TY - JOUR A1 - Noble, J.E. A1 - Wang, L. A1 - Cerasoli, E. A1 - Knight, A.E. A1 - Porter, R.A. A1 - Gray, E. A1 - Howe, C. A1 - Hannes, E. A1 - Corbisier, P. A1 - Wang, J. A1 - Wu, L. A1 - Altieri, I. A1 - Patriarca, M. A1 - Hoffmann, Angelika A1 - Resch-Genger, Ute A1 - Ebert, B. A1 - Voigt, Jan A1 - Shigeri, Y. A1 - Vonsky, M.S. A1 - Konopelko, L.A. A1 - Gaigalas, A.K. A1 - Bailey, M. J. A. T1 - An international comparability study to determine the sources of uncertainty associated with a non-competitive sandwich fluorescent ELISA JF - Clinical chemistry and laboratory medicine KW - ELISA KW - Fluorescence KW - Interferon KW - Uncertainty KW - Round Robin KW - Immunoassay KW - Quality assurance KW - Fluorescein PY - 2008 SN - 1434-6621 SN - 1437-8523 VL - 46 IS - 7 SP - 1033 EP - 1045 PB - De Gruyter CY - Berlin AN - OPUS4-18283 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Hoffmann, Katrin A1 - Dietzel, B. A1 - Schulz, B. A1 - Reck, Günter A1 - Hoffmann, Angelika A1 - Orgzall, I. A1 - Resch-Genger, Ute A1 - Emmerling, Franziska T1 - Combined structural and fluorescence studies of methyl-substituted 2,5-diphenyl-1,3,4-oxadiazoles - relation between electronic properties and packing motifs JF - Journal of molecular structure N2 - Prerequisite for the rational design of functional organic materials with tailor-made electronic properties is the knowledge of the structure–property relationship for the specific class of molecules under consideration. This encouraged us to systematically study the influence of the molecular structure and substitution pattern of aromatically substituted 1,3,4-oxadiazoles on the electronic properties and packing motifs of these molecules and on the interplay of these factors. For this purpose, seven diphenyl-oxadiazoles equipped with methyl substituents in the ortho- and meta-position(s) were synthesized and characterized. Absorption and fluorescence spectra in solution served here as tools to monitor substitution-induced changes in the electronic properties of the individual molecules whereas X-ray and optical measurements in the solid state provided information on the interplay of electronic and packing effects. In solution, the spectral position of the absorption maximum, the size of Stokes shift, and the fluorescence quantum yield are considerably affected by ortho-substitution in three or four ortho-positions. This results in blue shifted absorption bands, increased Stokes shifts, and reduced fluorescence quantum yields whereas the spectral position and vibrational structure of the emission bands remain more or less unaffected. In the crystalline state, however, the spectral position and shape of the emission bands display a strong dependence on the molecular structure and/or packing motifs that seem to control the amount of dye–dye-interactions. These observations reveal the limited value of commonly reported absorption and fluorescence measurements in solution for a straightforward comparison of spectroscopic results with single X-ray crystallography. This underlines the importance of solid state spectroscopic studies for a better understanding of the interplay of electronic effects and molecular order. KW - Diphenyl-oxadiazoles KW - X-ray structure KW - Packing motif KW - Optical properties KW - Fluorescence quantum yield PY - 2011 DO - https://doi.org/10.1016/j.molstruc.2010.11.071 SN - 0022-2860 SN - 1872-8014 SN - 0377-046X VL - 988 IS - 1-3 SP - 35 EP - 46 PB - Elsevier CY - Amsterdam AN - OPUS4-23253 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Hennig, Andreas A1 - Dietrich, Paul A1 - Hemmann, Felix A1 - Thiele, T. A1 - Borcherding, H. A1 - Hoffmann, Angelika A1 - Schedler, U. A1 - Jäger, Christian A1 - Resch-Genger, Ute A1 - Unger, Wolfgang T1 - En route to traceable reference standards for surface group quantifications by XPS, NMR and fluorescence spectroscopy JF - Analyst N2 - The fluorine content of polymer particles labelled with 2,2,2-trifluoroethylamine was reliably quantified with overlapping sensitivity ranges by XPS and solid-state NMR. This provides a first step towards reference materials for the metrological traceability of surface group quantifications. The extension of this concept to fluorescence spectroscopy is illustrated. KW - Polymer beads KW - Poly(acrylic acid)-grafted particles KW - XPS KW - SEM KW - Surface analysis KW - Fluorescent dye labeling KW - NMR PY - 2015 DO - https://doi.org/10.1039/c4an02248c SN - 0003-2654 SN - 1364-5528 VL - 140 IS - 6 SP - 1804 EP - 1808 PB - Royal Society of Chemistry CY - Cambridge AN - OPUS4-32794 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - GEN A1 - Resch-Genger, Ute A1 - Pfeifer, Dietmar A1 - Hoffmann, Katrin A1 - Flachenecker, Günter A1 - Hoffmann, Angelika A1 - Monte, C. ED - Resch-Genger, Ute ED - O.S. Wolfbeis, T1 - Linking fluorometry to radiometry with physical and chemical transfer standards: instrument characterization and traceable fluorescence measurements T2 - Standardization and Quality Assurance in Fluorescence Measurements I - Techniques N2 - Problems associated with the measurement of photoluminescence are briefly reviewed, including relevant instrument parameters affecting these measurements. Procedures for the characterization of relevant instruments are discussed, focusing on spectrofluorometers, and fit-for-purpose methods including suitable standards are recommended. The aim here is to increase the awareness of the importance of reliable instrument characterization and to improve the comparability of measurements of photoluminescence. KW - Calibration KW - Emission standards KW - Fluorescence intensity standards KW - Fluorescence standards KW - Quality assurance KW - Spectral correction PY - 2008 SN - 978-3-540-75206-6 DO - https://doi.org/10.1007/4243_2008_054 SN - 1617-1306 N1 - Serientitel: Springer Series on Fluorescence – Series title: Springer Series on Fluorescence VL - 5 IS - Part II SP - 65 EP - 99 PB - Springer AN - OPUS4-18301 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Hennig, Andreas A1 - Hoffmann, Angelika A1 - Borcherding, H. A1 - Thiele, T. A1 - Schedler, U. A1 - Resch-Genger, Ute T1 - Quantification of surface functional groups on polymer microspheres by supramolecular host-guest interactions JF - Chemical communications N2 - We introduce a method to determine the number of accessible functional groups on a polymer microsphere surface based on the interaction between the macrocyclic host cucurbit[7]uril (CB7) and a guest reacted to the microsphere surface. After centrifugation, CB7 in the supernatant is quantified by addition of a fluorescent dye. The difference between added and detected CB7 affords the number of accessible surface functional groups. KW - Cucurbituril KW - Acridine orange KW - Fluorescence KW - Polymer surface KW - Surface modification KW - Quantification PY - 2011 DO - https://doi.org/10.1039/c1cc11692d SN - 0022-4936 SN - 0009-241x SN - 1359-7345 SN - 1364-548x VL - 47 IS - 27 SP - 7842 EP - 7844 PB - Royal Society of Chemistry CY - Cambridge AN - OPUS4-24034 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Hennig, Andreas A1 - Borcherding, H. A1 - Jäger, Christian A1 - Hatami, Soheil A1 - Würth, Christian A1 - Hoffmann, Angelika A1 - Hoffmann, Katrin A1 - Thiele, T. A1 - Schedler, U. A1 - Resch-Genger, Ute T1 - Scope and limitations of surface functional quantification methods: exploratory study with poly(acrylic acid)-grafted micro- and nanoparticles JF - Journal of the American chemical society N2 - The amount of grafted poly(acrylic acid) on poly(methyl methacrylate) micro- and nanoparticles was quantified by conductometry, 13C solid-state NMR, fluorophore labeling, a supramolecular assay based on high-affinity binding of cucurbit[7]uril, and two colorimetric assays based on toluidine blue and nickel complexation by pyrocatechol violet. The methods were thoroughly validated and compared with respect to reproducibility, sensitivity, and ease of use. The results demonstrate that only a small but constant fraction of the surface functional groups is accessible to covalent surface derivatization independently of the total number of surface functional groups, and different contributing factors are discussed that determine the number of probe molecules which can be bound to the polymer surface. The fluorophore labeling approach was modified to exclude artifacts due to fluorescence quenching, but absolute quantum yield measurements still indicate a major uncertainty in routine fluorescence-based surface group quantifications, which is directly relevant for biochemical assays and medical diagnostics. Comparison with results from protein labeling with streptavidin suggests a porous network of poly(acrylic acid) chains on the particle surface, which allows diffusion of small molecules (cutoff between 1.6 and 6.5 nm) into the network. KW - Polymers KW - Surface groups KW - Quantification KW - Fluorescence PY - 2012 DO - https://doi.org/10.1021/ja302649g SN - 0002-7863 SN - 1520-5126 VL - 134 IS - 19 SP - 8268 EP - 8276 PB - American Chemical Society CY - Washington, DC AN - OPUS4-26002 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Hennig, Andreas A1 - Hoffmann, Angelika A1 - Borcherding, H. A1 - Thiele, T. A1 - Schedler, U. A1 - Resch-Genger, Ute T1 - Simple colorimetric method for quantification of surface carboxy groups on polymer particles JF - Analytical chemistry N2 - We present a novel, simple, and fast colorimetric method to quantify the total number of carboxy groups on polymer microparticle and nanoparticle surfaces. This method exploits that small divalent transition metal cations (M2+ = Ni2+, Co2+, Cd2+) are efficiently bound to these surface functional groups, which allows their extraction by a single centrifugation step. Remaining M2+ in the supernatant is subsequently quantified spectrophotometrically after addition of the metal ion indicator pyrocatechol violet, for which Ni2+ was identified to be the most suitable transition metal cation. We demonstrate that the difference between added and detected M2+ is nicely correlated to the number of surface carboxy groups as determined by conductometry, thereby affording a validated measure for the trueness of this procedure. The variation coefficient of ~5% found in reproducibility studies underlines the potential of this novel method that can find conceivable applications for the characterization of different types of poly(carboxylic acid)-functionalized materials, e.g., for quality control by manufacturers of such materials. KW - Polymer surface KW - Complexometry KW - Quantification KW - Pyrocatechol violet PY - 2011 DO - https://doi.org/10.1021/ac2007619 SN - 0003-2700 SN - 1520-6882 VL - 83 IS - 12 SP - 4970 EP - 4974 PB - American Chemical Society CY - Washington, DC AN - OPUS4-23892 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Resch-Genger, Ute A1 - Hoffmann, Katrin A1 - Hoffmann, Angelika T1 - Standardization of Fluorescence Measurements - Criteria for the Choice of Suitable Standards and Approaches to Fit-for-Purpose Calibration Tools JF - Annals of the New York academy of sciences N2 - This report summarizes problems associated with the comparability of measurements of photoluminescence and procedures for the characterization of relevant instruments, focusing on physical and chemical fluorescence standards. To provide recommendations on selecting and using such standards, we derive general and scope-specific requirements and quality criteria for suitable devices and materials and briefly address metrological requirements linked to the realization of comparable measurements. Special emphasis is dedicated to liquid and solid chromophore-based fluorescence standards developed or currently tested by us. KW - Fluorescence standard KW - Fluorescence spectroscopy KW - Spectral correction KW - day-to-day intensity standard KW - Calibration KW - Metrology KW - Fluorescence microscopy PY - 2008 DO - https://doi.org/10.1196/annals.1430.018 SN - 0077-8923 SN - 1749-6632 SN - 0094-8500 VL - 1130 SP - 35 EP - 43 PB - New York Academy of Sciences CY - New York, NY AN - OPUS4-17712 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Resch-Genger, Ute A1 - Bremser, Wolfram A1 - Pfeifer, Dietmar A1 - Spieles, Monika A1 - Hoffmann, Angelika A1 - DeRose, P.C. A1 - Zwinkels, J. C. A1 - Gauthier, F. A1 - Ebert, B. A1 - Taubert, D. A1 - Monte, C. A1 - Voigt, J. A1 - Hollandt, J. A1 - Macdonald, R. T1 - State-of-the art comparability of corrected emission spectra. 1. spectral correction with physical transfer standards and spectral fluorescence standards by expert laboratories JF - Analytical chemistry N2 - The development of fluorescence applications in the life and material sciences has proceeded largely without sufficient concern for the measurement uncertainties related to the characterization of fluorescence instruments. In this first part of a two-part series on the state-of-the-art comparability of corrected emission spectra, four National Metrology Institutes active in high-precision steady-state fluorometry performed a first comparison of fluorescence measurement capabilities by evaluating physical transfer standard (PTS)-based and reference material (RM)-based calibration methods. To identify achievable comparability and sources of error in instrument calibration, the emission spectra of three test dyes in the wavelength region from 300 to 770 nm were corrected and compared using both calibration methods. The results, obtained for typical spectrofluorometric (0°/90° transmitting) and colorimetric (45°/0° front-face) measurement geometries, demonstrated a comparability of corrected emission spectra within a relative standard uncertainty of 4.2% for PTS- and 2.4% for RM-based spectral correction when measurements and calibrations were performed under identical conditions. Moreover, the emission spectra of RMs F001 to F005, certified by BAM, Federal Institute for Materials Research and Testing, were confirmed. These RMs were subsequently used for the assessment of the comparability of RM-based corrected emission spectra of field laboratories using common commercial spectrofluorometers and routine measurement conditions in part 2 of this series (subsequent paper in this issue). KW - Fluorescence KW - Photoluminescence KW - Dye KW - Uncertainty KW - Method comparison KW - Standard KW - Method comparison KW - Spectral correction KW - Spectral fluorescence standard PY - 2012 DO - https://doi.org/10.1021/ac2034503 SN - 0003-2700 SN - 1520-6882 VL - 84 IS - 9 SP - 3889 EP - 3898 PB - American Chemical Society CY - Washington, DC AN - OPUS4-25838 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Resch-Genger, Ute A1 - Bremser, Wolfram A1 - Pfeifer, Dietmar A1 - Spieles, Monika A1 - Hoffmann, Angelika A1 - DeRose, P.C. A1 - Zwinkels, J. C. A1 - Gauthier, F. A1 - Ebert, B. A1 - Taubert, R.D. A1 - Voigt, J. A1 - Hollandt, J. A1 - Macdonald, R. T1 - State-of-the art comparability of corrected emission spectra. 2. Field laboratory assessment of calibration performance using spectral fluorescence standards JF - Analytical chemistry N2 - In the second part of this two-part series on the state-of-the-art comparability of corrected emission spectra, we have extended this assessment to the broader community of fluorescence spectroscopists by involving 12 field laboratories that were randomly selected on the basis of their fluorescence measuring equipment. These laboratories performed a reference material (RM)-based fluorometer calibration with commercially available spectral fluorescence standards following a standard operating procedure that involved routine measurement conditions and the data evaluation software LINKCORR developed and provided by the Federal Institute for Materials Research and Testing (BAM). This instrument-specific emission correction curve was subsequently used for the determination of the corrected emission spectra of three test dyes, X, QS, and Y, revealing an average accuracy of 6.8% for the corrected emission spectra. This compares well with the relative standard uncertainties of 4.2% for physical standard-based spectral corrections demonstrated in the first part of this study (previous paper in this issue) involving an international group of four expert laboratories. The excellent comparability of the measurements of the field laboratories also demonstrates the effectiveness of RM-based correction procedures. KW - Fluorescence KW - Photoluminescence KW - Dye KW - Uncertainty KW - Method comparison KW - Standard KW - Method comparison KW - Spectral correction KW - Spectral fluorescence standard PY - 2012 DO - https://doi.org/10.1021/ac203451g SN - 0003-2700 SN - 1520-6882 VL - 84 IS - 9 SP - 3899 EP - 3907 PB - American Chemical Society CY - Washington, DC AN - OPUS4-27081 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -