TY - JOUR A1 - Höhse, Marek A1 - Gornushkin, Igor B. A1 - Merk, Sven A1 - Panne, Ulrich T1 - Assessment of suitability of diode pumped solid state lasers for laser induced breakdown and Raman spectroscopy N2 - The potential of a diode pumped solid state (DPSS) laser for laser induced breakdown spectroscopy (LIBS) and Raman spectroscopy was investigated. The DPSS laser operating at repetition rates of 1 Hz–200 kHz, pulse energy of ~1 mJ, and pulse duration of ~20 ns was added to a conventional LIBS system equipped with the flashlamp-pumped Nd:YAG laser (10 Hz, 400 mJ, 6 ns) to directly compare the two laser systems. Despite the lower mass ablated per laser pulse (~several nanograms) and a weaker plasma, the total mass ablated per measurement interval (~1 s) and total emission signal were significantly higher than those obtained with the flashlamp pumped Nd:YAG laser. This resulted in the improved signal-to-noise ratio and better limits of detection (LOD). The best LODs were obtained for Cu, Cr and Ni in ultrapure standard iron samples, 0.7 ppm, 1 ppm and 5 ppm, correspondingly. The plasma from the DPSS laser was characterized in terms of its radiative properties, temperature, and shockwave propagation speed. The suitability of this laser for LIBS-Raman was demonstrated by mapping the elemental and molecular composition of a mineral sample. The results strongly imply that DPSS lasers are the promising source for LIBS, Raman, or combined LIBS-Raman spectroscopy. PY - 2011 DO - https://doi.org/10.1039/c0ja00038h SN - 0267-9477 SN - 1364-5544 VL - 26 IS - 2 SP - 414 EP - 424 PB - Royal Society of Chemistry CY - London AN - OPUS4-22943 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Herwig, Nadine A1 - Stephan, Kristin A1 - Panne, Ulrich A1 - Pritzkow, Wolfgang A1 - Vogl, Jochen T1 - Multi-element screening in milk and feed by SF-ICP-MS N2 - In this study, an analytical procedure for multi-element screening of 40 elements in milk and feed samples was developed. Three different digestion and two different calibration methods were tested for the best suitability. The analytical procedure for the quantification of minor and trace elements is based on sector field ICP-MS (SF-ICP-MS). The method validation revealed good agreement between the determined elemental mass fractions and the certified values of two milk and three feed reference materials. Milk samples of dairy cows at different feeding regimes were collected from two different farms. The results of our study showed significant differences of nine elemental mass fractions (Li, P, Mn, Co, Cu, Rb, Sr, Br, I) between milk from the two locations. Further, a correlation between the elemental mass fractions of milk and the ingested feed and water was observed. KW - Multi-element screening KW - SF-ICP-MS KW - Dairy cows KW - Milk KW - Feed KW - Water PY - 2011 DO - https://doi.org/10.1016/j.foodchem.2010.07.050 SN - 0308-8146 VL - 124 IS - 3 SP - 1223 EP - 1230 PB - Elsevier Ltd. CY - Amsterdam [u.a.] ; Jena AN - OPUS4-22042 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Merk, Virginia A1 - Schulte, Franziska A1 - Rooch, Heidemarie A1 - Feldmann, Ines A1 - Dörfel, Ilona A1 - Österle, Werner A1 - Panne, Ulrich A1 - Kneipp, Janina T1 - Surface-enhanced Raman scattering with silver nanostructures generated in situ in a sporopollenin biopolymer matrix N2 - Silver nanoparticles were generated based on citrate reduction in the ultrastructure of the sporopollenin biopolymer of Ambrosia artemisiifolia (ragweed) and Secale cereale (rye). The nanoparticles enable the acquisition of SERS spectra and thereby a vibrational characterization of the local molecular structure of sporopollenin. PY - 2011 DO - https://doi.org/10.1039/c0cc05326k SN - 0022-4936 SN - 0009-241x SN - 1359-7345 SN - 1364-548x N1 - Geburtsname von Merk, Virginia: Joseph, V. - Birth name of Merk, Virginia: Joseph, V. VL - 47 IS - 11 SP - 3236 EP - 3238 PB - Royal Society of Chemistry CY - Cambridge AN - OPUS4-23482 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bahlmann, Arnold A1 - Falkenhagen, Jana A1 - Weller, Michael G. A1 - Panne, Ulrich A1 - Schneider, Rudolf T1 - Cetirizine as pH-dependent cross-reactant in a carbamazepine-specific immunoassay N2 - High performance liquid chromatography (HPLC) was hyphenated with a previously reported carbamazepine-specific enzyme-linked immunosorbent assay (ELISA) as a screening approach to water analysis in order to identify possible interferences from transformation products. Treated wastewater was analysed and three substances were recognized by the antibody besides carbamazepine: the metabolites 10,11-dihydro-10,11-epoxycarbamazepine and 2-hydroxycarbamazepine plus the structurally not obviously related antihistamine cetirizine. The molar cross-reactivity against cetirizine was found to be pH-dependent and assessed to be 400% at pH 4.5 and 22% at pH 10.5. Performing the ELISA at pH 10.5 greatly improved the accuracy when carbamazepine was determined in surface and wastewater samples. KW - Immunoassay KW - ELISA KW - Cetirizin KW - Carbamazepin KW - pH-Abhängigkeit KW - Kreuzreaktivität PY - 2011 DO - https://doi.org/10.1039/c0an00928h SN - 0003-2654 SN - 1364-5528 VL - 136 IS - 7 SP - 1357 EP - 1364 PB - Royal Society of Chemistry CY - Cambridge AN - OPUS4-23546 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Weidner, Steffen A1 - Knappe, Patrick A1 - Panne, Ulrich T1 - MALDI-TOF imaging mass spectrometry of artifacts in 'dried droplet' polymer samples N2 - Matrix-assisted laser desorption/ionization-timeof-flight (MALDI-TOF) imaging of polystyrenes with various molecular masses was applied to study spatial molecular mass distribution of polymers in sample spots prepared by the 'dried droplet' method. When different solvents and target surfaces were examined, a segregation of single homologous polymers was observed depending upon the evaporation rate of the solvent. For the observed Patterns left by the evaporating droplet, a hypothesis is offered taking into account different hydrodynamic interactions and diffusion. The results illustrate that spot preparation using the conventionally 'dried droplet' method is prone to artifacts and should be avoided for reliable and reproducible MALDI mass spectrometry experiments with regards to the Determination of molecular masses and mass distributions. KW - MALDI KW - Mass spectrometry imaging KW - Polymer KW - Droplet PY - 2011 DO - https://doi.org/10.1007/s00216-011-4773-1 SN - 1618-2642 SN - 1618-2650 VL - 401 IS - 1 SP - 127 EP - 134 PB - Springer CY - Berlin AN - OPUS4-24538 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Walter, Astrid A1 - Panne, Ulrich A1 - Weller, Michael G. T1 - A novel immunoreagent for the specific and sensitive detection of the explosive triacetone triperoxide (TATP) N2 - Triacetone triperoxide (TATP) is a primary explosive, which was used in various terrorist attacks in the past. For the development of biosensors, immunochemical µ-TAS, electronic noses, immunological test kits, or test strips, the availability of antibodies of high quality is crucial. Recently, we presented the successful immunization of mice, based on the design, synthesis, and conjugation of a novel TATP derivative. Here, the long-term immunization of rabbits is shown, which resulted in antibodies of extreme selectivity and more than 1,000 times better affinity in relation to the antibodies from mice. Detection limits below 10 ng L-1 (water) were achieved. The working range covers more than four decades, calculated from a precision profile. The cross-reactivity tests revealed an extraordinary selectivity of the antibodies—not a single compound could be identified as a relevant cross-reactant. The presented immunoreagent might be a major step for the development of highly sensitive and selective TATP detectors particularly for security applications. KW - Organic peroxides KW - Terrorism KW - Biosensor development PY - 2011 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-240291 UR - http://www.mdpi.com/2079-6374/1/3/93 DO - https://doi.org/10.3390/bios1030093 SN - 2079-6374 VL - 1 IS - 3 SP - 93 EP - 106 PB - MDPI CY - Basel AN - OPUS4-24029 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Moreno-Gordaliza, E. A1 - Giesen, Charlotte A1 - Lázaro, A. A1 - Esteban-Fernández, D. A1 - Humanes, B. A1 - Canas, B. A1 - Panne, Ulrich A1 - Tejedor, A. A1 - Jakubowski, Norbert A1 - Gómez-Gómez, M.M. T1 - Elemental bioimaging in kidney by LA-ICP-MS as a tool to study nephrotoxicity and renal protective strategies in cisplatin therapies N2 - A laser ablation inductively coupled plasma mass spectrometry (LA–ICP–MS)-based methodology is presented for Pt, Cu, and Zn bioimaging on whole kidney 3 µm sagittal sections from rats treated with pharmacological doses of cisplatin, which were sacrificed once renal damage had taken place. Pt turned out to accumulate in the kidney cortex and corticomedullary junction, corresponding to areas where the proximal tubule S3 segments (the most sensitive cells to cisplatin nephrotoxicity) are located. This demonstrates the connection between platinum accumulation and renal damage proved by histological examination of HE-stained sections and evaluation of serum and urine biochemical parameters. Cu and Zn distribution maps revealed a significant displacement in cells by Pt, as compared to control tissues. A dramatic decrease in the Pt accumulation in the cortex was observed when cilastatin was coadministered with cisplatin, which can be related to its nephroprotective effect. Excellent imaging reproducibility, sensitivity (LOD 50 fg), and resolution (down to 8 µm) were achieved, demonstrating that LA–ICP–MS can be applied as a microscopic metal detector at cellular level in certain tissues. A simple and quick approach for the estimation of Pt tissue levels was proposed, based on tissue spiking. PY - 2011 DO - https://doi.org/10.1021/ac201933x SN - 0003-2700 SN - 1520-6882 VL - 83 IS - 20 SP - 7933 EP - 7940 PB - American Chemical Society CY - Washington, DC AN - OPUS4-24965 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Merk, Sven A1 - Shabanov, Sergej V. A1 - Gornushkin, Igor B. A1 - Panne, Ulrich T1 - Laser-induced plasma tomography by the Radon transform N2 - The Radon transform is tested as a method for reconstruction of the emissivity distribution of asymmetric laser induced plasmas. Two types of experiments were carried out. First, the plasma asymmetry is introduced via focusing the laser by a cylindrical lens to create plasma plumes elongated along the symmetry axis of the lens. Second, an asymmetric power distribution across the laser beam is created by reflecting the latter from a damaged mirror. Various effects on the quality of the plasma emissivity reconstructed by the Radon tomography method are investigated. The understanding of these effects appears to be essential to design a proper experimental setup to study LIBS plasmas by the Radon tomography method. It is demonstrated that the Radon tomography can successfully be used for experimental studies of asymmetric LIBS plasmas. KW - Radon transform KW - Laser induced plasma KW - LIBS KW - Plasma tomography PY - 2011 DO - https://doi.org/10.1039/c1ja10187k SN - 0267-9477 SN - 1364-5544 VL - 26 SP - 2483 EP - 2492 PB - Royal Society of Chemistry CY - London AN - OPUS4-24889 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Giesen, Charlotte A1 - Mairinger, T. A1 - Khoury, L. A1 - Müller, Larissa A1 - Jakubowski, Norbert A1 - Panne, Ulrich T1 - Multiplexed immunohistochemical detection of tumor markers in breast cancer tissue using laser ablation inductively coupled plasma mass spectrometry N2 - We optimized multiplexed immunohistochemistry (IHC) on breast cancer tissue. Up to 20 tumor markers are routinely evaluated for one patient, and thus, a common analysis results in a series of time consuming staining procedures. As an alternative, we used lanthanides for labeling of primary antibodies, which are applied in IHC. Laser ablation (LA) ICPMS was elaborated as a detection tool for multiplexed IHC of tissue sections. In this study, we optimized sample preparation steps and LA ICPMS parameters to achieve a sufficient signal-to-background ratio. The results prove the high selectivity of applied antibodies, which was sustained after labeling. Up to three tumor markers (Her 2, CK 7, and MUC 1) were detected simultaneously in a single multiplex analysis of a 5 µm thin breast cancer tissue at a laser spot size of 200 µm. Furthermore, the LA ICPMS results indicate a significantly higher expression level of MUC 1 compared to Her 2 and CK 7, which was not obvious from the conventionally stained tissue sections. PY - 2011 DO - https://doi.org/10.1021/ac2016823 SN - 0003-2700 SN - 1520-6882 N1 - Geburtsname von Müller, Larissa: Wäntig, L. - Birth name of Müller, Larissa: Wäntig, L. VL - 83 IS - 21 SP - 8177 EP - 8183 PB - American Chemical Society CY - Washington, DC AN - OPUS4-24966 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Giesen, Charlotte A1 - Müller, Larissa A1 - Mairinger, T. A1 - Drescher, Daniela A1 - Kneipp, Janina A1 - Roos, P.H. A1 - Panne, Ulrich A1 - Jakubowski, Norbert T1 - Iodine as an elemental marker for imaging of single cells and tissue sections by laser ablation inductively coupled plasma mass spectrometry N2 - A new laser ablation (LA)-ICP-MS method for single cell and cell nucleus imaging was developed. Therein, iodine was employed as an elemental dye for fibroblast cells and for thin tissue sections. At an incubation time of 60 s, iodine is located mainly within the cell nuclei. This effect was illustrated in fibroblast cells, and iodine signal within the cell nucleus was as high as 5 × 104 cps at 4 µm laser spot size. The surrounding cytoplasm was iodinated as well, but to a lesser extent. The spatial resolution attained was sufficient to detect even smaller cell nuclei within a liver biopsy tissue. Furthermore, iodine was successfully employed for biomolecule labeling and we demonstrated that iodine signal increased with increasing thickness of a palatine tonsil tissue. Thus, the use of iodine as an internal standard to correct for tissue inhomogeneities in LA-ICP-MS was investigated for the simultaneous detection of two tumor markers (Her 2 and CK 7) in breast cancer tissue. Additionally, lanthanide background resulting from glass ablation can be corrected for by Eu standardization. PY - 2011 DO - https://doi.org/10.1039/c1ja10227c SN - 0267-9477 SN - 1364-5544 N1 - Geburtsname von Müller, Larissa: Wäntig, L. - Birth name of Müller, Larissa: Wäntig, L. VL - 26 IS - 11 SP - 2160 EP - 2165 PB - Royal Society of Chemistry CY - London AN - OPUS4-24964 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Schlegel, Moritz-Caspar A1 - Müller, Urs A1 - Panne, Ulrich A1 - Emmerling, Franziska T1 - Deciphering the sulfate attack of cementitious materials by high-resolution micro-x-ray diffraction N2 - The durability of cementitious materials depends, among others, on their resistance against chemical attack during the service life of a building. Here, we present an approach to analyze changes in the phase composition due to chemical attack in the form of sulfate ingress within the microstructure. Micro-X-ray (µX-ray) diffraction using synchrotron radiation in Debye–Scherrer (transmission) geometry allowed a spatial resolution of 10 µm. Phase transformations in the wake of damaging processes were observed in a detailed high-resolution imaging study. In comparison, samples containing supplementary cementitious materials were investigated and used to reconstruct the influence of different degeneration processes in detail. Additionally, reaction fronts within the bulk were localized by micro-X-ray fluorescence analysis. The experimental setup provided the possibility for analyzing the phase assemblage of a given sample without destroying the microstructure. The specimens for phase analysis are thick sections of the primary material and can be used for further microscopic analysis of the microstructure and microchemistry, e.g., scanning electron microscopy–energy-dispersive X-ray spectroscopy (SEM–EDX) or Raman spectroscopy. PY - 2011 DO - https://doi.org/10.1021/ac200181g SN - 0003-2700 SN - 1520-6882 VL - 83 IS - 10 SP - 3744 EP - 3749 PB - American Chemical Society CY - Washington, DC AN - OPUS4-23669 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Dutschke, Andreas A1 - Lohrer, Christian A1 - Kurth, Lutz A1 - Seeger, Stefan A1 - Barthel, Mathias A1 - Panne, Ulrich T1 - Aerosol emissions from outdoor firework displays N2 - The results of investigations on the emission of particulate matter from three outdoor firework displays are presented. During these events, possibly harmful and toxic aerosol emissions were measured. Aerosols in the size range from 30 to 450 nm were detected. The maxima in the observed particle size distribution were between 80 and 175 nm, and the highest value determined was ~1.3 · 106 particles per cm³, while the highest mass concentration was approximately 1.8 mg m–3. The time-averaged particle mass concentration did, however, not exceed 0.7 mg m–3 (over 15 min). KW - Firework display KW - Particle size range KW - Particulate matter KW - Threshold values PY - 2011 DO - https://doi.org/10.1002/ceat.201100080 SN - 0930-7516 SN - 1521-4125 VL - 34 IS - 12 SP - 2044 EP - 2050 PB - Wiley-VCH Verl. CY - Weinheim AN - OPUS4-24929 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Leiterer, Jork A1 - Panne, Ulrich A1 - Thünemann, Andreas A1 - Weidner, Steffen T1 - Container-less polymerization in acoustically levitated droplets: an analytical study by GPC and MALDI-TOF mass spectrometry N2 - Molecular masses and end groups of polystyrene (PS) formed in a novel container-less polymerization strategy, based on levitated droplets in an acoustic trap, were determined by Gel Permeation Chromatography (GPC) and Matrix-assisted Laser Desorption/Ionization Time of Flight Mass spectrometry (MALDI-TOF MS). KW - Ultraschallfalle KW - MALDI Massenspektrometrie KW - Polymerisation PY - 2011 DO - https://doi.org/10.1039/c0ay00390e SN - 1759-9660 SN - 1759-9679 VL - 3 IS - 1 SP - 70 EP - 73 PB - RSC Publ. CY - Cambridge AN - OPUS4-23655 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Schlegel, Moritz-Caspar A1 - Müller, Urs A1 - Panne, Ulrich A1 - Emmerling, Franziska A1 - Meng, Birgit ED - Rogge, A. ED - Meng, B. T1 - Neue Wege in der Zementanalytik - Hochaufgelöste in-situ Sy-XRD T2 - 52. DAfStb-Forschungskolloquium CY - Berlin, Deutschland DA - 2011-11-07 PY - 2011 SN - 978-3-9814281-0-0 SP - 18 EP - 24 PB - BAM Bundesanstalt für Materialforschung und -prüfung CY - Berlin AN - OPUS4-25197 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Kühn, Andreas A1 - Scharf, Oliver A1 - Ordavo, I. A1 - Riesemeier, Heinrich A1 - Reinholz, Uwe A1 - Radtke, Martin A1 - Berger, Achim A1 - Ostermann, Markus A1 - Panne, Ulrich T1 - Pushing the limits for fast spatially resolved elemental distribution patterns N2 - A new setup for fast spatially resolved measurements of elemental trace amounts under total reflection conditions using a new colour X-ray camera is described. Samples prepared on conventional total reflection X-ray fluorescence (TXRF) reflectors were measured at BESSY II synchrotron. A spatial resolution of 50 × 50 µm² was obtained, while the required time for the investigation of a 10 × 10 mm² sample is 30 seconds. The set-up is up to 350 times faster than conventional X-ray fluorescence systems for elemental traces. The major components of the X-ray camera are polycapillary optics and a pn-CCD chip with an active area of 13 × 13 mm². This area is divided into 264 × 264 pixels of 48 × 48 µm². A full X-ray spectrum with a resolution of 152 eV @ 5.9 keV and a chip temperature of 246 K is recorded for each pixel. The chip has a read-out rate of 400 Hz. PY - 2011 DO - https://doi.org/10.1039/c1ja10069f SN - 0267-9477 SN - 1364-5544 VL - 26 SP - 1986 EP - 1989 PB - Royal Society of Chemistry CY - London AN - OPUS4-24386 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -