TY - JOUR A1 - Gusarova, Tamara A1 - Hodoroaba, Vasile-Dan A1 - Matschat, Ralf A1 - Kipphardt, Heinrich A1 - Panne, Ulrich T1 - Exploitation of the hollow cathode effect for sensitivity enhancement of Grimm-type DC glow discharge optical emission spectroscopy N2 - The hollow cathode (HC) effect was investigated in non-cooled 15 mm deep drilled flat metallic analytical samples that were easy to prepare. The deep cavity used ("complete HC" in contrast to "recessed HC" with 2-3 mm deep cavity) intensified the HC effect notably and therefore distinctly improved the detection power of the common GD-OES with planar cathodes. A signal enhancement of up to a factor of 150, not reported earlier, in comparison with flat conventional samples was achieved. A better separation of the analytical lines from spectral interferences was observed when the HC assembly was applied. Additionally, an effect of strongly enhanced intensities of atomic lines and somewhat decreased intensities of ionic lines was detected in the case of HC in comparison to usual planar cathodes. The investigations were carried out with samples of copper, steel and zinc matrices using both the same and individually optimised glow discharge (GD) electrical parameters. KW - Hollow cathode KW - Grimm type glow discharge KW - Emission spectroscopy KW - GD-OES KW - Sensitivity PY - 2009 DO - https://doi.org/10.1039/b814977a SN - 0267-9477 SN - 1364-5544 VL - 24 SP - 680 EP - 684 PB - Royal Society of Chemistry CY - London AN - OPUS4-19600 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Richter, Silke A1 - Roethke, A. A1 - Rienitz, O. A1 - Matschat, Ralf A1 - Schiel, D. A1 - Jaehrling, R. A1 - Goerlitz, V. A1 - Kipphardt, Heinrich T1 - SI-traceable monoelemental solutions on the highest level of accuracy: 25 years from the foundation of CCQM to recent advances in the development of measurement methods N2 - Within the Working Group on Inorganic Analysis (IAWG) of the Consultative Committee for Amount of Substance: Metrology in Chemistry and Biology (CCQM) international key comparisons and pilot studies related to inorganic analysis are carried to ensure consistency in this field at the highest level. Some of these comparisons deal directly with the preparation and characterization of monoelemental solutions or with topics, closely related. The importance of monoelemental solutions lies in the fact that almost every measurement in inorganic analysis relies on the comparison with either a reference material, or references in form of solutions, usually (mono)elemental solutions. All quantitative measurement approaches, e.g. isotope dilution or standard addition, need an accurate reference solution made from a well characterized reference material, prepared under full gravimetric control. These primary (monoelemental) solutions do not only serve as arbitrary references/calibration solutions, but they also link up measurement results to the International System of units (SI), this way establishing the so-called metrological traceability to a measurement unit of the SI. Without such solutions on the highest possible level of accuracy and with the smallest possible associated uncertainties (for e.g. element content and/or impurities), an analysis itself can never be as good as it could be with appropriate reference solutions. This article highlights select key comparisons and pilot studies dealing with monoelemental solution related topics within the IAWG from the foundation of CCQM – 25 years ago – up to latest achievements in the field of inorganic analysis. KW - Metrology KW - SI Traceability KW - CCQM PY - 2019 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-496080 DO - https://doi.org/10.1088/1681-7575/ab5636 SN - 0026-1394 SN - 1681-7575 SP - 1 EP - 23 PB - IOP CY - Bristol AN - OPUS4-49608 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Tiebe, Carlo A1 - Mieller, Björn A1 - Maiwald, Michael A1 - Kipphardt, Heinrich A1 - Tuma, Dirk A1 - Prager, Jens A1 - Schukar, Marcus A1 - Strohhäcker, J. T1 - Sensoren und Analytik für Sicherheit und Prozesskontrolle in Wasserstofftechnologien N2 - Die Nutzung von Sensortechnologien, insbesondere im Bereich der Gasdetektion mit einem Schwerpunkt auf Wasserstoff, spielt eine entscheidende Rolle in verschiedenen Anwendungsbereichen der Wasserstofftechnologie. Sicherheitsüberwachung, Leckdetektion und Prozesskontrolle gehören zu den prominenten Anwendungsgebieten dieser Sensortechnologien. Ein zentrales Ziel ist die Erkennung von freigesetztem Wasserstoff sowie die genaue Bestimmung des Wasserstoff-Luftverhältnisses mithilfe von Gassensoren. Dies ist von entscheidender Bedeutung, um potenzielle Gefahren frühzeitig zu erkennen und angemessene Maßnahmen zu ergreifen. Ein weiterer Schwerpunkt dieses Beitrags liegt auf der Analytik und der Verwendung zertifizierter Referenzmaterialien in Verbindung mit Metrologie für die Wasserstoffspeicherung. Dies gewährleistet eine präzise und zuverlässige Charakterisierung von Wasserstoff und unterstützt die Entwicklung sicherer Speichertechnologien. Im Rahmen des Euramet-Vorhabens Metrology for Advanced Hydrogen Storage Solutions (MefHySto) wird eine Kurzvorstellung präsentiert. Der Vortrag stellt zwei zerstörungsfreie Prüfverfahren zum strukturellen Zustandsüberwachung (Structural Health Monitoring, SHM) für Wasserstofftechnologien vor. Insbesondere die Fehlstellenerkennung mittels geführter Ultraschallwellen spielt eine bedeutende Rolle bei der Lebensdauerüberwachung von Wasserstoffspeichern. Ein weiterer Aspekt ist die Anwendung faseroptischer Sensorik zur Schadensfrüherkennung von Wasserstoffspeichern. Diese zerstörungsfreien Prüfverfahren ermöglichen eine präzise und frühzeitige Identifizierung von Schäden, was die Sicherheit und Effizienz von Wasserstoffspeichersystemen entscheidend verbessert. T2 - DVGW Kongress H2 Sicherheit CY - Online meeting DA - 15.11.2023 KW - H2Safety@BAM KW - Gassensorik KW - Metrologie KW - zertifizierte Referenzmaterialien KW - Zerstörungsfreie Prüfung KW - Ultraschall KW - Faseroptik PY - 2023 AN - OPUS4-59230 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bayat, Mehmet Emin A1 - Kipphardt, Heinrich A1 - Tiebe, Carlo A1 - Tuma, Dirk A1 - Engelhard, Carsten T1 - Trace-Level Ammonia–Water Interactions in Hydrogen: Challenges in Gas Purity Analysis Using Optical-Feedback Cavity-Enhanced Absorption Spectroscopy (OF-CEAS) N2 - Ammonia is a critical impurity in hydrogen fuel due to its irreversible poisoning effect on proton exchange membrane fuel cells. Therefore, international standards (e.g., ISO 14687) set a stringent threshold of 100 nmol/mol. Furthermore, with the growing potential use of ammonia as a hydrogen carrier, its accurate quantification is becoming increasingly important. However, the presence of trace humidity poses analytical challenges, as ammonia may interact with water or interfaces, thereby affecting its detectability. Therefore, the goal of this work is to enable accurate trace ammonia quantification for hydrogen purity measurements through fundamental studies of the methodological challenges. Here, low-pressure sampling (ultra)long-path Optical-Feedback Cavity-Enhanced Absorption Spectroscopy (OF-CEAS) was applied with an effective optical path length of approximately 6.17 km. We studied three average amounts of ammonia: (38.2 ± 0.8) nmol/mol, (74.8 ± 0.7) nmol/mol, and (112.1 ± 1.2) nmol/mol. Furthermore, these amounts were investigated at trace-humidity levels ranging from 0.8 to 8.5 ppmV. We observed a systematic, nonlinear, and humidity-dependent positive measurement bias of up to + (1.0 ± 0.2) nmol/mol at the maximum investigated trace-humidity volume fraction of 8.5 ppmV. This bias was not caused by spectral interference but rather by water-induced accumulation of ammonia within the optical cavity. Moreover, time-resolved measurements in the presence of trace ammonia showed that water desorption follows first-order kinetics, whereas water adsorption followed mixed-order kinetics with an apparent reaction order of 1.57 ± 0.03. Distinct hydration states of surface-bound ammonia were identified, whereas under dry conditions and with increasing amounts of ammonia, enhanced surface adhesion through intermolecular clustering was observed. In addition, the presence of ammonium species within the sorption layer was indirectly confirmed by our experiments. In conclusion, we provide a deeper insight into trace-level ammonia–water interactions and establish a framework for optimizing methodologies, particularly for (ultra)long-path optical gas measurement systems. KW - Ammonia KW - Hydrogen KW - OF-CEAS KW - Humidity KW - Surface Interactions KW - Adsorption PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-643627 DO - https://doi.org/10.1021/acsmeasuresciau.5c00105 SN - 2694-250X SP - 1 EP - 15 PB - American Chemical Society (ACS) AN - OPUS4-64362 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Bayat, Mehmet A1 - Kipphardt, Heinrich A1 - Tiebe, Carlo A1 - Engelhard, Carsten T1 - Effect of Trace Humidity on Trace Ammonia Quantification N2 - Hydrogen is receiving growing attention as a promising medium for energy storage and transportation. Proton exchange membrane (PEM) fuel cells are a key technology for energy conversion from hydrogen. However, high purity hydrogen is required for this technology, as described in standards such as ISO 14687:2019. A critical impurity that impairs PEM efficiency is ammonia, for which a limit value of 100 nmol/mol applies. In addition, the maximum amount of humidity is specified with 5 µmol/mol. The presence of humidity in hydrogen at levels up to 50 times higher than ammonia raises the critical question of whether ammonia can still be quantified accurately in the lower nmol/mol range without interference from water. In addition, the general question arises as to whether the critical analyte, ammonia, can still be detected at all, given the expected surface effects at such amounts of trace humidity. In this work, we describe the influence of trace humidity on the surface behavior of adsorbed ammonia on stainless steel. Samples with ammonia at approximately 100 nmol/mol in hydrogen, with trace humidity levels ranging from 3 µmol/mol to 10 µmol/mol, were carefully prepared using dynamic mixing. Humidification was conducted using an advanced saturation method. These samples were investigated under atmospheric conditions using low pressure sampling Optical Feedback Cavity Enhanced Absorption Spectroscopy (OFCEAS). From the observed time-dependent signals, kinetics were determined and analyzed, yielding insights on the strength of the adsorption behavior of the analytes. T2 - ANAKON 2025 CY - Leipzig, Germany DA - 10.03.2025 KW - OF-CEAS KW - Hydrogen KW - Ammonia KW - Water KW - Humidity KW - Kinetics KW - Adsorption PY - 2025 UR - https://veranstaltungen.gdch.de/microsite/index.cfm?l=11713&sp_id=2&selSiteID=vplanner AN - OPUS4-64046 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Heinrich, Hans-Joachim A1 - Kipphardt, Heinrich T1 - Application of methane as a gaseous modifier for the determination of silicon using electrothermal atomic absorption spectrometry N2 - For determination of silicon in aqueous solutions by electrothermal atomic absorption spectrometry methane/argon mixtures as a gaseous modifier were applied during the pyrolysis step to improve the analytical performance. The beneficial effects observed on thermal stabilization, signal enhancement and shape of absorbance signals were attributed to the thermal decomposition products of methane, which were hydrogen and carbon black (soot). Using a 5% CH4 mixture with argon, the optimized pyrolysis and atomization temperatures were 1350 °C and 2450 °C, respectively. A flushing step following the pyrolysis was mandatory to avoid background absorption and accelerated deposition of pyrolytic graphite. Characteristic masses of 50 and 30 pg were obtained for standard transversely heated graphite atomizer (THGA) tubes and end-capped THGA tubes, respectively, which were lower than with other previously applied modifiers. A limit of detection of 0.2 µg L- 1 (3 s, n = 10) has been obtained. In addition, this gaseous modifier did not contribute to contamination which often was significant when a liquid modifier solution was co-injected. The proposed method has been applied to the determination of silicon in ultrapure water, nitric and hydrochloric acids. KW - Electrothermal atomic absorption KW - Spectrometry KW - Silicon determination KW - Methane KW - Modifier KW - Ultrapure reagents PY - 2012 DO - https://doi.org/10.1016/j.sab.2012.04.001 SN - 0584-8547 SN - 0038-6987 VL - 70 SP - 68 EP - 73 PB - Elsevier B.V. CY - Amsterdam AN - OPUS4-26050 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Balski, Matthias A1 - Emmerling, Franziska A1 - Kipphardt, Heinrich A1 - Panne, Ulrich T1 - Determination of boron in silicon without use of additional complexing agents N2 - Methods for the determination of boron in various matrices described in the literature usually employ complexing agents like mannitol to retain the volatile boron species during matrix evaporation steps. However, also relatively high boron recoveries from silicon containing samples have been reported when no complexing agents were added to the digestion acids. The mechanism behind this matrix-dependant recovery has been investigated by studying the boron recovery in the analysis of solar grade silicon. It was found that the NH4+ ion formed by the reduction of nitric acid during sample digestion is responsible for the higher recovery of boron, which leads to a possible analysis method without the use of complexing agents if the sample preparation procedure is carefully optimized. KW - Boron KW - Complexing agents KW - Matrix evaporation KW - Sample enrichment PY - 2014 DO - https://doi.org/10.1039/c4ay00410h SN - 1759-9660 SN - 1759-9679 VL - 6 IS - 12 SP - 4003 EP - 4008 PB - RSC Publ. CY - Cambridge AN - OPUS4-30956 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Richter, Matthias A1 - Baldan, A. A1 - Beelen, R. A1 - de Boed, E. J.J. A1 - Tuma, Dirk A1 - Kipphardt, Heinrich A1 - Wilke, Olaf A1 - van der Veen, A. M.H. T1 - Primary gas standards for the determination of sulfur-based impurities at trace level in hydrogen N2 - Hydrogen fuel quality needs to comply with ISO 14687:2025 to avoid harmful impact on applications using proton-exchange membrane (PEM) fuel cells. For total sulfur as one of the most impactful contaminants, an amount fraction of 4 nmol/mol has been set as threshold (so-called Grade D quality). In this study, novel gaseous primary gas standards (PGS) of seven sulfur compounds were prepared either gravimetrically at 1000 nmol/mol, 100 nmol/mol, and 10 nmol/mol, or dynamically diluted down to 4 nmol/mol in hydrogen and argon matrices with relative expanded uncertainties well below 10 % (k = 2) and proven stability of at least 9–12 months. Two different cylinder passivation treatments were compared, with one treatment identified as unsuitable since reactions took place within the mixture. By performing a cross-check study, equivalence between both sets of PGS could be demonstrated (≤ 5 % relative deviation for most of the sulfur species). KW - Hydrogen quality KW - ISO 14687 KW - Sulfur impurities KW - Primary gas standards KW - TD-GC/SCD PY - 2026 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-656759 DO - https://doi.org/10.1016/j.ijhydene.2026.154461 SN - 0360-3199 VL - 224 SP - 1 EP - 13 PB - Elsevier Ltd. AN - OPUS4-65675 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -