TY - JOUR A1 - Kipphardt, Heinrich A1 - Matschat, Ralf A1 - Vogl, Jochen A1 - Gusarova, Tamara A1 - Czerwensky, Michael A1 - Heinrich, Hans-Joachim A1 - Hioki, A. A1 - Konopelko, L.A. A1 - Methven, B. A1 - Miura, T. A1 - Petersen, O. A1 - Riebe, Gundel A1 - Sturgeon, R. A1 - Turk, G.C. A1 - Yu, L. L. T1 - Purity determination as needed for the realisation of primary standards for elemental determination: status of international comparability N2 - Within the National Metrology Institutes (NMIs) and designated laboratories, an interlaboratory comparison, CCQM-P107, was conducted to verify the degree of international comparability concerning the results of purity analysis. The mass fractions of Ag, Bi, Cd, Cr, Ni, Tl at the lower mg/kg-level in a high purity zinc material were determined, but the real measurand in metrological sense was the sum of the six mass fractions. Homogeneity was investigated by glow discharge mass spectrometry, reference values were obtained using isotope dilution mass spectrometry. Six NMIs participated, contributing eight independent data sets. The agreement amongst the results of the participants, their median and the agreement with the reference values were usually excellent and in almost all cases below the target uncertainty of 30% relative. In this manner, the accuracy of results and the comparability between the participants was demonstrated to be established. KW - Interlaboratory comparison KW - Purity analysis KW - High purity metals KW - CCQM-P107 KW - Zinc PY - 2010 DO - https://doi.org/10.1007/s00769-009-0557-0 SN - 0949-1775 SN - 1432-0517 VL - 15 IS - 1 SP - 29 EP - 37 PB - Springer CY - Berlin AN - OPUS4-20767 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Kipphardt, Heinrich A1 - Dudzus, Thomas A1 - Meier, Klaus A1 - Recknagel, Sebastian A1 - Hedrich, Martina A1 - Matschat, Ralf T1 - Measurement of oxygen and nitrogen in high purity metals used as national standards for elemental analysis in Germany by classical carrier gas hot extraction (HE) and HE after activation with photons N2 - The special importance of the analysis of non-metals in high purity metals, which will serve as national standards for elemental analysis in Germany, is illustrated for oxygen and nitrogen. The typical range of the mass fraction of oxygen and nitrogen in these materials is below 10.MU.g/g, often close to 1.MU.g/g. The two methods applied at BAM for these measurements are classical carrier gas hot extraction (HE) and carrier gas hot extraction after activation with photons (PAA-HE). The approach, the methods and their advantages and limitations are discussed. Comparative results from the measurement of oxygen and nitrogen in Cu, Fe, Ga, Pb, Sn and W are presented. (author abst.) T2 - 8th International Conference on Ultra-High Purity Base Metals (UHPM-2001) CY - Berlin, Germany DA - 2001-06-25 KW - Oxygen determination KW - Nitrogen determination KW - High purity metals KW - Carrier gas hot extraction KW - Photon activation analysis PY - 2002 DO - https://doi.org/10.2320/matertrans.43.98 SN - 1345-9678 SN - 1347-5320 VL - 43 IS - 2 SP - 98 EP - 100 PB - Japan Institute of Metals CY - Sendai AN - OPUS4-1361 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Matschat, Ralf A1 - Czerwensky, Michael A1 - Pattberg, S. A1 - Heinrich, Hans-Joachim A1 - Tutschku, S. T1 - High Purity Metals as Primary Calibration Materials for Elemental Analysis - Their Importance and Their Certification N2 - The Bundesanstalt fuer Materialforschung und -pruefung, BAM (Federal Institute for Materials Research and Testing) continues to establish a system of primary reference materials to meet the demands of metrological traceability. The materials act as national standards in the field of elemental analysis. For all elements of the periodic table -excepting those that are gases or radioactive- two different kinds of reference materials are being certified. The substances are of very high purity and of defined stoichiometry. Pure elements and metals are used as far as possible. They are certified by determining the trace contents of most elements of the periodic table at very low levels using different trace element analysis methods. Recent application of these methods is described and examples of the certification of some pure metals (copper, iron and lead) are given. (author abst.) T2 - 8th International Conference on Ultra-High Purity Base Metals (UHPM-2001) CY - Berlin, Germany DA - 2001-06-25 KW - Trace elemental analysis KW - Atomic spectrometry KW - High purity metals KW - Certification KW - Primary calibration materials KW - Reference materials KW - National standards PY - 2002 UR - https://www.researchgate.net/publication/235927297_High_purity_metals_as_primary_calibration_materials_for_elemental_analysis_-_Their_importance_and_their_certification SN - 1345-9678 SN - 1347-5320 VL - 43 IS - 2 SP - 90 EP - 97 PB - Japan Institute of Metals CY - Sendai AN - OPUS4-1600 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -