TY - JOUR A1 - von der Au, Marcus A1 - Faßbender, Sebastian A1 - Chronakis, Michail A1 - Vogl, Jochen A1 - Meermann, Björn T1 - Size determination of nanoparticles by ICP-ToF-MS using isotope dilution in microdroplets N2 - Within this work, the combination of a microdroplet generator and an ICP-ToF-MS for nanoparticle analysis is presented. For the size determination of platinum nanoparticles an on-line isotope dilution analysis approach was developed. The 194Pt/195Pt isotopic ratio was used for the characterization of the particles, while the 182W/183W isotopic ratio was monitored simultaneously for mass bias correction. The on-line ID-MDG-sp-ICP-ToF-MS approach was deployed for the size determination of three platinum nanoparticle samples (50 nm, 63 nm, 70 nm); for validation, complementary size characterization techniques (sp-ICP-ToF-MS and TEM) were used. The robustness of this technique was evidenced, by using sodium chloride concentrations up to 100 mg L−1 as a matrix component. Our new on-line ID MDG-sp-ICP-ToF-MS approach is a promising tool for the fast and reliable determination of nanoparticles' size in severe matrix concentrations, e.g., environmental samples. KW - ICP-ToF-MS KW - Nanoparticles KW - Isotope Dilution PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-552727 DO - https://doi.org/10.1039/D2JA00072E SN - 0267-9477 VL - 37 IS - 6 SP - 1203 EP - 1207 PB - Royal Society of Chemistry AN - OPUS4-55272 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Chronakis, Michail Ioannis A1 - von der Au, Marcus A1 - Faßbender, Sebastian A1 - Vogl, Jochen A1 - Meermann, Björn T1 - MDG-ICP-ToF-MS: A Versatile Tool for Quantification in the field of Single Particle ICP-MS Using Isotope Dilution N2 - In the interest of exploring their potential in the field of single particle analysis, a Microdroplet Generator (MDG) was coupled to an ICP-ToF-MS. Isotopic Dilution Analysis was also incorporated for the size determination of three different Platinum nanoparticles samples (50, 63 and 70 nm). The performance of the technique was validated by comparison to traditional size characterization techniques (sp-ICP-ToF-MS, TEM), while the robustness of the technique was proven by incorporating NaCl in the samples’ matrix, up to 100 mg/L. T2 - European Winter Conference on Plasma Spectrochemistry CY - Ljubljana, Slovenia DA - 29.01.2023 KW - MDG KW - Single Particle KW - Isotope Dilution PY - 2023 AN - OPUS4-57643 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -