TY - JOUR A1 - Hein, Sebastian A1 - Philipp, Rosemarie A1 - Duewer, D.L. A1 - Gasca-Aragon, H. A1 - Lippa, K. A1 - Toman, B. A1 - Bremser, Wolfram A1 - Calderon, M.A.Á. A1 - Castorena, A.P. A1 - Caixeiro, J.M.R. A1 - Do Rego, E.C.P. A1 - Rosso, A. A1 - De Arechavaleta, R. A1 - O'Connor, G. A1 - Holcombe, G. A1 - Ellison, S.L.R. A1 - Splett, J.D. A1 - Schantz, M.M. A1 - Sharpless, K.E. A1 - Croft, M. A1 - Archer, M. A1 - De Vos, J. A1 - Konopelko, L.A. A1 - Kustikov, Y. T1 - Final report on CCQM-K79: Comparison of value-assigned CRMs and PT materials: Ethanol in aqueous matrix N2 - The 2010 CCQM-K79 'Comparison of value-assigned CRMs and PT materials: Ethanol in aqueous media' is the second key comparison directly testing the chemical measurement services provided to customers by National Metrology Institutes (NMIs) and Designated Institutes (DIs). CCQM-K79 compared the assigned ethanol values of proficiency test (PT) and certified reference materials (CRMs) using measurements made on these materials under repeatability conditions. Nine NMIs submitted 27 CRM or value-assigned PT materials for evaluation. These materials represent many of the higher-order reference materials then available for this commercially and forensically important measurand. The assigned ethanol mass fraction in the materials ranged from 0.1 mg/kg to 334 mg/kg. All materials were stored and prepared according the specifications provided by each NMI. Samples were processed and analyzed under repeatability conditions by one analytical team using a gas chromatography with flame ionization detection (GC-FID) method of demonstrated trueness and precision. Given the number of materials and the time required for each analysis, the majority of the measurements were made in two measurement campaigns ('runs'). Due to a shipping delay from one NMI, an unanticipated third campaign was required. In all three campaigns, replicate analyses (three injections of one preparation separated in time) were made for one randomly selected unit of each of the 27 materials. Nine of the 27 materials were gravimetrically diluted before measurement to provide solutions with ethanol mass fraction in the established linear range of the GC-FID method. The repeatability measurement value for each analyzed solution was estimated as the mean of all replicate values. The within- and between-campaign variance components were estimated using one-way ANOVA. Markov Chain Monte Carlo Bayesian analysis was used to estimate 95% level-of-confidence coverage intervals for the mean values. Uncertainty-weighted generalized distance regression was used to establish the key comparison reference function (KCRF) relating the assigned values to the repeatability measurements. On the basis of leave-one-out cross-validation, all of the assigned values for all 27 materials were deemed equivalent at the 95% level of confidence. These materials were used to define the KCRF. Parametric bootstrap Monte Carlo was used to estimate 95% level-of-confidence coverage intervals for the degrees of equivalence of materials, d ± U95(d), and of the participating NMIs, D ± U95(D). Because of the very wide range of ethanol mass fraction in the materials, these degrees of equivalence are expressed in percent relative form: %d ± U95(%d) and %D ± U95(%D). The median of the absolute values of the %D for the participating NMIs is less than 0.05% with a median U95(%D) of less than 1%. These results demonstrate that the participating NMIs have the ability to correctly value-assign CRMs and proficiency test materials for ethanol in aqueous media and similar measurands. PY - 2013 U6 - https://doi.org/10.1088/0026-1394/50/1A/08005 SN - 0026-1394 SN - 1681-7575 VL - 50 IS - 1A / Technical Suppl. 2013 / 08005 SP - 1 EP - 49 PB - Inst. of Physics Publ. CY - Bristol AN - OPUS4-27730 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Urquiza, M.P. A1 - Torres, M.M. A1 - Mitani, Y. A1 - Schantz, M.M. A1 - Duewer, D.L. A1 - May, W.E. A1 - Parris, R.M. A1 - Wise, S. A. A1 - Kaminski, Katja A1 - Philipp, Rosemarie A1 - Win, Tin A1 - Rosso, A. A1 - Kim, D.H. A1 - Ishikawa, K. A1 - Krylov, A.I. A1 - Kustikov, Y.A. A1 - Baldan, A. T1 - Final report on CCQM-K47: Volatile organic compounds in methanol N2 - KEY COMPARISON At the October 2005 CCQM Organic Analysis Working Group Meeting (IRMM, Belgium), the decision was made to proceed with a Key Comparison study (CCQM-K47) addressing the calibration function for the determination of volatile organic compounds (VOCs) used for water quality monitoring. This was coordinated by CENAM and NIST. Benzene, o-xylene, m-xylene and p-xylene were chosen as representative VOCs. The solvent of choice was methanol. Key Comparison CCQM-K47 demonstrated the capabilities of participating NMIs to identify and measure the four target VOCs in a calibration solution using GC-based methods. The measurement challenges in CCQM-K47, such as avoiding volatility loss, achieving adequate chromatographic resolution and isolating potential interferences, are typical of those required for value-assigning volatile reference materials. Participants achieving comparable measurements for all four VOCs in this Key Comparison should be capable of providing reference materials and measurements for VOCs in solutions when present at concentration levels greater than 10 µg/g. PY - 2013 U6 - https://doi.org/10.1088/0026-1394/50/1A/08021 SN - 0026-1394 SN - 1681-7575 VL - 50 SP - 08021, 1 EP - 24 PB - Inst. of Physics Publ. CY - Bristol AN - OPUS4-31071 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Westwood, S. A1 - Josephs, R. A1 - Choteau, T. A1 - Daireaux, A. A1 - Wielgosz, R. A1 - Davies, S. A1 - Moad, M. A1 - Chan, B. A1 - Munoz, A. A1 - Conneely, P. A1 - Ricci, M. A1 - Do Rego, E.C.P. A1 - Garrido, B.C. A1 - Violante, F.G.M. A1 - Windust, A. A1 - Dai, X. A1 - Huang, T. A1 - Zhang, W. A1 - Su, F. A1 - Quan, C. A1 - Wang, H. A1 - Lo, M. A1 - Wong, W. A1 - Gantois, F. A1 - Lalerle, B. A1 - Dorgerloh, Ute A1 - Koch, Matthias A1 - Klyk-Seitz, Urszula-Anna A1 - Pfeifer, Dietmar A1 - Philipp, Rosemarie A1 - Piechotta, Christian A1 - Recknagel, Sebastian A1 - Rothe, Robert A1 - Yamazaki, T. A1 - Zakaria, O. B. A1 - Castro, E. A1 - Balderas, M. A1 - González, N. A1 - Salazar, C. A1 - Regalado, L. A1 - Valle, E. A1 - Rodríguez, L. A1 - Laguna, L.Á.. A1 - Ramírez, P. A1 - Avila, M. A1 - Ibarra, J. A1 - Valle, L. A1 - Arce, M. A1 - Mitani, Y. A1 - Konopelko, L. A1 - Krylov, A. A1 - Lopushanskaya, E. A1 - Lin, T.T. A1 - Liu, Q. A1 - Kooi, L.T. A1 - Fernandes-Whaley, M. A1 - Prevoo-Franzsen, D. A1 - Nhlapo, N. A1 - Visser, R. A1 - Kim, B. A1 - Lee, H. A1 - Kankaew, P. A1 - Pookrod, P. A1 - Sudsiri, N. A1 - Shearman, K. A1 - Gören, A.C. A1 - Bilsel, G. A1 - Yilmaz, H. A1 - Bilsel, M. A1 - Cergel, M. A1 - Coskun, F.G. A1 - Uysal, E. A1 - Gündüz, S. A1 - Ün, I. A1 - Warren, J. A1 - Bearden, D.W. A1 - Bedner, M. A1 - Duewer, D.L. A1 - Lang, B.E. A1 - Lippa, K.A. A1 - Schantz, M.M. A1 - Sieber, J.R. T1 - Final report on key comparison CCQM-K55.c (L-(+)-Valine): Characterization of organic substances for chemical purity N2 - KEY COMPARISON Under the auspices of the Organic Analysis Working Group (OAWG) of the Comité Consultatif pour la Quantité de Matière (CCQM) a key comparison, CCQM K55.c, was coordinated by the Bureau International des Poids et Mesures (BIPM) in 2012. Twenty National Measurement Institutes or Designated Institutes and the BIPM participated. Participants were required to assign the mass fraction of valine present as the main component in the comparison sample for CCQM-K55.c. The comparison samples were prepared from analytical grade L-valine purchased from a commercial supplier and used as provided without further treatment or purification. Valine was selected to be representative of the performance of a laboratory's measurement capability for the purity assignment of organic compounds of low structural complexity [molecular weight range 100–300] and high polarity (pKOW > –2). The KCRV for the valine content of the material was 992.0 mg/g with a combined standard uncertainty of 0.3 mg/g. The key comparison reference value (KCRV) was assigned by combination of KCRVs assigned from participant results for each orthogonal impurity class. The relative expanded uncertainties reported by laboratories having results consistent with the KCRV ranged from 1 mg/g to 6 mg/g when using mass balance based approaches alone, 2 mg/g to 7 mg/g using quantitative 1H NMR (qNMR) based approaches and from 1 mg/g to 2.5 mg/g when a result obtained by a mass balance method was combined with a separate qNMR result. The material provided several analytical challenges. In addition to the need to identify and quantify various related amino acid impurities including leucine, isoleucine, alanine and a-amino butyrate, care was required to select appropriate conditions for performing Karl Fischer titration assay for water content to avoid bias due to in situ formation of water by self-condensation under the assay conditions. It also proved to be a challenging compound for purity assignment by qNMR techniques. There was overall excellent agreement between participants in the identification and the quantification of the total and individual related structure impurities, water content, residual solvent and total non-volatile content of the sample. Appropriate technical justifications were developed to rationalise observed discrepancies in the limited cases where methodology differences led to inconsistent results. The comparison demonstrated that to perform a qNMR purity assignment the selection of appropriate parameters and an understanding of their potential influence on the assigned value is critical for reliable implementation of the method, particularly when one or more of the peaks to be quantified consist of complex multiplet signals. PY - 2014 U6 - https://doi.org/10.1088/0026-1394/51/1A/08010 SN - 0026-1394 SN - 1681-7575 VL - 51 SP - 08010, 1 EP - 44 PB - Inst. of Physics Publ. CY - Bristol AN - OPUS4-31072 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Zeleny, R. A1 - Voorspoels, S. A1 - Ricci, M. A1 - Becker, Roland A1 - Jung, Christian A1 - Bremser, Wolfram A1 - Sittidech, M. A1 - Panyawathanakit, N. A1 - Wong, W. F. A1 - Choi, S.M. A1 - Lo, K.C. A1 - Yeung, W. Y. A1 - Kim, D.H. A1 - Han, J. A1 - Ryu, J. A1 - Mingwu, S. A1 - Chao, W. A1 - Schantz, M.M. A1 - Lippa, K.A. A1 - Matsuyama, S. T1 - Evaluation of the state-of-the-art measurement capabilities for selected PBDEs and decaBB in plastic by the international intercomparison CCQM-P114 N2 - An international intercomparison involving eight national metrology institutes (NMIs) was conducted to establish their current measurement capabilities for determining five selected congeners from the brominated flame retardant classes polybrominated diphenyl ethers and polybrominated biphenyls. A candidate reference material consisting of polypropylene fortified with technical mixtures of penta-, octa- and decabromo diphenyl ether and decabromo biphenyl, which was thoroughly assessed for material homogeneity and stability, was used as study material. The analytical procedures applied by the participants differed with regard to sample pre-treatment, extraction, clean-up, employed calibrants and type of calibration procedure as well as regarding analytical methods used for separation, identification and quantification of the flame retardant congeners (gas chromatography coupled to an electron capture detector (GC-ECD), gas chromatography-mass spectrometry in the electron ionisation mode (GC-EI-MS), gas chromatography-mass spectrometry in the electron capture negative ionisation mode (GC-ECNI-MS), and liquid chromatography-inductive coupled plasma-mass spectrometry (LC-ICP-MS)). The laboratory means agreed well with relative standard deviations of the mean of means of 1.9%, 4.8%, 5.5% and 5.4% for brominated diphenyl ether (BDE) 47, 183 and 209 and for the brominated biphenyl (BB) congener 209, respectively. For BDE 206, a relative standard deviation of 28.5% was obtained. For all five congeners, within-laboratory relative standard deviations of six measurements obtained under intermediate precision conditions were between 1% and 10%, and reported expanded measurements uncertainties typically ranged from 4% to 10% (8% to 14% for BDE 206). Furthermore, the results are in good agreement with those obtained in the characterization exercise for determining certified values for the flame retardant congeners in the same material. The results demonstrate the state-of-the-art measurement capabilities of NMIs for quantifying representative BDE congeners and BB 209 in a polymer. The outcome of this intercomparison (pilot study) in conjunction with possible improvements for employing exclusively calibrants with thoroughly assessed purity suggests that a key comparison aiming at underpinning calibration and measurement capability (CMC) claims of NMIs can be conducted. KW - Flame retardants KW - Polymer KW - Polybrominated diphenyl ethers (PBDEs) KW - Polybrominated biphenyls (PBBs) KW - International intercomparison PY - 2010 U6 - https://doi.org/10.1007/s00216-009-3314-7 SN - 1618-2642 SN - 1618-2650 VL - 396 IS - 4 SP - 1501 EP - 1511 PB - Springer CY - Berlin AN - OPUS4-22905 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Duewer, D.L. A1 - May, W.E. A1 - Parris, R.M. A1 - Schantz, M.M. A1 - Wise, S. A. A1 - Piechotta, Christian A1 - Philipp, Rosemarie A1 - Win, Tin A1 - Avila, M. A1 - Pérez Urquiza, M. A1 - Ulberth, F. A1 - Kim, B. A1 - Ishikawa, K. A1 - Chen, D. A1 - Krylov, A.I. A1 - Kustidov, Y.A. A1 - Lopushanskaya, E.M. T1 - Final report on CCQM-K38: Determination of PAHs in solution PY - 2009 U6 - https://doi.org/10.1088/0026-1394/46/1A/08003 SN - 0026-1394 SN - 1681-7575 VL - 46 IS - 08003 SP - 1 EP - 13 PB - Inst. of Physics Publ. CY - Bristol AN - OPUS4-20029 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Westwood, S. A1 - Josephs, R. A1 - Daireaux, A. A1 - Wielgosz, R. A1 - Davies, S. A1 - Kang, M. A1 - Ting, H. A1 - Philipp, Rosemarie A1 - Malz, Frank A1 - Shimizu, Y. A1 - Frias, E. A1 - Pérez, M. A1 - Apps, P. A1 - Fernandes-Whaley, M. A1 - De Vos, B. A1 - Wiagnon, K. A1 - Ruangrittinon, N. A1 - Wood, S. A1 - Duewer, D.L. A1 - Schantz, M.M. A1 - Bedner, M. A1 - Hancock, D. A1 - Esker, J. T1 - An international comparison of mass fraction purity assignment of theophylline: CCQM pilot study CCQM-P20.e (Theophylline) PY - 2009 U6 - https://doi.org/10.1088/0026-1394/46/1A/08019 SN - 0026-1394 SN - 1681-7575 VL - 46 IS - 08019 SP - 1 EP - 37 PB - Inst. of Physics Publ. CY - Bristol AN - OPUS4-20030 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -