TY - JOUR A1 - Traub, Heike A1 - Wälle, M. A1 - Koch, J. A1 - Panne, Ulrich A1 - Matschat, Ralf A1 - Kipphardt, Heinrich A1 - Günther, D. T1 - Evaluation of different calibration strategies for the analysis of pure copper and zinc samples using femtosecond laser ablation ICP-MS N2 - Solution-doped metal powder pellets as well as aspirated liquids were used as calibration samples to analyze pure copper and zinc certified reference materials (CRMs) by femtosecond laser ablation ICP-MS. It was demonstrated that calibration by copper pellets resulted in relative deviations up to 20%, whereas fs-LA-ICP-MS among copper-based CRMs led to inaccuracies in the same range unless nominal mass fractions were chosen to be <3 mg/kg. Calibration by zinc pellets generally provided better accuracy. Depending on the analyte considered, deviations below 10% were obtained even for mass fractions close to the limit of quantification. Our data, therefore, indicate solution-doped metal powder pellets to be suitable as calibration samples for fs-LA-ICP-MS of metals. Furthermore, the utilization of liquid standards for calibration was found to result in stronger deviations of up to 50% for both copper and zinc samples which, in addition, turned out to be dependent on the plasma conditions. KW - Laser ablation KW - ICP-MS KW - Calibration KW - Copper KW - Zinc KW - Metal PY - 2009 U6 - https://doi.org/10.1007/s00216-009-3061-9 SN - 1618-2642 SN - 1618-2650 VL - 395 IS - 5 SP - 1471 EP - 1480 PB - Springer CY - Berlin AN - OPUS4-20402 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -