TY - CONF A1 - Deresch, Andreas A1 - Jechow, Mirko A1 - Gollwitzer, Christian ED - Chimenti, D.E. ED - Bond, L.J. ED - Thompson, D.O. T1 - Spectral characterization of storage phosphor imaging plates N2 - This work presents attenuation and sensitivity measurements of radiographic imaging plates (IPs) with quasimonoenergetic X-rays in the 8–60 kiloelectronvolt range. The measurements are used to validate theoretical predictions. A short overview of the theoretical model is given. The model can be used to describe the sensitivity of different detector types to a wide range of X-ray energies. T2 - 40th Annual review of progress in quantitative nondestructive evaluation CY - Baltimore, Maryland, USA DA - 21.07.2013 KW - Radiography KW - Imaging plates KW - Spectral sensitivity PY - 2014 SN - 978-0-7354-1212-5 SN - 978-0-7354-1211-8 U6 - https://doi.org/10.1063/1.4865040 SN - 0094-243X N1 - Serientitel: AIP conference proceedings – Series title: AIP conference proceedings VL - 1581 SP - 1786 EP - 1792 PB - AIP Publishing AN - OPUS4-30757 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ewert, Uwe A1 - Beckmann, Jörg A1 - Bellon, Carsten A1 - Jaenisch, Gerd-Rüdiger A1 - Zscherpel, Uwe A1 - Jechow, Mirko T1 - Fehlergrößenabhängige Kontrastreduktion und zusätzliche Unschärfe durch Streustrahlung - Film und digitale Detektoren im Vergleich N2 - Die radiografische Bildqualität in ihrer klassischen Definition hängt von Kontrast, Rauschen und Unschärfe ab. Diese Parameter werden normalerweise mit Bildgüteprüfkörpern (BPK) bestimmt. Die Streustrahlung reduziert im Allgemeinen immer die Bildqualität in der Radiografie. Verschiedene Einflussgrößen werden unter Berücksichtigung von Techniken, wie Film-Radiografie, Computer-Radiografie mit Speicherfolien und digitale Radiographie mit Matrixdetektoren, analysiert. In Lehrbüchern und Standards werden das Streuverhältnis k und verschiedene Kontraste definiert. Für digitale Medien werden zusätzliche Parameter wie das Signal-Rausch-Verhältnis SNR und das Kontrast-Rausch-Verhältnis CNR eingeführt. In diesem Zusammenhang wird der Einfluss der Streustrahlung aus dem Messobjekt und dem Detektor bestehend aus Kassette bzw. Gehäuse und sensitiver Detektorschicht auf die Bildqualität ermittelt. Die numerische Modellierung wird eingesetzt, um die Streustrahlung und die Primärstrahlung aus dem Objekt separat zu bestimmen. Die Streustrahlung aus dem Objekt und Detektor erzeugen eine Abbildung des Objektes, die dem Primärstrahlungsbild überlagert ist, wobei das Streubild eine höhere Unschärfe aufweist. Dieser Effekt hängt von der Fehlergröße, dem Abstand zwischen Detektor und Objekt und der Zwischenfilterung ab. KW - Digitale Radiologie KW - Kontrast-Rausch-Verhältnis KW - Streuverhältnis KW - Röntgen Simulation PY - 2008 SN - 0025-5300 VL - 50 IS - 10 SP - 560 EP - 568 PB - Hanser CY - München AN - OPUS4-18346 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - GEN A1 - Ewert, Uwe A1 - Beckmann, Jörg A1 - Bellon, Carsten A1 - Jaenisch, Gerd-Rüdiger A1 - Zscherpel, Uwe A1 - Jechow, Mirko A1 - Großer, A. T1 - Fehlergrößenabhängige Kontrastreduktion und zusätzliche Unschärfe durch Streustrahlung reduzierendie Bildqualität radiologischer Abbildungen - Film und digitale Detektoren im Vergleich T2 - DACH-Jahrestagung 2008 - Zerstörungsfreie Materialprüfung, ZfP in Forschung, Entwicklung und Anwendung CY - St. Gallen, Schweiz DA - 2008-04-28 KW - Digitale Radiologie KW - Streuverhältnis KW - Strahlenkontrast KW - Unschärfe PY - 2008 SN - 978-3-940283-09-2 IS - DGZfP-BB 113 (Di.1.A.1) SP - 133 EP - 144 AN - OPUS4-18349 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Ewert, Uwe A1 - Heyne, Klara A1 - Zscherpel, Uwe A1 - Jechow, Mirko A1 - Bavendiek, K. ED - Thompson, D.O. ED - Chimenti, D.E. T1 - Optimum exposure conditions for computed radiography depending on fixed pattern noise and efficiency of imaging plate-scanner systems N2 - The presently active standards on Computed Radiography (CR) need a major revision. It was observed by many users that the image quality for class B of EN 14784-2 is not achievable under the same exposure conditions as used for film exposure. A mathematical model was developed and tested, which allows the calculation of the image quality, proven by image quality indicators (IQI), depending on the fixed pattern noise and the efficiency of the used imaging plate (IP) scanner system. All tested IP scanner systems provide a fixed correlation between the measured signal/noise ratio (SNR) and the grey values in the digital images. The maximum achievable SNRmax depends on the fixed pattern noise of the plate (high dose limit). Depending on the exposure dose an optimum visibility of IQIs can be predicted by calculation and measured considering the attenuation coefficient for calculation of the contrast/noise ratio (CNR). The diameter of the just visible wire is proportional to 1/sqrt(CNR). The optimum tube voltage for best visibility of IQIs and maximum CNR depend on the exposure dose. The optimum tube voltage for best visibility is achieved only at exposures with high dose. A dose dependant optimal tube voltage was not observed for film radiography. T2 - 37th Annual Review of Progress in Quantitative Nondestructive Evaluation CY - San Diego, CA, USA DA - 18.07.2010 KW - Computed radiography KW - Computed radiology KW - Non destructive testing KW - Image quality KW - Signal-to-noise ratio KW - Contrast-to-noise ratio KW - Imaging plates KW - Optimum tube voltage KW - Standards PY - 2011 SN - 978-0-7354-0888-3 U6 - https://doi.org/10.1063/1.3591892 SN - 0743-0760 SN - 0094-243X N1 - Serientitel: AIP conference proceedings – Series title: AIP conference proceedings IS - 1335 SP - 493 EP - 500 PB - American Institute of Physics CY - Melville, NY AN - OPUS4-27540 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ewert, Uwe A1 - Zscherpel, Uwe A1 - Heyne, Klara A1 - Jechow, Mirko A1 - Bavendiek, K. T1 - Image quality in digital industrial radiography KW - ZfP KW - Radiographie KW - Bildqualität KW - Digitale Detektoren PY - 2012 SN - 0025-5327 VL - 70 IS - 8 SP - 955 EP - 964 PB - Society for Nondestructive Testing CY - Columbus, Ohio AN - OPUS4-26699 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Jaenisch, Gerd-Rüdiger A1 - Ewert, Uwe A1 - Jechow, Mirko T1 - Scatter radiation in radiography T2 - 10th European conference on non-destructive testing CY - Moscow, Russia DA - 2010-06-07 KW - Simulation KW - Streustrahlung KW - Abbildungskontrast KW - Bildqualität PY - 2010 IS - Paper 4.02.18 SP - 1 EP - 9 AN - OPUS4-22765 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Jechow, Mirko A1 - Zscherpel, Uwe A1 - Redmer, Bernhard A1 - Ewert, Uwe T1 - Dependence of Signal-to-Noise Ratio on Radiation Energy and Grey Value in Digital Industrial Radiology T2 - International symposium on digital industrial radiology and computed tomography, DIR 2011 der DGZfP CY - Berlin, Germany DA - 2011-06-20 PY - 2011 AN - OPUS4-26664 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Jechow, Mirko A1 - Beckmann, Jörg A1 - Ewert, Uwe A1 - Przybilla, Waldemar A1 - Redmer, Bernhard A1 - Zscherpel, Uwe T1 - Die richtige Auswahl von Vorder- und Hinterfolien für die Computer-Radiographie mit Speicherfolien T2 - DGZfP-Jahrestagung 2010 CY - Erfurt, Germany DA - 2010-05-10 PY - 2010 AN - OPUS4-22633 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Osterloh, Kurt A1 - Fratzscher, Daniel A1 - Jechow, Mirko A1 - Bücherl, T. A1 - Schillinger, B. A1 - Hasenstab, A. A1 - Zscherpel, Uwe A1 - Ewert, Uwe T1 - Limited view tomography of wood with fast and thermal neutrons N2 - Neutrons are absorbed particularly by hydrogen containing materials so they can be used as a tool for visualising heterogeneous density distributions of organic materials. Penetration limits are set by the layer thicknesses and the neutron energies applied. In case of specimens with a flat shape the situation may be encountered that the object only could be penetrated in selected directions. In addition, the overall size may exceed the beam diameter and thus the viewing window if a certain region of interest should be studied by tomography without destroying the integrity of the specimen. This study shows the capabilities and limits of thermal and fast neutrons to investigate flat wooden specimens such as boards and girders by neutron tomography under such circumstances. Taking projections was impaired either by the limits of penetrability or by the total size of the object. As a consequence, projections were included for reconstruction only from a limited angular range of 90°. It could be shown that an approach based on the slice theorem was capable to visualise structural features along the beam directions while simply omitting the perpendicular ones without causing additional artefacts. Small samples with a thickness of up to 2 cm but several times broader could be studied with the ANTARES facility of the FRM II neutron source in Garching providing thermal neutrons while larger objects required a beam of higher energy as available in the NECTAR facility of the same institution. The fast (fission) neutrons (1.5 – 2 MeV) of this site allowed investigating an area of interest inside a girder with a cross section of 23.5 x 49 cm². Internal features such as inclusions could be detected as well as a heterogeneous density distribution in glue layers. T2 - International symposium on digital industrial radiology and computed tomography CY - Berlin, Germany DA - 20.06.2011 KW - Computed tomography KW - Slice theoreme KW - Limited angle tomography PY - 2011 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:kobv:b43-250663 SN - 978-3-940283-34-4 VL - 128 SP - 1 EP - 5 PB - Deutsche Gesellschaft für Zerstörungsfreie Prüfung (DGZfP) CY - Berlin AN - OPUS4-25066 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Osterloh, Kurt A1 - Jechow, Mirko A1 - Fratzscher, Daniel A1 - Wrobel, Norma A1 - Zscherpel, Uwe A1 - Ewert, Uwe A1 - Bücherl, T. A1 - Hasenstab, A. A1 - Weiss, P. A1 - Weiss, C. A1 - Tannert, T. T1 - Strukturmerkmale im Holz - radiologisch gesehen N2 - Die radiologische Untersuchung von Holz ist hinsichtlich der Durchführung und Auswertung komplex. Im Gegensatz zu künstlich hergestellten Materialien zeichnet sich Holz besonders durch seine Heterogenität und Individualität aus, kein Stück ist wie das andere. Dies erschwert bei der radiologischen Inspektion die Fehlerdetektion, da diese teilweise schwer von natürlichen Wachstumsmerkmalen zu unterscheiden sind. Insektenfraßgänge und Hohlräume durch biologischen Abbau sind deutlich erkennbar, schwieriger wird es bei feinen Rissen und Anzeichen für Fäulnis. Metallische Einbauten wie Nägel und Bolzen sind natürlich gut zu erkennen, großflächigere Platten können jedoch den Einblick in die benachbarten Gebiete versperren. Klebstoffschichten sind im Röntgenbild nur dann sichtbar, wenn sie exakt parallel zum Strahlengang getroffen werden. Verschiedene Strukturmerkmale treten also radiologisch recht unterschiedlich in Erscheinung, deshalb liegt es nahe, zu deren Erkennung mehr als nur eine Durchstrahlungstechnologie anzuwenden. Während bei immobilen Objekten, d.h. Bauwerken und Bäumen, mobile Durchstrahlungstechniken mit Röntgen- oder Gammastrahlen (Se-75) in Frage kommen, stehen im Labor auch (schnelle) Neutronen alternativ zur Verfügung. Die verschiedenen Techniken, einschließlich der Tomographie mit Neutronen, werden an zwei größeren Prüfteilen demonstriert, einer Brettschichtholzprobe und einem Knotenpunkt aus einer Holzbrücke mit eingebauter Stahlplatte. Neutronen lassen Materialunterschiede differenzierter hervortreten und durchdringen sperrige Metallteile besser, trotz verringerter Bildqualität. An Verbesserungen wird derzeit gearbeitet. T2 - DGZfP-Jahrestagung 2010 CY - Erfurt, Deutschland DA - 2010-05-10 KW - Mobile Radiographie KW - Holzbauteilprüfung PY - 2010 SN - 978-3-940283-26-9 IS - DGZfP-BB 122 (Poster 14) SP - 1 EP - 2 CY - Berlin AN - OPUS4-22532 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -