TY - JOUR A1 - de Oliveira Guilherme Buzanich, Ana A1 - Radtke, Martin A1 - Reinholz, Uwe A1 - Riesemeier, Heinrich A1 - Emmerling, Franziska ED - de Oliveira Guilherme Buzanich, Ana T1 - Time- and spatial-resolved X-ray absorption fine structure (XAFS) spectroscopy in a single-shot: new analytical possibilities for in situ material characterization N2 - A new concept that comprises both time- and lateral-resolved X-ray absorption fine-structure information simultaneously in a single shot is presented. This uncomplicated set-up was tested at the BAMline at BESSY-II (Berlin, Germany). The primary broadband beam was generated by a double multilayer monochromator. The transmitted beam through the sample is diffracted by a convexly bent Si (111) crystal, producing a divergent beam. This, in turn, is collected by either an energy-sensitive area detector, the so-called color X-ray camera, or by an area-sensitive detector based on a CCD camera, in θ-2θ geometry. The first tests were performed with thin metal foils and some iron oxide mixtures. A time resolution of lower than 1 s together with a spatial resolution in one dimension of at least 50 µm is achieved. KW - Single-shot XAFS KW - Time resolution KW - Spatial resolution KW - Divergent XAFS PY - 2016 DO - https://doi.org/10.1107/S1600577516003969 SN - 1600-5775 VL - 23 SP - 769 EP - 776 PB - International Union of Crystallography AN - OPUS4-38370 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Kulow, Anicó A1 - de Oliveira Guilherme Buzanich, Ana A1 - Reinholz, Uwe A1 - Emmerling, Franziska A1 - Hampel, S. A1 - Fittschen, UEA A1 - Streli, C. A1 - Radtke, Martin T1 - Reconstruction for coded aperture full-field x-ray fluorescence imaging N2 - X-ray fluorescence imaging is a well-established tool in materials characterization. In this work, we present the adaption of coded aperture imaging to full-field X-ray fluorescence imaging at the synchrotron. Coded aperture imaging has its origins in astrophysics, and has several advantages: Coded apertures are relatively easy to fabricate, achromatic, allow a high photon throughput, and high angular acceptance. Coded aperture imaging is a two-step-process, consisting of the measurement process and a reconstruction step. Different programs have been written, for the raytracing/forward projection and the reconstruction. Experiments with coded aperture in combination with a Color X-ray Camera and an energy-dispersive area detector, have been conducted at the BAMline. Measured samples were successfully reconstructed, and gave a 9.1-fold increase in count rate compared to a polycapillary optic. KW - Synchrotron KW - BAMline KW - Coded Aperture PY - 2022 SN - 1097-0002 VL - 65 SP - 57 EP - 70 PB - Cambridge University Press CY - Cambridge AN - OPUS4-56350 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Cakir, Cafer Tufan A1 - Piotrowiak, T. A1 - Reinholz, Uwe A1 - Ludwig, A. A1 - Emmerling, Franziska A1 - Streli, C. A1 - de Oliveira Guilherme Buzanich, Ana A1 - Radtke, Martin T1 - Exploring the Depths of Corrosion: A Novel GE-XANES Technique for Investigating Compositionally Complex Alloys N2 - In this study, we propose the use of nondestructive, depth-resolved, element-specific characterization using grazing exit X-ray absorption near-edge structure spectroscopy (GE-XANES) to investigate the corrosion process in compositionally complex alloys (CCAs). By combining grazing exit X-ray fluorescence spectroscopy (GE-XRF) geometry and a pnCCD detector, we provide a scanning-free, nondestructive, depth-resolved analysis in a sub-micrometer depth range, which is especially relevant for layered materials, such as corroded CCAs. Our setup allows for spatial and energy-resolved measurements and directly extracts the desired fluorescence line, free from scattering events and other overlapping lines. We demonstrate the potential of our approach on a compositionally complex CrCoNi alloy and a layered reference sample with known composition and specific layer thickness. Our findings indicate that this new GE-XANES approach has exciting opportunities for studying surface catalysis and corrosion processes in real-world materials. KW - Degradation mechanisms KW - Grazin exit XANES KW - Depth resolved XANES KW - Compositional complex alloys KW - Corrosion PY - 2023 DO - https://doi.org/10.1021/acs.analchem.3c00404 VL - 95 SP - 4810 EP - 4818 PB - ACS Publications AN - OPUS4-57823 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Reinholz, Uwe T1 - Archäometrie an der BAMline T2 - Lange Nacht der Wissenschaften, BAM, BESSY II CY - Berlin, Germany DA - 2010-06-05 PY - 2010 AN - OPUS4-21379 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Polte, Jörg A1 - Emmerling, Franziska A1 - Radtke, Martin A1 - Reinholz, Uwe A1 - Riesemeier, Heinrich A1 - Thünemann, Andreas T1 - Real-time monitoring of copolymer stabilized growing gold nanoparticles N2 - A great challenge in the production of nanoparticles with defined sizes and properties is to control their growth in situ. We developed a dedicated combined small-angle X-ray scattering (SAXS) and X-ray absorption spectroscopy (XAS) setup to monitor nanoparticle formation in solution. The capabilities of simultaneously deriving particle sizes and oxidation states of atoms/ions are illustrated for the formation of spherical gold nanoparticles by the reduction of hydrogen tetrachloroaureate (HAuCl4). Particles with initial radii of 4.60 ± 0.10 nm and final radii of 5.67 ± 0.10 nm were produced in a levitated droplet with a volume of 4 µL. An ethylene oxide/propylene oxide triblock copolymer PEO-PPO-PEO (Pluronic F-127) functions as reducing agent and colloidal stabilizer. XANES shows in situ how the gold was reduced in the droplet from Au(III) to Au(0), and simultaneously SAXS recorded the size distribution of the formed nanoparticles. It is shown that the final particle number is reached quickly. Thereafter, only the particles’ sizes increase. Comparison of XANES and SAXS shows that the quantity of Au(0) is higher than the quantity of gold located in the nanoparticles while the particles are growing. Finally, all the Au(0) is found in the nanoparticles. We tentatively attribute this finding to the pseudo crown ether effect of the polymer surfactant that kinetically stabilizes gold atoms when formed from gold ions within their protecting cavity. A simple “burst” mechanism for the gold nanoparticle formation is the consequence. The possibility of an inhomogeneous particles structure with an enhanced density near the particle surface is discussed. PY - 2010 DO - https://doi.org/10.1021/la903829q SN - 0743-7463 SN - 1520-5827 VL - 26 IS - 8 SP - 5889 EP - 5894 PB - American Chemical Society CY - Washington, DC AN - OPUS4-21223 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Polte, Jörg A1 - Kraehnert, R. A1 - Radtke, Martin A1 - Reinholz, Uwe A1 - Riesemeier, Heinrich A1 - Thünemann, Andreas A1 - Emmerling, Franziska T1 - New insights of the nucleation and growth process of gold nanoparticles via in situ coupling of SAXS and XANES N2 - Although metallic nanoparticles play an important role in the area of nanotechnology, a coherent mechanistic explanation for the evolution of the particles during their chemical synthesis has not yet been provided in many cases. To gain a profound understanding of the growth mechanism of colloidal nanoparticles, new approaches using Small Angle X-Ray Scattering (SAXS) combined with X-ray absorption near-edge structure (XANES) are presented. This combination allows for insights into two prominent syntheses routes of gold nanoparticles (GNP): The 'slow' reaction using sodium citrate (30-90 min) as a reducing agent and the 'fast' reaction employing NaBH4 (within few seconds). In the first case data derived with the coupled XANES and SAXS suggests a four-step particle formation mechanism. For the second system a time resolution in the order of 100-200 ms was achieved by coupling a common laboratory SAXS instrument with a microstructured mixer, which allows data acquisition in a continuous-flow mode. The results indicate a coalescence driven growth process. Based on the capabilities to deduce the size, number and polydispersity of the particles, the results of both methods enable the development of mechanistic schemes explaining the different phases of particle formation and growth, thus providing a basis for improved control over the synthesis processes. KW - SAXS KW - XANES KW - Nanoparticle formation PY - 2010 DO - https://doi.org/10.1088/1742-6596/247/1/012051 SN - 1742-6588 SN - 1742-6596 VL - 247 IS - 1 SP - 012051-1 - 012051-10 PB - IOP Publ. CY - Bristol, UK AN - OPUS4-22684 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -