by thermally unconfined infrared-laser desorption N2 - An ionization scheme for fast online mass spectrometric interrogation of levitated droplets is presented. That renewed method comprises the output of an a Er:YAG laser at λ = 2.94 µm which is in resonance with the OH stretch vibration band of solvents like water and alcohols. A temporal pulse width larger than the time needed for pressure redistribution and also above the temperature redistribution time constant was found to lead to soft evaporation/ionization. Despite these mild desorption conditions, no additional postionization is found to be needed. Accordingly, the ionization is found to be very soft resulting in entirely intact analyte ions and concentration dependent cluster ions. Resulting mass spectra of small amino acids and large antibiotics are presented showing the versatility of the introduced technique. Above a critical mass of m ≈ 1 kDa, the formed ions carry multiple charges as it is typical for thermospray or electrospray ionization. The detection technique enables fast contactless analysis of the chemical composition of levitated microreactors and, thus, paves the way for future contactless reaction monitoring. KW - Laser desorption ionization KW - Levitated droplets KW - Ambient mass spectrometry PY - 2015 DO - https://doi.org/10.1021/acs.analchem.5b01495 SN - 0003-2700 SN - 1520-6882 VL - 87 IS - 16 SP - 8323 EP - 8327 PB - American Chemical Society CY - Washington, DC AN - OPUS4-35134 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Kern, Simon A1 - Michalik-Onichimowska, Aleksandra A1 - Riedel, Jens A1 - Panne, Ulrich A1 - King, Rudibert A1 - Maiwald, Michael ED - Maiwald, Michael T1 - “Click” analytics for “click” chemistry – a simple method for calibration-free evaluation of online NMR spectra N2 - Currently research in chemical manufacturing moves towards flexible plug-and-play approaches focusing on modular plants, capable of producing small scales on-demand with short down-times between individual cam-paigns. This approach allows for efficient use of hardware, a faster optimization of the process conditions, and thus, an accelerated introduction of new products to the market. Driven mostly by the search for chemical syntheses under biocompatible conditions, so-called “click” chemistry rapidly became a growing field of research. The re-sulting simple one-pot reactions are so far only scarcely accompanied by an adequate optimization via compara-bly straightforward and robust analysis techniques. Here we report on a fast and reliable calibration-free online high field NMR monitoring approach for technical mixtures. It combines a versatile fluidic system, continuous-flow measurement with a time interval of 20 s per spectrum, and a robust, automated algo-rithm to interpret the ob-tained data. All spectra were acquired using a 500 MHz NMR spectrometer (Varian) with a dual band flow probe having a 1/16 inch polymer tubing working as a flow cell. Single scan 1H spectra were recorded with an acquisition time of 5 s, relaxation delay of 15 s. T2 - 12. Kolloquium des Arbeitskreises Prozessanalytik CY - Berlin, Germany DA - 28.11.2016 KW - “Click” chemistry KW - Online NMR KW - Online monitoring PY - 2016 SP - 72 EP - 74 AN - OPUS4-38385 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bierstedt, Andreas A1 - Panne, Ulrich A1 - Riedel, Jens T1 - Confinement and enhancement of an airborne atmospheric laser-induced plasma using an ultrasonic acoustic resonator N2 - Optical elemental analysis in the gas phase typically relies on electrically driven plasmas. As an alternative approach, laser-induced plasmas (LIPs) have been suggested but have so far been only scarcely used. Here, a novel signal enhancement strategy for laser-based airborne plasma optical Emission spectroscopy for gas phase analytics is presented. In contrast to an electrically driven plasma, in the laser-induced analogue dynamic matter transport equilibrium builds up. The latter results in a rarefied density regime in the plasma core itself, surrounded by an area of compressed matter. The central rarefaction leads to a decrease in plasma intensity and analyte number density, both of which are detrimental for analytical purposes. Since the repetitive ignition of LIPs is a transient process, a restoration of the former gaseous medium by other dynamically equilibrated diffusion processes would be favourable. The presented combination of an airborne LIP and an ultrasonic acoustic resonator yields a fourfold signal enhancement while the Background contribution of ubiquitous air is at the same time effectively suppressed. Since the entire enhancement effect occurs without contact, no additional sources for abrasive sample contamination are introduced. KW - DPSS laser KW - Laser-induced plasma KW - High repetition rate KW - Ultrasonic acoustic resonator KW - Optical emission spectroscopy PY - 2018 DO - https://doi.org/10.1039/C7JA00297A SN - 0267-9477 SN - 1364-5544 VL - 33 IS - 1 SP - 135 EP - 140 PB - Royal Society of Chemistry CY - London AN - OPUS4-43619 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Riedel, Jens T1 - Recent Trends in Inexpensive Lithium Isotopic Analysis N2 - Lithium exists in two stable isotopes, 6Li and 7Li. The ratio of these in every ore varies depending on the geological history of the sample, thus providing a tool for fingerprinting the distinct origin of Li containing samples. Determination of the exact isotope ratio for e.g. designation of provenance today relies on expensive and bulky instrumentation such as multi `collector inductively coupled plasma mass spectrometry` (MC-ICP-MS). These instruments, however, are known to bear pitfalls in the characterization of particular elements including Lithium. BAM recently developed two alternative analytical devices for this task, solely relying on inexpensive optical spectroscopy in combination with state-of-the-art multivariate data analysis such as Machine learning algorithms. Both techniques have been comprehensively studied using certified reference materials and comparing the results to MC-ICP-MS results and could be shown to result in comparable figures of merit, paving the way for a more general accessibility to provenance determination instrumentation. The results also pave the way towards even further simplification of the laboratory infrastructure demands and to further include additional elements into the isotopic fingerprinting methodology. T2 - Lithium-Days CY - Halle, Germany DA - 06.12.2021 KW - Lithium KW - Isotopes KW - Optical Spectroscopy PY - 2021 AN - OPUS4-53966 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Michalik-Onichimowska, Aleksandra A1 - Beitz, T. A1 - Panne, Ulrich A1 - Löhmannsröben, H.-G. A1 - Riedel, Jens T1 - Microsecond mid-infrared laser pulses for atmospheric pressure laser ablation/ionization of liquid samples N2 - In many laser based ionization techniques with a subsequent drift time separation, the laser pulse generating the ions is considered as the start time t0. Therefore, an accurate temporal definition of this event is crucial for the resolution of the experiments. In this contribution, the laser induced plume dynamics of liquids evaporating into atmospheric pressure are visualized for two distinctively different laser pulse widths, Δt = 6 nanoseconds and Δτ = 280 microseconds. For ns-pulses the expansion of the generated vapour against atmospheric pressure is found to lead to turbulences inside the gas phase. This results in spatial and temporal broadening of the nascent clouds. A more equilibrated expansion, without artificial smearing of the temporal resolution can, in contrast, be observed to follow μs-pulse excitation. This leads to the counterintuitive finding that longer laser pulses results in an increased temporal vapour formation definition. To examine if this fume expansion also eventually results in a better definition of ion formation, the nascent vapour plumes were expanded into a linear drift tube ion mobility spectrometer (IMS). This time resolved detection of ion formation corroborates the temporal broadening caused by collisional impeding of the supersonic expansion at atmospheric pressure and the overall better defined ion formation by evaporation with long laser pulses. A direct comparison of the observed results strongly suggests the coexistence of two individual ion formation mechanisms that can be specifically addressed by the use of appropriate laser sources. KW - Plume KW - Laser ablation KW - Ion mobility spectrometry KW - Pulse duration KW - Ionization PY - 2017 DO - https://doi.org/10.1016/j.snb.2016.06.155 VL - 238 SP - 298 EP - 305 PB - Elsevier CY - Amsterdam, NL AN - OPUS4-37158 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Kern, Simon A1 - Michalik-Onichimowska, Aleksandra A1 - Riedel, Jens A1 - Panne, Ulrich A1 - King, R. A1 - Maiwald, Michael T1 - “Click” analytics for “click” chemistry – a simple method for calibration-free evaluation of online NMR spectra N2 - Currently research in chemical manufacturing moves towards flexible plug-and-play approaches focusing on modular plants, capable of producing small scales ondemand with short down-times between individual campaigns. This approach allows for efficient use of hardware, a faster optimization of the process conditions, and thus, an accelerated introduction of new products to the market. Driven mostly by the search for chemical syntheses under biocompatible conditions, so-called “click” chemistry rapidly became a growing field of research. The resulting simple one-pot reactions are so far only scarcely accompanied by an adequate optimization via comparably straightforward and robust analysis techniques. Here we report on a fast and reliable calibration-free online high field NMR monitoring approach for technical mixtures. It combines a versatile fluidic system, continuous-flow measurement with a time interval of 20 s per spectrum, and a robust, automated algorithm to interpret the obtained data. All spectra were acquired using a 500 MHz NMR spectrometer (Varian) with a dual band flow probe having a 1/16 inch polymer tubing working as a flow cell. Single scan 1H spectra were recorded with an acquisition time of 5 s, relaxation delay of 15 s. As a proof-of-concept, the thiol-ene coupling between N-boc cysteine methyl ester and allyl alcohol was conducted in non-deuterated solvents while its time-resolved behaviour was characterised with step tracer experiments Through the application of spectral modeling the signal area for each reactant can be deconvoluted in the online spectra and thus converted to the respective concentrations or molar ratios. The signals which were suitable for direct integration were used herein for comparison purposes of both methods. T2 - 4th European Conference on Process Analytics and Control Technology (EuroPACT 2017) CY - Potsdam, Germany DA - 10.05.2017 KW - Online NMR Spectroscopy KW - Process Analytical Technology KW - Click Chemistry KW - Reaction Monitoring KW - Process control KW - Indirect Hard Modeling KW - Spectral Modeling PY - 2017 SP - 156 EP - 157 CY - Frankfurt a. M. AN - OPUS4-40232 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bierstedt, Andreas A1 - Kersten, H. A1 - Glaus, Reto A1 - Gornushkin, Igor B. A1 - Panne, Ulrich A1 - Riedel, Jens T1 - Characterization of an airborne laser-spark ion source for ambient mass spectrometry N2 - An airborne laser plasma is suggested as an ambient ion source for mass spectrometry. Its fundamental physical properties, such as an excellent spatial and temporal definition, high electron and ion densities and a high effective cross section in maintaining the plasma, make it a promising candidate for future applications. For deeper insights into the plasma properties, the optical plasma emission is examined and compared to mass spectra. The results show a seemingly contradictory behavior, since the emitted light reports the plasma to almost entirely consist of hot elemental ions, while the corresponding mass spectra exhibit the formation of intact molecular species. Further experiments, including time- resolved shadowgraphy, spatially resolved mass spectrometry, as well as flow-dependent emission spectroscopy and mass spectrometry, suggest the analyte molecules to be formed in the cold plasma vicinity upon interaction with reactive species formed inside the hot plasma center. Spatial separation is maintained by concentrically expanding pressure waves, inducing a strong unidirectional diffusion. The accompanying rarefaction inside the plasma center can be compensated by a gas stream application. This replenishing results in a strong increase in emission brightness, in local reactive species concentration, and eventually in direct mass spectrometric sensitivity. To determine the analytical performance of the new technique, a comparison with an atmospheric pressure chemical ionization (APCI) source was conducted. Two kitchen herbs, namely, spearmint and basil, were analyzed without any sample pretreatment. The presented results demonstrate a considerably higher sensitivity of the presented laser-spark ionization technique. KW - Laser-spark KW - Laser induced plasma KW - Ambient mass spectrometry KW - Optical emission spectroscopy KW - Ionization PY - 2017 DO - https://doi.org/10.1021/acs.analchem.6b04178 SN - 0003-2700 SN - 1520-6882 VL - 89 IS - 6 SP - 3437 EP - 3444 PB - American Chemical Society CY - Washington, DC, USA AN - OPUS4-39474 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Porizka, Pavel A1 - Demidov, Alexandr A1 - Kaiser, J. A1 - Keivanian, J. A1 - Gornushkin, Igor B. A1 - Panne, Ulrich A1 - Riedel, Jens T1 - Laser-induced breakdown spectroscopy for in situ qualitative and quantitative analysis of mineral ores N2 - In this work, the potential of laser-induced breakdown spectroscopy (LIBS) for discrimination and analysis of geological materials was examined. The research was focused on classification of mineral ores using their LIBS spectra prior to quantitative determination of copper. Quantitative analysis is not a trivial task in LIBS measurement because intensities of emission lines in laser-induced plasmas (LIP) are strongly affected by the sample matrix (matrix effect). To circumvent this effect, typically matrix-matched standards are used to obtain matrix-dependent calibration curves. If the sample set consists of a mixture of different matrices, even in this approach, the corresponding matrix has to be known prior to the downstream data analysis. For this categorization, the multielemental character of LIBS spectra can be of help. In this contribution, a principal component analysis (PCA) was employed on the measured data set to discriminate individual rocks as individual matrices against each other according to their overall elemental composition. Twenty-seven igneous rock samples were analyzed in the form of fine dust, classified and subsequently quantitatively analyzed. Two different LIBS setups in two laboratories were used to prove the reproducibility of classification and quantification. A superposition of partial calibration plots constructed from the individual clustered data displayed a large improvement in precision and accuracy compared to the calibration plot constructed from all ore samples. The classification of mineral samples with complex matrices can thus be recommended prior to LIBS system calibration and quantitative analysis. KW - Laser-induced breakdown spectroscopy KW - LIBS KW - Chemometrics KW - Principal component analysis KW - Geochemical analysis PY - 2014 DO - https://doi.org/10.1016/j.sab.2014.08.027 SN - 0584-8547 SN - 0038-6987 VL - 101 SP - 155 EP - 163 PB - Elsevier CY - Amsterdam AN - OPUS4-32588 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -