TY - JOUR A1 - Simons, C. A1 - Mans, C. A1 - Hanning, S. A1 - Janßen, A. A1 - Radtke, Martin A1 - Reinholz, Uwe A1 - Ostermann, Markus A1 - Michaelis, Matthias A1 - Wienold, Julia A1 - Alber, D. A1 - Kreyenschmidt, M. T1 - Study on microscopic homogeneity of polymeric candidate reference materials BAM H001-BAM H010 by means of synchrotron µ-XRF and LA-ICP-MS N2 - In this study the microscopic homogeneity of the newly developed plastic reference materials BAM H001-BAM H010 was investigated. The materials consist of an acrylonitryle-butadien-styrene terpolymer, doped with different amounts of the elements Br, Cd, Cr, Hg and Pb. For the quantitative determination of the degree of homogeneity, a procedure proposed by Kempenaers et al. (Fresenius J. Anal. Chem., 2001, 369, 733-737) was used. On every sample an extensive number of different points were analyzed and standard deviation for every element mentioned above was used to calculate a minimal sampling mass that is necessary to reach a certain level of uncertainty caused by inhomogeneity (mmin,5%). The experiments were taken out with synchrotron µ-XRF (SR µ-XRF) at BESSYII in Berlin and by laser ablation inductively coupled plasma mass spectroscopy (LA-ICP-MS). The calculated values for mmin,5% of both techniques showed comparable results for all elements. It could be shown that the materials are suitable for calibration of micro analytic techniques if at least 64 µg are used. KW - XRF KW - LA-ICP-MS KW - RoHS KW - CRM PY - 2010 U6 - https://doi.org/10.1039/b917008a SN - 0267-9477 SN - 1364-5544 VL - 25 SP - 40 EP - 43 PB - Royal Society of Chemistry CY - London AN - OPUS4-20938 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ostermann, Markus A1 - Berger, Achim A1 - Mans, C. A1 - Simons, C. A1 - Hanning, S. A1 - Janßen, A. A1 - Kreyenschmidt, M. T1 - Results and effects of homogeneity and stability studies carried out by X-ray fluorescence on the new reference material BAM-H010 N2 - The Federal Institute for materials research and testing (BAM)in cooperation with the Fachhochschule Münster has developed a new certified polymeric reference material, BAM-H010. The reference material consists of acrylonitrile–butadiene–styrene terpolymer (ABS) and has been doped with different contents of the elements Br, Cd, Cr, Hg and Pb. The monitoring of these elements is demanded in the EU directive 2002/95/EG, commonly referred to as RoHS. We have shown that the uncertainty contribution of the thermal long-term stability (ustability) is negligible compared to the contribution to the uncertainty caused by the method used for the characterization of the elemental content and the batch inhomogeneity (uhomogeneity). The stability of theABS material during exposure of the samples with hard X-radiation leads to a degradation of the ABS matrix with either long- or short-chain products. KW - Reference material KW - Plastic KW - RoHS KW - WEEE KW - X-ray fluorescence PY - 2011 U6 - https://doi.org/10.1007/s00769-011-0806-x SN - 0949-1775 SN - 1432-0517 VL - 16 IS - 10 SP - 515 EP - 522 PB - Springer CY - Berlin AN - OPUS4-24249 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rackwitz, Vanessa A1 - Ostermann, Markus A1 - Panne, Ulrich A1 - Hodoroaba, Vasile-Dan T1 - Performance of mu-XRF with SEM/EDS for trace analysis on the example of RoHS relevant elements - measurement, optimisation and predivtion of the detection limits N2 - For ten years µ-XRF (micro-focus X-ray fluorescence) analysis has been performed with SEM/EDS (scanning electron microscope with an energy dispersive X-ray detector) so that non-destructive analysis of elements at trace level concentrations below 100 µg g-1 becomes possible. This can be considered as a valuable completion of the classical electron probe microanalysis by EDS, an analytical method 'suffering' from rather poor limits of detection in the range of one to two orders of magnitude higher than those of µ-XRF. Based on a representative actual application, namely analysis of RoHS relevant elements at trace concentration levels, the performance of the rather new analytical method with respect to its limits of detection is systematically evaluated. CRMs (certified reference materials) specially prepared to support the quantitative XRF analysis of RoHS relevant elements were employed. On the other side, based on calculations of µ-XRF spectra according to a recently developed physical model the optimization of the analytical performance is also successfully undertaken. KW - RoHS KW - XRF KW - muXRF KW - SEM/EDS KW - Detection limits KW - Trace elements KW - Modelling PY - 2013 U6 - https://doi.org/10.1039/c3ja50064k SN - 0267-9477 SN - 1364-5544 VL - 28 IS - 9 SP - 1466 EP - 1474 PB - Royal Society of Chemistry CY - London AN - OPUS4-29045 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -