TY - JOUR A1 - Holzweber, Markus A1 - Unger, Wolfgang A1 - Hodoroaba, Vasile-Dan T1 - Ionic liquids as a reference material candidate for the quick performance check of energy dispersive X-ray spectrometers for the low energy range below 1 keV N2 - Ionic liquids (ILs) are proposed as simple and efficient test materials to evaluate the performance of energy dispersive X-ray spectrometers (EDS) in the low energy range below 1 keV. By only one measurement, C Kα, N Kα, O Kα, and F Kα X-ray lines can be excited. Additionally, the S Kα line at 2.3 keV and, particularly, the S L series at 149 eV complete the picture with X-ray lines offered by the selected ILs. The well-known (certifiable) elemental composition of the ILs selected in the present study can be used to check the accuracy of results produced with the available EDS quantification routines in the low energy range, simultaneously, for several low atomic number elements. A comparison with other reference materials in use for testing the performance of EDS in the low energy range is included. KW - Ionic liquids KW - EDX KW - CRM KW - Reference material KW - Low energy KW - X-rays PY - 2016 UR - http://pubs.acs.org/doi/abs/10.1021/acs.analchem.6b01444 U6 - https://doi.org/10.1021/acs.analchem.6b01444 SN - 0003-2700 SN - 1520-6882 VL - 88 IS - 14 SP - 6967 EP - 6970 PB - ACS Publications CY - 1155 Sixteenth Street N.W., Washington, DC 20036 AN - OPUS4-36912 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Holzweber, Markus A1 - Hodoroaba, Vasile-Dan A1 - Unger, Wolfgang T1 - What about ionic liquids as a 'hot' certified reference material candidate to check your EDS below 1 keV? N2 - Energy-dispersive X-ray spectrometry (EDX) is one of the most applied methods used for the analysis of the chemical composition of solids and thin films. Recent progress in EDS (Energy Dispersive X-ray Spectrometer) technology has increased the general performance also in the energy range < 1 keV addressing low Z elements. Suitable test materials to be employed especially to check the low-energy EDS performance – also in line with ISO 15632 - are rather limited and mainly based on C K and F K lines. In order to obtain valid results in laboratories accredited in compliance with ISO/IEC 17025 it is necessary to periodically check the instrument performance. T2 - M&M 2015 - Microscopy & Microanalysis CY - Portland, Oregon, USA DA - 02.08.2015 KW - Reference material KW - EDX KW - Ionic liquids PY - 2015 U6 - https://doi.org/10.1017/S1431927615009356 VL - 21 SP - Suppl. 3, Paper 0856 AN - OPUS4-34211 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -