TY - JOUR A1 - Abad Andrade, Carlos Enrique A1 - Florek, S. A1 - Becker-Ross, H. A1 - Huang, M.-D. A1 - de Oliveira Guilherme Buzanich, Ana A1 - Radtke, Martin A1 - Lippitz, Andreas A1 - Hodoroaba, Vasile-Dan A1 - Schmid, Thomas A1 - Heinrich, Hans-Joachim A1 - Recknagel, Sebastian A1 - Jakubowski, Norbert A1 - Panne, Ulrich T1 - Zirconium permanent modifiers for graphite furnaces used in absorption spectrometry: understanding their structure and mechanism of action N2 - The mechanism of action of zirconium permanent modifiers on graphite surfaces was investigated in order to understand its influence on the analytical signal in atomic and molecular absorption spectrometry (AAS/MAS). For this, the molecule formation of CaF was studied, which is used for the indirect analytical determination of fluorine in high-resolution continuum source graphite furnace molecular absorption spectrometry (HR-CS-GFMAS). The kinetics of this reaction was established by monitoring its molecular spectrum at different atomisation temperatures. An Arrhenius plot showed a pseudo-first order reaction with respect to fluorine (n = 1). An intermediate state was isolated, and its structure was elucidated by spectroscopic methods: scanning electron microscopy with energy dispersive X-ray spectroscopy (SEMEDX), X-ray photoelectron spectroscopy (XPS), X-ray absorption spectroscopy (XANES and EXAFS), and Raman microspectroscopy. We propose here a mechanism, where ZrO2 acts as a heterogeneous catalyst: after a pyrolytic step, an intermediate state of ZrO(OCaF) is activated, and at higher temperatures, CaF(g) is released from the zirconium-coated graphite surface. No evidence of the formation of zirconium carbide was found. Consequently, as the CaF formation is catalysed by a heterogeneous catalyst, surface modifications with ZrO2 nanoparticles and ZrO xerogels were investigated in order to increase the surface area. Their influence was evaluated in the molecule formation of CaF, CaCl, CaBr, and CaI. Graphite furnace modification with zirconium oxide nanoparticles proves to be the best choice for fluorine analysis with a signal enhancement of more than eleven times with respect a non-coated graphite furnace. However, the influence of zirconium modifications in the analytical signals of Cl, and I is lower than the F signals or even negative in case of the Br. Understanding zirconium modifiers as heterogeneous catalysts offers a new perspective to AAS and MAS, and reveals the potential of surface analytical methods for development of improved permanent modifiers and graphite furnace coatings. KW - Zirconium KW - HR-CS-MAS KW - Graphite furnace KW - Nanoparticles KW - Xerogel KW - Calcium monofluoride KW - Absorption spectrometry PY - 2018 UR - https://pubs.rsc.org/en/content/articlelanding/2018/ja/c8ja00190a U6 - https://doi.org/10.1039/C8JA00190A SN - 0267-9477 VL - 33 IS - 12 SP - 2034 EP - 2042 PB - Royal Society of Chemistry AN - OPUS4-46775 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Abad Andrade, Carlos Enrique A1 - Florek, S. A1 - Becker-Ross, H. A1 - Huang, M.-D. A1 - Heinrich, Hans-Joachim A1 - Recknagel, Sebastian A1 - Vogl, Jochen A1 - Jakubowski, Norbert A1 - Panne, Ulrich T1 - Determination of boron isotope ratios by high-resolution continuum source molecular absorption spectrometry using graphite furnace vaporizers N2 - Boron isotope amount ratios n(10B)/n(11B) have been determined by monitoring the absorption spectrum of boron monohydride (BH) in a graphite furnace using high-resolution continuum source molecular absorption spectrometry (HR-CS-MAS). Bands (0→0) and (1→1) for the electronic transition X1Σ+ → A1Π were evaluated around wavelengths 433.1 nm and 437.1 nm respectively. Clean and free of memory effect molecular spectra of BH were recorded. In order to eliminate the memory effect of boron, a combination of 2% (v/v) hydrogen gas in argon and 1% trifluoromethane in argon, an acid solution of calcium chloride and mannitol as chemical modifiers was used. Partial least square regression (PLS) for analysis of samples and reference materials were applied. For this, a spectral library with different isotopes ratios for PLS regression was built. Results obtained around the 433.1 nm and 437.1 nm spectral regions are metrologically compatible with those reported by mass spectrometric methods. Moreover, for the evaluated region of 437 nm, an accuracy of 0.15‰ is obtained as the average deviation from the isotope reference materials. Expanded uncertainties with a coverage factor of k = 2 range between 0.15 and 0.44‰. This accuracy and precision are compatible with those obtained by mass spectrometry for boron isotope ratio measurements. KW - Boron isotopes KW - Isotope ratios KW - Boron monohydride KW - Molecular absorption KW - High-resolution continuum source absorption spectrometry KW - Graphite furnace KW - Memory effect KW - HR-CS-MAS PY - 2017 UR - http://www.sciencedirect.com/science/article/pii/S0584854717302537 U6 - https://doi.org/10.1016/j.sab.2017.08.012 SN - 0584-8547 VL - 136 SP - 116 EP - 122 PB - Elsevier CY - Amsterdam, The Netherlands AN - OPUS4-42071 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Abad Andrade, Carlos Enrique A1 - Florek, S. A1 - Becker-Ross, H. A1 - Huang, M.-D. A1 - Lopez-Linares, F. A1 - Poirier, L. A1 - Jakubowski, Norbert A1 - Recknagel, Sebastian A1 - Panne, Ulrich T1 - Shake, shut, and go – A fast screening of sulfur in heavy crude oils by highresolution continuum source graphite furnace molecular absorption spectrometry via GeS molecule detection N2 - A fast and simple method for sulfur quantification in crude oils was developed by using high-resolution continuum source graphite furnace molecular absorption spectrometry (HR-CS-GFMAS). For this, heavy crude oil samples were prepared as microemulsion (shake) and injected into a graphite furnace (shut). Finally, the concentration of sulfur was determined by monitoring in situ the transient molecular spectrum of GeS at wavelength 295.205nm after adding a germanium solution as molecular forming agent (and go). Zirconium dioxide in the form of nanoparticles (45–55nm) was employed as a permanent modifier of the graphite furnace. Calibration was done with an aqueous solution standard of ammonium sulfate, and a characteristic mass (m0) of 7.5ng was achieved. The effectiveness of the proposed method was evaluated analizing, ten heavy crude oil samples with Sulfur amounts ranging between 0.3 and 4.5% as well as two NIST standard reference materials, 1620c and 1622e. Results were compared with those obtained by routine ICP-OES analysis, and no statistical relevant differences were found. KW - Heavy crude oil KW - Sulfur KW - HR-CS-MAS KW - Germanium sulfide KW - Microemulsion PY - 2019 U6 - https://doi.org/10.1016/j.sab.2019.105671 SN - 0584-8547 VL - 160 SP - 105671 PB - Elsevier B.V. CY - Amsterdam, Netherlands AN - OPUS4-48747 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Abad Andrade, Carlos Enrique A1 - Mimus, S. A1 - Recknagel, Sebastian A1 - Jakubowski, N. A1 - Panne, Ulrich A1 - Becker-Ross, H. A1 - Huang, M.-D. T1 - Determination of organic chlorine in water via AlCl derivatization and detection by high-resolution continuum source graphite furnace molecular absorption spectrometry N2 - High-resolution continuum source graphite furnace molecular absorption spectrometry (HR-CS-GF-MAS) was employed for determining adsorbable organic chlorine (AOCl) in water. Organic chlorine was indirectly quantified by monitoring the molecular absorption of the transient aluminum monochloride molecule (AlCl) around a wavelength of 261.42 nm in a graphite furnace. An aluminum solution was used as the molecularforming modifier. A zirconium coated graphite furnace, as well as Sr and Ag solutions were applied as modifiers for a maximal enhancement of the absorption signal. The pyrolysis and vaporization temperatures were 600 °C and 2300 °C, respectively. Non-spectral interferences were observed with F, Br, and I at concentrations higher than 6 mg L-1, 50 mg L-1, and 100 mg L-1, respectively. Calibration curves with NaCl, 4-chlorophenol, and trichlorophenol present the same slope and dynamic range, which indicates the chlorine atom specificity of the method. This method was evaluated and validated using synthetic water samples, following the current standard DIN EN ISO 9562:2004 for the determination of the sum parameter adsorbable organic halides (AOX) for water quality. These samples contain 4-chlorophenol as the chlorinated organic standard in an inorganic chloride matrix. Prior to analysis, organic chlorine was extracted from the inorganic matrix via solid-phase extraction with a recovery rate >95%. There were no statistically significant differences observed between measured and known values and for a t-test a confidence level of 95% was achieved. The limits of detection and characteristic mass were found to be 48 and 22 pg, respectively. The calibration curve was linear in the range 0.1–2.5 ng with a correlation coefficient R2 = 0.9986. KW - Chlorides KW - Chlorine KW - Graphite furnace KW - Spectrometry KW - Diatomic molecule KW - Water KW - AlCl PY - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:kobv:b43-534426 SN - 1759-9660 VL - 13 IS - 33 SP - 3724 EP - 3730 PB - The Royal Society of Chemistry CY - London, UK AN - OPUS4-53442 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Abdelshafi, Nahla A. A1 - Panne, Ulrich A1 - Schneider, Rudolf T1 - Screening for cocaine on Euro banknotes by a highly sensitive enzyme immunoassay N2 - This study focused on quantitative detection of cocaine on Euro banknotes in Germany. A sensitive direct competitive immunoassay was developed and optimized with a limit of detection (LOD) of 5.6 ng/L. Exhaustive cocaine extraction by solvent was tested using different methanol concentrations and buffered solutions. Crossreactivity studies were performed to determine the degree of interference of cocaine metabolites with the immunoassay. Sixty-five Euro banknotes obtained from different districts in Berlin were evaluated. A 100% contamination frequency with cocaine was detected. A comparison between the amount of cocaine extracted by cotton swabbing of one square centimeter of the banknote showed a good correlation for lower contamination levels. This assay showed high sensitivity of detecting pg of cocaine per 1 cm2 of one banknote by swabbing 1 cm2: 0, 14, and 21 pg/cm2. Moreover, three notes of different denominations revealed high cocaine concentration; 1.1 mg/note, and twice 55 μg/note. KW - Cocaine KW - Euro banknotes KW - Drugs on bills KW - Enzyme-Linked Immunosorbent Assay KW - ELISA PY - 2017 U6 - https://doi.org/10.1016/j.talanta.2017.01.004 SN - 0039-9140 SN - 1873-3573 VL - 165 SP - 619 EP - 624 PB - Elsevier AN - OPUS4-39150 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Andrade, M. A. B. A1 - Stindt, Arne A1 - Panne, Ulrich A1 - Riedel, Jens A1 - Adamowski, J. C. T1 - Numerical and experimental determination of the sound pressure distribution in single-axis acoustic levitators N2 - This paper presents a numerical and an experimental procedure to obtain the pressure field in single-axis acoustic levitators. Numerically, the pressure field is determined by a matrix method based on the Rayleigh integral that take into account the multiple wave reflections that occur between the transducer and the reflector. The numerical pressure field is compared with the acoustic pressure measured by an earplug microphone, that is connected to a hollow needle. The tip of the needle is moved by a µm translation stage through the field and the signal is recorded using a lock-in amplifier locked to the levitator frequency. The pressure field obtained numerically show good agreement with that obtained experimentally. T2 - 2012 IEEE International Ultrasonics Symposium (IUS) CY - Dresden, Germany DA - 2012-10-07 KW - Acoustic levitation KW - Matrix method KW - Pressure field PY - 2012 SN - 978-1-4673-4562-0 U6 - https://doi.org/10.1109/ULTSYM.2012.0690 SN - 1948-5719 SP - 2754 EP - 2757 AN - OPUS4-32587 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bahlmann, Arnold A1 - Carvalho, José Joao A1 - Weller, Michael G. A1 - Panne, Ulrich A1 - Schneider, Rudolf T1 - Immunoassays as high-throughput tools: monitoring spatial and temporal variations of carbamazepine, caffeine and cetirizine in surface and wastewaters N2 - Carbamazepine (CBZ), caffeine and cetirizine were monitored by enzyme-linked immunosorbent assays (ELISAs) in surface and wastewaters from Berlin, Germany. This fast and cost-efficient method enabled to assess the spatial and temporal variation of these anthropogenic markers in a high-throughput screening. CBZ and cetirizine were detected by the same antibody, which selectively discriminates between both compounds depending on the pH value used in the incubation step. To our best knowledge, this is the first dual-analyte immunoassay working with a single antibody. The frequent sampling with 487 samples being processed allowed for the repeated detection of unusually high concentrations of CBZ and caffeine. ELISA results correlate well with the ones obtained by liquid chromatography tandem mass spectrometry (LC-MS/MS). Caffeine concentrations found in surface waters were elevated by combined sewer overflows after stormwater events. During the hay fever season, the concentrations of the antihistamine drug cetirizine increased in both surface and wastewaters. Caffeine was almost completely removed during wastewater treatment, while CBZ and cetirizine were found to be more persistent. The maximum concentrations of caffeine, CBZ and cetirizine found in influent wastewater by LC–MS/MS were 470, 5.0 and 0.49 µg L-1, while in effluent wastewater the concentrations were 0.22, 4.5 and 0.51 µg L-1, respectively. For surface waters, concentrations up to 3.3, 4.5 and 0.72 µg L-1 were found, respectively. KW - ELISA KW - Carbamazepine KW - Caffeine KW - Cetirizine KW - Environmental monitoring KW - Anthropogenic markers PY - 2012 U6 - https://doi.org/10.1016/j.chemosphere.2012.05.020 SN - 0045-6535 SN - 0366-7111 VL - 89 IS - 11 SP - 1278 EP - 1286 PB - Elsevier Science CY - Kidlington, Oxford AN - OPUS4-26187 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bahlmann, Arnold A1 - Falkenhagen, Jana A1 - Weller, Michael G. A1 - Panne, Ulrich A1 - Schneider, Rudolf T1 - Cetirizine as pH-dependent cross-reactant in a carbamazepine-specific immunoassay N2 - High performance liquid chromatography (HPLC) was hyphenated with a previously reported carbamazepine-specific enzyme-linked immunosorbent assay (ELISA) as a screening approach to water analysis in order to identify possible interferences from transformation products. Treated wastewater was analysed and three substances were recognized by the antibody besides carbamazepine: the metabolites 10,11-dihydro-10,11-epoxycarbamazepine and 2-hydroxycarbamazepine plus the structurally not obviously related antihistamine cetirizine. The molar cross-reactivity against cetirizine was found to be pH-dependent and assessed to be 400% at pH 4.5 and 22% at pH 10.5. Performing the ELISA at pH 10.5 greatly improved the accuracy when carbamazepine was determined in surface and wastewater samples. KW - Immunoassay KW - ELISA KW - Cetirizin KW - Carbamazepin KW - pH-Abhängigkeit KW - Kreuzreaktivität PY - 2011 U6 - https://doi.org/10.1039/c0an00928h SN - 0003-2654 SN - 1364-5528 VL - 136 IS - 7 SP - 1357 EP - 1364 PB - Royal Society of Chemistry CY - Cambridge AN - OPUS4-23546 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bahlmann, Arnold A1 - Weller, Michael G. A1 - Panne, Ulrich A1 - Schneider, Rudolf T1 - Monitoring carbamazepine in surface and wastewaters by an immunoassay based on a monoclonal antibody N2 - The pharmaceutical compound carbamazepine (CBZ) is an emerging pollutant in the aquatic environment and may potentially be used as a wastewater marker. In this work, an enzyme-linked immunosorbent assay (ELISA) for the detection of carbamazepine in surface and sewage waters has been developed. The heterogeneous immunoassay is based on a commercially available monoclonal antibody and a novel enzyme conjugate (tracer) that links the hapten via a hydrophilic peptide (triglycine) spacer to horseradish peroxidase. The assay achieves a limit of detection of 24 ng/L and a quantitation range of 0.05-50 µg/L. The analytical performance and figure of merits were compared to liquid chromatography-tandem mass spectrometry after solid-phase extraction. For nine Berlin surface water samples and one wastewater sample, a close correlation of results was observed. A constant overestimation relative to the CBZ concentration of approximately 30% by ELISA is probably caused by the presence of 10,11-epoxy-CBZ and 2-hydroxy-CBZ in the samples. The ELISA displayed cross-reactivities for these compounds of 83% and 14%, respectively. In a first screening of 27 surface water samples, CBZ was detected in every sample with concentrations between 0.05 and 3.2 µg/L. Since no sample cleanup is required, the assay allowed for the determination of carbamazepine with high sensitivity at low costs and with much higher throughput than with conventional methods. KW - Carbamazepine KW - ELISA KW - Antibody KW - Immunoassay KW - Surface water KW - LC-MS/MS KW - Pharmaceuticals PY - 2009 U6 - https://doi.org/10.1007/s00216-009-2958-7 SN - 1618-2642 SN - 1618-2650 VL - 395 IS - 6 SP - 1809 EP - 1820 PB - Springer CY - Berlin AN - OPUS4-20927 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Balski, Matthias A1 - Emmerling, Franziska A1 - Kipphardt, Heinrich A1 - Panne, Ulrich T1 - Determination of boron in silicon without use of additional complexing agents N2 - Methods for the determination of boron in various matrices described in the literature usually employ complexing agents like mannitol to retain the volatile boron species during matrix evaporation steps. However, also relatively high boron recoveries from silicon containing samples have been reported when no complexing agents were added to the digestion acids. The mechanism behind this matrix-dependant recovery has been investigated by studying the boron recovery in the analysis of solar grade silicon. It was found that the NH4+ ion formed by the reduction of nitric acid during sample digestion is responsible for the higher recovery of boron, which leads to a possible analysis method without the use of complexing agents if the sample preparation procedure is carefully optimized. KW - Boron KW - Complexing agents KW - Matrix evaporation KW - Sample enrichment PY - 2014 U6 - https://doi.org/10.1039/c4ay00410h SN - 1759-9660 SN - 1759-9679 VL - 6 IS - 12 SP - 4003 EP - 4008 PB - RSC Publ. CY - Cambridge AN - OPUS4-30956 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -