TY - JOUR A1 - Ordavo, I. A1 - Ihle, S. A1 - Arkadiev, V. A1 - Scharf, Oliver A1 - Soltau, H. A1 - Bjeoumikhov, A. A1 - Bjeoumikhova, S. A1 - Buzanich, Günter A1 - Gubzhokov, R. A1 - Günther, A. A1 - Hartmann, R. A1 - Holl, P. A1 - Kimmel, N. A1 - Kühbacher, M. A1 - Lang, M. A1 - Langhoff, N. A1 - Liebel, A. A1 - Radtke, Martin A1 - Reinholz, Uwe A1 - Riesemeier, Heinrich A1 - Schaller, G. A1 - Schopper, F. A1 - Strüder, L. A1 - Thamm, C. A1 - Wedell, R. T1 - A new pnCCD-based color X-ray camera for fast spatial and energy-resolved measurements N2 - We present a new high resolution X-ray imager based on a pnCCD detector and a polycapillary optics. The properties of the pnCCD like high quantum efficiency, high energy resolution and radiation hardness are maintained, while color corrected polycapillary lenses are used to direct the fluorescence photons from every spot on a sample to a corresponding pixel on the detector. The camera is sensitive to photons from 3 to 40 keV with still 30% quantum efficiency at 20 keV. The pnCCD is operated in split frame mode allowing a high frame rate of 400 Hz with an energy resolution of 152 eV for Mn Kα (5.9 keV) at 450 kcps. In single-photon counting mode (SPC), the time, energy and position of every fluorescence photon is recorded for every frame. A dedicated software enables the visualization of the elements distribution in real time without the need of post-processing the data. A description of the key components including detector, X-ray optics and camera is given. First experiments show the capability of the camera to perform fast full-field X-Ray Fluorescence (FF-XRF) for element analysis. The imaging performance with a magnifying optics (3×) has also been successfully tested. KW - X-ray CCD camera KW - pnCCD KW - Fast X-ray imaging KW - XRF KW - Full-field X-ray fluorescence KW - Elemental analysis KW - High quantum efficiency KW - High energy resolution KW - Polycapillary optics PY - 2011 DO - https://doi.org/10.1016/j.nima.2011.05.080 SN - 0168-9002 SN - 0167-5087 VL - 654 SP - 250 EP - 257 PB - North-Holland CY - Amsterdam AN - OPUS4-24370 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Scharf, Oliver A1 - Ihle, S. A1 - Ordavo, I. A1 - Arkadiev, V. A1 - Bjeoumikhov, A. A1 - Bjeoumikhova, S. A1 - Buzanich, Günter A1 - Gubzhokov, R. A1 - Günther, A. A1 - Hartmann, R. A1 - Kühbacher, M. A1 - Lang, M. A1 - Langhoff, N. A1 - Liebel, A. A1 - Radtke, Martin A1 - Reinholz, Uwe A1 - Riesemeier, Heinrich A1 - Soltau, H. A1 - Strüder, L. A1 - Thünemann, Andreas A1 - Wedell, R. T1 - Compact pnCCD-based X-ray camera with high spatial and energy resolution: a color X-ray camera N2 - For many applications there is a requirement for nondestructive analytical investigation of the elemental distribution in a sample. With the improvement of X-ray optics and spectroscopic X-ray imagers, full field X-ray fluorescence (FF-XRF) methods are feasible. A new device for high-resolution X-ray imaging, an energy and spatial resolving X-ray camera, is presented. The basic idea behind this so-called 'color X-ray camera' (CXC) is to combine an energy dispersive array detector for X-rays, in this case a pnCCD, with polycapillary optics. Imaging is achieved using multiframe recording of the energy and the point of impact of single photons. The camera was tested using a laboratory 30 µm microfocus X-ray tube and synchrotron radiation from BESSY II at the BAMline facility. These experiments demonstrate the suitability of the camera for X-ray fluorescence analytics. The camera simultaneously records 69696 spectra with an energy resolution of 152 eV for manganese Kα with a spatial resolution of 50 µm over an imaging area of 12.7 × 12.7 mm². It is sensitive to photons in the energy region between 3 and 40 keV, limited by a 50 µm beryllium window, and the sensitive thickness of 450 µm of the chip. Online preview of the sample is possible as the software updates the sums of the counts for certain energy channel ranges during the measurement and displays 2-D false-color maps as well as spectra of selected regions. The complete data cube of 264 × 264 spectra is saved for further qualitative and quantitative processing. KW - Color X-ray camera KW - X-ray fluorescence KW - pnCCD KW - Polycapillary optic KW - Synchrotron radiation PY - 2011 DO - https://doi.org/10.1021/ac102811p SN - 0003-2700 SN - 1520-6882 VL - 83 IS - 7 SP - 2532 EP - 2538 PB - American Chemical Society CY - Washington, DC AN - OPUS4-23948 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Mans, C. A1 - Simons, C. A1 - Hanning, S. A1 - Janßen, A. A1 - Alber, D. A1 - Radtke, Martin A1 - Reinholz, Uwe A1 - Bühler, A. A1 - Kreyenschmidt, M. T1 - New polymeric candidate reference materials for XRF and LA-ICP-MS - development and preliminary characterization N2 - A set of ten calibration materials based on acrylonitrile-butadiene-styrene terpolymer (ABS) containing the elements Br, Pb, Cd, Cr, and Hg was produced in order to control polymer samples in regard to the restriction on the use of certain hazardous substances (RoHS) directive. The materials were produced with respect to the special demands of X-ray fluorescence analysis (XRF) and inductively coupled plasma mass spectrometry in combination with laser ablation (LA-ICP-MS). The mass fractions of all elements were 0-1500 mg/kg and the materials were produced as granulates and solid discs with a diameter of 40 mm and thicknesses of 1, 2, and 6 mm. It could be shown that the signals measured by XRF strongly depend on the thickness of polymeric samples. Macroscopic homogeneity was assessed with XRF measuring a total of 180 samples. The microscopic homogeneities were determined for a material with average elemental mass fractions with the aid of synchrotron radiation (SR) µ-XRF and LA-ICP-MS. Sufficient macroscopic and microscopic homogeneities for all elements could be achieved. It was observed that organic additives show a better homogeneity than oxides. XRF and LA-ICP-MS were successfully calibrated with the new materials. Limits of detection were in the range of 0.4-22 mg/kg for XRF and 2.3-26.8 mg/kg for LA-ICP-MS. The materials are considered as candidate reference materials (RM) by the Federal Institute of Materials Research and Testing (BAM, Germany). KW - XRF KW - LA.ICP-MS Plastics KW - RoHS KW - CRM PY - 2009 DO - https://doi.org/10.1002/xrs.1120 SN - 0049-8246 VL - 38 IS - 1 SP - 52 EP - 57 PB - Wiley CY - Chichester AN - OPUS4-18681 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Strüder, L. A1 - Hartmann, R. A1 - Holl, P. A1 - Ihle, S. A1 - Huth, M. A1 - Schmidt, J. A1 - Tham, Ch. A1 - Kanngießer, B. A1 - Baumann, J. A1 - Renno, A. A1 - Grenzer, J. A1 - Radtke, Martin A1 - Abboud, A. A1 - Pietsch, U. A1 - Soltau, H. T1 - High Speed, High Resolution imaging spectrometers based on pnCCDs for XRF and XRD applications N2 - For many years pnCCDs have been well known as X-ray detectors for spectroscopic imaging in many fields of science: X-Ray Fluorescence analysis (XRF), X-ray Diffraction (XRD) with light sources in large accelerator facilities as well as with laboratory light sources or with X-rays from celestial sources in X-ray astronomy. A brief introduction in GEXRF (Grazing Emission XRF) measurements with a laboratory laser produced plasma source will be given, PIXE (Particle Induced X-ray Emission) measurements and D2XRF (Double Dispersive X-Ray Fluorescence) and Slicing experiments with pnCCDs coupled to polycapillary optics performed at the BESSY synchrotron will be shown. Energy-dispersive Laue diffraction with ultra-hard X-rays for the analysis of defects in metals will conclude the overview of spectroscopic X-ray imaging measurements in the field of structure and dynamics of matter. KW - PnCCD KW - Synchrotron KW - XRF PY - 2016 DO - https://doi.org/10.1017/S1431927616001355 VL - 22 IS - S3 SP - 100 EP - 101 AN - OPUS4-38884 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Reiche, I. A1 - Muller, K. A1 - Alberic, M. A1 - Scharf, O. A1 - Wahning, A. A1 - Bjeoumikhov, A. A1 - Radtke, Martin A1 - Simon, R. T1 - Discovering vanished paints and naturally formed gold nanoparticles on 2800 years old Phoenician ivories using SR-FF-microXRF with the color X-ray camera N2 - Phoenician ivory objects (8th century B.C., Syria) from the collections of the Badisches Landesmuseum, Karlsruhe, Germany, have been studied with full field X-ray fluorescence microimaging, using synchrotron radiation (SR-FF-microXRF). The innovative Color X-ray Camera (CXC), a full-field detection device (SLcam), was used at the X-ray fluorescence beamline of the ANKA synchrotron facility (ANKA-FLUO, KIT, Karlsruhe, Germany) to noninvasively study trace metal distributions at the surface of the archeological ivory objects. The outstanding strength of the imaging technique with the CXC is the capability to record the full XRF spectrum with a spatial resolution of 48 µm on a zone of a size of 11.9 × 12.3 mm² (264 × 264 pixels). For each analyzed region, 69696 spectra were simultaneously recorded. The principal elements detected are P, Ca, and Sr, coming from the ivory material itself; Cu, characteristic of pigments; Fe and Pb, representing sediments or pigments; Mn, revealing deposited soil minerals; Ti, indicating restoration processes or correlated with Fe sediment traces; and Au, linked to a former gilding. This provides essential information for the assessment of the original appearance of the ivory carvings. The determined elemental maps specific of possible pigments are superimposed on one another to visualize their respective distributions and reconstruct the original polychromy and gilding. Reliable hypotheses for the reconstruction of the original polychromy of the carved ivories are postulated on this basis. KW - SR-FF-microXRF KW - Elemental imaging KW - Ivory KW - Paints KW - Au NPs KW - Egyptian blue KW - Fe PY - 2013 DO - https://doi.org/10.1021/ac4006167 SN - 0003-2700 SN - 1520-6882 VL - 85 IS - 12 SP - 5857 EP - 5866 PB - American Chemical Society CY - Washington, DC AN - OPUS4-30222 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Aldrabee, A. A1 - Wriekat, A. A1 - Abu Saleem, K. A1 - Radtke, Martin T1 - Non-destructive SR-XRF analysis of ancient Mamluk-Ayyubid glazed pottery fragments from Karak Castle, Jordan N2 - Sixteen Ayyubid-Mamluk glazed pottery sherds were analyzed in order to identify and characterize the elemental composition to determine their provenance. The tested sherds were collected from the historical site of Karak Castle, southern Jordan. Chemical analysis for the sixteen samples has been carried out using Synchrotron Radiation X-ray Fluorescence Spectrometry (SR-XRF) Technique. Furthermore, the semiquantitative analysis of the elements Fe – Cu – Zn – Br – Rb – Sr – Y – Zr – Nb – Mo – Pd – Ag – Cd and Pb has been performed for the samples based on Principal Component Analysis (PCA) and hierarchical Cluster Analysis with Bray-Curtis in order to define grouping of different glazed pottery by obtaining information on their similarity and clustering. The results of chemical analysis provided persuasive evidence that the Karak Castle pottery sherds have at least three different sources of provenance. KW - BAMline KW - Karak Castle KW - Synchrotron KW - XRF PY - 2015 VL - 8 IS - 3 SP - 157 EP - 163 AN - OPUS4-43489 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rauwolf, M. A1 - Turyanskaya, A. A1 - Roschger, A. A1 - Prost, J. A1 - Simon, R. A1 - Scharf, O. A1 - Radtke, Martin A1 - Schoonjans, Tom A1 - de Oliveira Guilherme Buzanich, Ana A1 - Klaushofer, K. A1 - Wobrauschek, P. A1 - Hofstaetter, J. G. A1 - Roschger, P. A1 - Streli, C. T1 - Synchrotron radiation micro X-ray fluorescence spectroscopy of thin structures in bone samples: comparison of confocal and color X-ray camera setups N2 - In the quest for finding the ideal synchrotron-radiation-induced imaging method for the investigation of trace element distributions in human bone samples, experiments were performed using both a scanning confocal synchrotron radiation micro X-ray fluorescence (SR-mXRF) (FLUO beamline at ANKA) setup and a full-field color X-ray camera (BAMline at BESSY-II) setup. As zinc is a trace element of special interest in bone, the setups were optimized for its detection. The setups were compared with respect to count rate, required measurement time and spatial resolution. It was demonstrated that the ideal method depends on the element of interest. Although for Ca (a major constituent of the bone with a low energy of 3.69 keV for its K alpha XRF line) the color X-ray camera provided a higher resolution in the plane, for Zn (a trace element in bone) only the confocal SR-µXRF setup was able to sufficiently image the distribution. KW - Synchrotron KW - BAMline KW - BESSY KW - XRF KW - X-ray Color Camera PY - 2017 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-391944 DO - https://doi.org/10.1107/S1600577516017057 SN - 1600-5775 VL - 24 SP - 307 EP - 311 PB - International Union of Crystallography AN - OPUS4-39194 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - INPR A1 - Nowak, S. H. A1 - Petric, M. A1 - Buchriegler, J. A1 - Bjeoumikhov, A. A1 - Bjeoumikhova, Z. A1 - von Borany, J. A1 - Munnik, F. A1 - Radtke, Martin A1 - Renno, A. D. A1 - Reinholz, Uwe A1 - Scharf, O. A1 - Tilgner, J. A1 - Wedell, R. T1 - Road to micron resolution with a color X-ray camera - Polycapillary optics characterization N2 - In a color X-ray camera spatial resolution is achieved by means of a polycapillary optic conducting X-ray photons from small regions on a sample to distinct energy dispersive pixels on a CCD matrix. At present, the resolution limit of color X-ray camera systems can go down to several microns and is mainly restricted by Pixel dimensions. The recent development of an efficient subpixel resolution algorithm allows a release from pixel size, limiting the resolution only to the quality of theoptics. In this work polycapillary properties that influence the spatial resolution are systematized and assessed both theoretically and experimentally. It is demonstrated that with the current technological Level reaching one micron resolution is challenging, but possible. KW - Color X-ray camera KW - Polycapillary optics PY - 2017 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-404404 DO - https://doi.org/10.48550/arXiv.1705.08939 SN - 2331-8422 SP - 1 EP - 11 PB - Cornell University CY - Ithaca, NY AN - OPUS4-40440 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Fittschen, U.E.A. A1 - Möckel, R. A1 - Schreiner, M. A1 - klinger, M. A1 - Radtke, Martin A1 - Meyer, B. A1 - Guhl, S. A1 - Renno, A. A1 - Godinho, J. A1 - Gloaguen, R. A1 - Gutzmer, J. T1 - Bundling analytical capacities to understand phase formation in recycling of functional materials N2 - Transitioning from combustion engine-driven transportation to e-mobility demands a paradigm shift – from a system geared to maximize energy efficiency (i.e. fuel consumption) to a system that may be constrained by the availability of high technology (critical) metals required for electrical energy storage systems and drives. In the wake of these developments efforts in securing new resources of these metals from recycling of end-of-life products are increasing steadily. Recycling of Li-Ion batteries has recently been evaluated. The results pinpoint to a critical need for understanding slag Formation and its dependence on metal components like Mn under extreme conditions. This will allow researchers to predict optimal Operation setting and to react quickly to changing market demands (which may be Li or Co at one point but may also shift to Ni or rare earth elements (REE)). The long-term goal is to control the formation of specific phases in slags allowing for a Maximum yield of elements of interest and optimal recovery in the separation processes that follows. The combination of data on the physical micro structure and local chemistry of the multi-Phase products during and after processing will help to understand and derive thermodynamic and kinetic data on its formation. In this paper we are giving an overview on the analytical challenges and approaches to provide robust data on local element concentration and species (especially Mn which is a common component of next generation Li-ion batteries cathodes), spanning the dimensions from the nanometer scale to the bulk material. The complementary interactions of X-rays and electrons make them ideal probes to collect Interface and “in-depth” information. Before- and -after studies as well as in situ structural changes and Phase (trans)formation, changes in elemental and elemental species (e.g. oxidation state) distribution may be tracked by X-ray diffraction (XRD), X-ray fluorescence microscopy and X-ray Absorption spectroscopy. The application of such advanced analytical tools will not only provide essential clues during early lab-based experiments towards the development of new recycling technologies, but may also be deployed for on-line and in-line monitoring of industrial processes. KW - Synchrotron KW - XANES KW - Slags KW - Battery PY - 2019 DO - https://doi.org/10.4028/www.scientific.net/MSF.959.183 SN - 1662-9752 VL - 959 SP - 183 EP - 190 PB - Trans Tech Publ. AN - OPUS4-48900 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Nowak, S.H. A1 - Bjeoumikhov, A. A1 - Von Borany, J. A1 - Buchriegler, J. A1 - Munnik, F. A1 - Petric, M. A1 - Renno, A.D. A1 - Radtke, Martin A1 - Reinholz, Uwe A1 - Scharf, O. A1 - Strüder, L. A1 - Wedell, R. A1 - Ziegenrücker, R. T1 - Examples of XRF and PIXE imaging with few microns resolution using SLcam® a color X-ray camera N2 - We present results of recent development of the color X-ray camera, type SLcam®, allowing detection of X-ray images with few microns resolution. Such spectral resolution is achieved with the use of high-quality polycapillary optics combined with sub-pixel resolution. Imaging of Siemens star resolution test chart reveals that the resolution limit of SLcam® can go down to nearly 5µm. Several real sample examples of measurements carried out at the laboratory, synchrotron, and particle-induced X-ray emission beamlines are shown. This is the first time SLcam® is used as particle-induced X-ray emission detector. KW - X-ray KW - Color X-ray camera KW - CXC KW - Synchrotron radiation KW - BESSY KW - BAMline PY - 2015 DO - https://doi.org/10.1002/xrs.2590 SN - 0049-8246 VL - 44 IS - 3 SP - 135 EP - 140 PB - Wiley & Sons, Ltd. CY - Chichester AN - OPUS4-33170 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Nowak, S.H. A1 - Bjeoumikhov, A. A1 - Von Borany, J. A1 - Buchriegler, J. A1 - Munnik, F. A1 - Petric, M. A1 - Radtke, Martin A1 - Renno, A.D. A1 - Reinholz, Uwe A1 - Scharf, O. A1 - Wedell, R. T1 - Sub-pixel resolution with a color X-ray camera N2 - The color X-ray camera SLcam® is a full-field, single photon detector providing scanning-free, energy and spatially resolved X-ray imaging. Spatial resolution is achieved with the use of polycapillary optics guiding X-ray photons from small regions on a sample to distinct energy dispersive pixels on a charged-coupled device detector. Applying sub-pixel resolution, signals from individual capillary channels can be distinguished. Therefore, the SLcam® spatial resolution, which is normally limited to the pixel size of the charge-coupled device, can be improved to the size of individual polycapillary channels. In this work a new approach to a sub-pixel resolution algorithm comprising photon events also from the pixel centers is proposed. The details of the employed numerical method and several sub-pixel resolution examples are presented and discussed. KW - Color X-ray camera KW - Sub pixel analysis PY - 2015 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-334948 DO - https://doi.org/10.1039/c5ja00028a SN - 0267-9477 SN - 1364-5544 SP - 1890 EP - 1897 PB - Royal Society of Chemistry CY - London AN - OPUS4-33494 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Guilherme, A. A1 - Buzanich, Günter A1 - Radtke, Martin A1 - Reinholz, Uwe A1 - Coroado, J. A1 - Dos Santos, J.M.F. A1 - Carvalho, M.L. T1 - Synchrotron micro-XRF with compound refractive lenses (CRLs) for tracing key elements on Portuguese glazed ceramics N2 - Several glazed ceramic fragments (XVIth to XVIIIth centuries) from two production centers in Portugal (Coimbra and Lisbon) were the object of this study. The ones from Coimbra comprise two sets of samples: faiences and tiles. The ones from Lisbon are only tiles (azulejos, in Portuguese). The three main divisions of such pieces are the ceramic support (body), glaze and surface decoration. The system decoration/glaze is not easy to investigate, due to the high heterogeneity resulting from the mixing procedures and firing temperatures used. Moreover, the ability of the pigment to diffuse into the base glaze varies depending on the composition of both the pigment and the glaze in terms of fusibility. In order to raster these effects, high resolution techniques are required. In this work, synchrotron micro-X-ray fluorescence (µ-XRF) analysis was performed to monitor the profiles of the characteristic elements from the colors used and the glaze in well prepared cross-sections of the samples. Key elements are: Co for blue, Mn for purple, Cu for green, Sb for yellow, Pb for the glaze and Fe for the body. The major difference observed is that faiences have glaze thicknesses between 150 and 200 µm and tiles have glaze thicknesses between 350 and 400 µm. Furthermore, in faiences all the pigments except the yellow ones are well dispersed into the glassy matrix, while in tiles, all of them are just partially diffused throughout the glaze. However, differences between the tiles from Coimbra and Lisbon were observed. In the samples from Lisbon, a higher intake from the pigment throughout the glaze is observed. KW - Synchrotron micro-XRF KW - Compound refractive lenses (CRLs) KW - Portuguese glazed ceramics PY - 2012 DO - https://doi.org/10.1039/c2ja30030c SN - 0267-9477 SN - 1364-5544 VL - 27 IS - 6 SP - 966 EP - 974 PB - Royal Society of Chemistry CY - London AN - OPUS4-26789 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bjeoumikhov, A. A1 - Bjeoumikhova, Zemfira A1 - Riesemeier, Heinrich A1 - Radtke, Martin A1 - Wedell, R. T1 - Propagation of synchrotron radiation through nanocapillary structures N2 - The propagation of synchrotron radiation through nanocapillary structures with channel sizes of 200 nm and periods in the micrometer size has been studied experimentally. It was shown that the propagation through individual capillary channels has a mode formation character. Furthermore it was shown that during the propagation through capillary channels the coherence of synchrotron radiation is partially conserved. Interference of beams propagating through different capillary channels is observed which leads to a periodically modulated distribution of the radiation intensity in a plane far from the exit of the structure. These investigations are of high relevance for the understanding of X-ray transmission through nanocapillaries and the appearance of wave properties at this size scale. KW - Polycapillary structure KW - Nanocapillary KW - X-ray optics KW - Wave properties KW - Synchrotron radiation KW - X-ray diffraction PY - 2007 DO - https://doi.org/10.1016/j.physleta.2007.01.092 SN - 0031-9163 SN - 0375-9601 VL - 366 IS - 4-5 SP - 283 EP - 288 PB - North-Holland Publ. Co. CY - Amsterdam AN - OPUS4-20580 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -