TY - JOUR A1 - Urquiza, M.P. A1 - Torres, M.M. A1 - Mitani, Y. A1 - Schantz, M.M. A1 - Duewer, D.L. A1 - May, W.E. A1 - Parris, R.M. A1 - Wise, S. A. A1 - Kaminski, Katja A1 - Philipp, Rosemarie A1 - Win, Tin A1 - Rosso, A. A1 - Kim, D.H. A1 - Ishikawa, K. A1 - Krylov, A.I. A1 - Kustikov, Y.A. A1 - Baldan, A. T1 - Final report on CCQM-K47: Volatile organic compounds in methanol N2 - KEY COMPARISON At the October 2005 CCQM Organic Analysis Working Group Meeting (IRMM, Belgium), the decision was made to proceed with a Key Comparison study (CCQM-K47) addressing the calibration function for the determination of volatile organic compounds (VOCs) used for water quality monitoring. This was coordinated by CENAM and NIST. Benzene, o-xylene, m-xylene and p-xylene were chosen as representative VOCs. The solvent of choice was methanol. Key Comparison CCQM-K47 demonstrated the capabilities of participating NMIs to identify and measure the four target VOCs in a calibration solution using GC-based methods. The measurement challenges in CCQM-K47, such as avoiding volatility loss, achieving adequate chromatographic resolution and isolating potential interferences, are typical of those required for value-assigning volatile reference materials. Participants achieving comparable measurements for all four VOCs in this Key Comparison should be capable of providing reference materials and measurements for VOCs in solutions when present at concentration levels greater than 10 µg/g. PY - 2013 U6 - https://doi.org/10.1088/0026-1394/50/1A/08021 SN - 0026-1394 SN - 1681-7575 VL - 50 SP - 08021, 1 EP - 24 PB - Inst. of Physics Publ. CY - Bristol AN - OPUS4-31071 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -