TY - JOUR A1 - Ecke, Alexander A1 - Bell, Jérémy A1 - Schneider, Rudolf T1 - A three-dimensional microfluidic flow cell and system integration for improved electrochemical substrate detection in HRP/TMB-based immunoassays N2 - Immunoassays, based on the recognition and capture of analytes by highly selective antibodies, are now used extensively in all areas of diagnostics, but the challenge is to further integrate them into online sensors. To improve the transition from laboratory immunoassays to immunosensors, we have developed a complete flow system, based on a microfluidic core flow cell to enable automated detection of one of the most commonly used immunoassay substrates, TMB, by chronoamperometry. The architecture and fluidic optimisation of the system showed that a specially designed 3D flow cell allows higher flow rates (500 μL min−1) than a standard enlarged microfluidic channel (50 μL min−1) resulting in a significantly shorter detection time of 30 seconds per sample and making the system more robust against interferences due to bubble formation in the chip. The electrochemical measurements showed an improved signal-to-noise ratio (SNR) and thus higher sensitivity for a model immunoassay for diclofenac (SNR = 59), compared to the analytical performance of a conventional laboratory microplate-based assay with optical detection (SNR = 19). In general, this system facilitates the conversion of any conventional immunoassay into an immunosensor with automatic and continuous detection. KW - Microfluidic KW - Immunoassay KW - Electrochemical KW - Mikrofluidik KW - Immunassay KW - Elektrochemie PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-580159 DO - https://doi.org/10.1039/d3sd00095h VL - 2 SP - 887 EP - 892 PB - Royal Society of Chemistry CY - London, United Kingdom AN - OPUS4-58015 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Johann, Sergej A1 - Mansurova, Maria A1 - Kohlhoff, Harald A1 - Gkertsos, Aris A1 - Neumann, Patrick P. A1 - Bell, Jérémy A1 - Bartholmai, Matthias T1 - Wireless Mobile Sensor Device for in-situ Measurements with Multiple Fluorescent Sensors N2 - This paper describes a wireless mobile prototype able to perform optical measurements by means of a miniatur-ized spectrometer for low light analysis, e.g. fluorescent sensors. Evaluations, calculations, calibration management and result display are performed by a computer or a standard tablet. The device was designed primarily to detect traces of oil in drinking or ground water and for the analyses of crude oils. However, it can also address a wide range of fluorescent sensors. The fast and user-friendly inspection of water quality or oil properties, as well as the adaptability and mobility, make the device attractive for a variety of users. Further application areas could be easily imple-mented by adapting the optics and the software (database, data processing and calibration plots, etc.) T2 - IEEE Sensors 2018 CY - New Delhi, India DA - 28.10.2018 KW - Wireless mobile sensor device KW - Fluorescent sensor KW - Embedded system KW - Water quality KW - Oil PY - 2018 SN - 978-1-5386-4707-3 SP - 1067 EP - 1070 PB - IEEE CY - New Delhi, India AN - OPUS4-46556 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - GEN A1 - Ecke, Alexander A1 - Bell, Jérémy A1 - Schneider, Rudolf T1 - From Laboratory Immunoassays to Immunosensors with an Integrated Microfluidic Flow System N2 - This research summary describe efforts to the development of a three-dimensional microfluidic flow cell for improved electrochemical substrate detection in HRP/TMB-based immunoassays. This work provides a valuable advancement in the sensitivity and integration of immunoassay technologies, aiding in the refinement of diagnostic tools. KW - Antikörper KW - Immunoassay KW - Microfluidics KW - Electrochemical detection KW - Biosensor PY - 2025 UR - https://www.elveflow.com/microfluidics-research-summaries/from-laboratory-immunoassays-to-immunosensors-with-an-integrated-microfluidic-flow-system/ SP - 1 EP - 4 CY - Paris AN - OPUS4-62674 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Burnage, Samual A1 - Bell, Jérémy A1 - Wan, Wei A1 - Kislenko, Evgeniia A1 - Rurack, Knut T1 - Combining a hybrid chip and tube microfluidic system with fluorescent molecularly imprinted polymer (MIP) core–shell particles for the derivatisation, extraction, and detection of peptides with N-terminating phosphorylated tyrosine N2 - The reliable identification and quantitation of phosphorylated amino acids, peptides and proteins is one of the key challenges in contemporary bioanalytical research, an area of particular interest when attempting to diagnose and treat diseases at an early stage. We have developed a synthetic probe for targeting phosphorylated amino acids, based on core–shell submicron-sized particles consisting of a silica core, coated with a molecularly imprinted polymer (MIP) shell. The MIP layer contains a fluorescent probe crosslinker which binds selectively to phosphorylated tyrosine (pY) moieties with a significant imprinting factor (IF) and responds with a "light-up” fluorescence signal. The bead-based ratiometric detection scheme has been successfully transferred to a microfluidic chip format and its applicability to rapid assays has been exemplarily shown by discriminating a pY-terminating oligopeptide against its nonphosphorylated counterpart. Such miniaturised devices could lead to an automated pY or pY N-terminated peptide measurement system in the future. The setup combines a modular microfluidic system for amino acid derivatisation, extraction (by micropillar co-flow) and selective adsorption and detection with the fluorescent MIP core–shell particle probes. A miniaturised optical assembly for low-light fluorescence measurements was also developed, based on miniaturised opto-electronic parts and optical fibres. The emission from the MIP particles upon binding of pY or pY N-terminated peptides could be monitored in real-time. KW - Microfluidics KW - Molecularly imprinted polymers KW - Phosphorylated peptides KW - Fluorescence KW - Core-shell particles PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-569204 DO - https://doi.org/10.1039/d2lc00955b SN - 1473-0197 VL - 23 IS - 3 SP - 466 EP - 474 PB - Royal Society of Chemistry CY - Cambridge AN - OPUS4-56920 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Lempke, L. A1 - Fischer, Tobias A1 - Bell, Jérémy A1 - Kraus, Werner A1 - Rurack, Knut A1 - Krause, N. T1 - Gold-catalyzed allene cycloisomerization for pyrrole synthesis: towards highly fluorinated BODIPY dyes N2 - A novel synthetic strategy toward highly fluorinated BODIPY dyes with exceptional photostabilities relying on sustainable gold catalysis has been developed. A key to the tailored pyrrole precursors is the gold catalysis performed in ionic liquids as the reaction medium, allowing a facile recycling of the catalysts. The dyes prepared are well-matching with the spectral windows of popular rhodamine dyes and possess high brightness while showing a distinctly higher photostability than the rhodamines especially in aprotic solvents. KW - Gold-Katalyse KW - BODIPY-Farbstoffe KW - Photostabilität KW - Fluor-Substitution KW - Fluoreszenz PY - 2015 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-344692 DO - https://doi.org/10.1039/c4ob02671c SN - 1477-0520 SN - 1477-0539 VL - 13 IS - 12 SP - 3787 EP - 3791 PB - RSC CY - Cambridge AN - OPUS4-34469 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Mansurova, Maria A1 - Johann, Sergej A1 - Kohlhoff, Harald A1 - Rurack, Knut A1 - Bartholmai, Matthias A1 - Bell, Jérémy T1 - On-Site Analytical Tool Based on Crude Oil Fluorescence and Chemometrics for the Rapid Determination of the Nature and Essential Properties of Oil Spills N2 - With the reduction of large oil spills because of stricter regulations and safety measures, the question of how to manage smaller oil spills arises. Few on-site analytical tools are available for first responders or other law enforcement personnel to rapidly test for crude oil in the early management of localized polluted areas. The approach reported here relies on well-described computer-assisted multivariate data analysis of the intrinsic fluorescence fingerprints of crude oils to build a multivariate model for the rapid classification of crude oils and the prediction of their properties. Thanks to a dedicated robust portable reader, the method allowed classification and accurate prediction of various properties of crude oil samples like density (according to API, the American Petroleum Institute and viscosity as well as composition parameters such as volume fractions of paraffins or aromatics. In this way, autonomous operation in on-site or in-the-field applications becomes possible based on the direct (undiluted and untreated) measurement of samples and a rapid, tablet-operated readout system to yield a robust and simple analytical test with superior performance. Testing in real-life scenarios allowed the successful classification and prediction of a number of oil spill samples as well as weathered samples that closely resemble samples collected by first responders. KW - Oil spills KW - Fluorescence KW - PCA KW - Petroleum KW - Rapid test KW - Portable PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-595442 DO - https://doi.org/10.1021/acsestwater.3c00648 VL - 4 IS - 2 SP - 621 EP - 627 PB - American Chemical Society (ACS) AN - OPUS4-59544 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Bell, Jérémy T1 - Microfluidic tubing system for mercury detection utilizing fluorescent sensor nanoparticles N2 - Currently, the development of portable devices to detect traces of chemical species by real time monitoring is a highly topical research area in the chemical sciences. Indeed, the lowering of standards for pollutants in drinking water requires the development of new technologies and devices. In this context, gated sensory nanoparticle systems for selective and sensitive recognition of targets such as heavy metal ions, small molecules or oligonucleotides are excellent candidates[1,2]. Provided that such sensory delivery systems are highly porous nanoparticles loaded with indicator dyes and capped with bulky stoppers that can be uncapped only by the designated analyte, an integration with a microfluidic tubing system designed for droplet-assisted liquid-liquid extraction enable separation of released from still particle-confined dye by phase-transfer extraction and subsequent selective detection by fluorescence. The incorporation in a simple microfluidic tubing system is an innovative way towards the realization of a sensor for trace analysis of target molecules in real time. T2 - 9th International Conference on "Instrumental Methods of Analysis-Modern Trends and Applications" (IMA 2015) CY - Kalamata, Grece DA - 2015-09-20 PY - 2015 AN - OPUS4-34774 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Gotor, Raúl A1 - Rurack, Knut A1 - Bell, Jérémy T1 - Fluorescent Paper Strips for the Detection of Diesel Adulteration with Smartphone Read-out N2 - Three fluorescent molecular rotors of 4-dimethylamino-4-nitrostilbene (4-DNS) were investigated for their potential use as viscosity probes to indicate the content of kerosene in diesel/kerosene blends, a wide-spread activity to adulterate fuel. In solvents with low viscosity, the dyes rapidly deactivate via a so-called twisted intramolecular charge transfer state, efficiently quenching the fluorescence. Measurements of diesel/kerosene blends revealed a good linear correlation between the decrease in fluorescence and the increase of the fraction of the less viscous kerosene in diesel/kerosene blends. Immobilization of the hydroxy derivative 4-DNS-OH in cellulose paper yielded test strips that preserve the fluorescent indicator's behavior. Combination of the strips with a reader based on a smartphone and a controlling app allowed to create a simple field test. The method can reliably detect the presence of kerosene in diesel from 7 to 100%, outperforming present standard methods for diesel adulteration. KW - Diesel KW - Kerosene KW - Adulteration KW - Fluorescent sensor KW - Smartphone KW - Test strips PY - 2018 UR - https://www.jove.com/video/58019 DO - https://doi.org/10.3791/58019 VL - 141 SP - e58019-1 EP - e58019-6 PB - MyJoVE Corporation CY - Cambridge, MA, USA AN - OPUS4-46567 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Bell, Jérémy T1 - Microfluidic platform and fluorescent sensor particles for the detection of the herbicide 2,4-D in worldwide samples N2 - Fluorescent sensory MIP (molecularly imprinted polymer) particles were combined with a droplet-based 3D microfluidic system for the sensitive and selective determination of 2,4-dichlorophenoxyacetic acid (2,4-D) in water samples. 2,4-D being an important and widely used herbicide to regulate plant growth, its extensive use leads to food or ground water contamination. Analytical assays based on MIP have emerged as a valuable tool in the field of environmental analysis thanks to low production costs, stability, format adaptability and their ability to recognize a wide variety of targets, such as the critical neutral organic molecules present in water: herbicides, pesticides, antibiotics, etc. To accomplish this, a tailor-made fluorescent indicator cross-linker was designed that translates directly the binding event into an enhanced fluorescence signal. This phenoxazinone-type cross-linker was co-polymerized into a thin MIP layer grafted from the surface of silica microparticles. The latter was achieved by integration of the fluorescent core-shell MIP sensor particles into a modular microfluidic platform that allows for an in-line phase-transfer assay, extracting the analyte from aqueous sample droplets into organic phase droplets that contain the sensor particles. This tool offering a novel simple and rapid way for the detection of herbicides, real-time fluorescence determination of 2,4-D down to 20 nM was realized with the system and applied for the analysis of worlwide surface water samples. T2 - INTERNATIONAL CONFERENCE ON CHEMISTRY AND THE ENVIRONMENT - ICCE 2017 CY - Oslo, Norway DA - 18.06.2017 KW - Microfluidics KW - MIP KW - Fluorescence KW - Pesticide PY - 2017 AN - OPUS4-42454 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Abdelshafi, Nahla A. A1 - Bell, Jérémy A1 - Rurack, Knut A1 - Schneider, Rudolf T1 - Microfluidic electrochemical immunosensor for the trace analysis of cocaine in water and body fluids N2 - Quick but accurate testing and on‐the‐spot monitoring of cocaine in oral fluids and urine continues to be an important toxicological issue. In terms of drug testing, a number of devices have been introduced into the market in recent decades, notably for workplace inspection or roadside testing. However, these systems do not always fulfill the requirements in terms of reliability, especially when low cut‐off levels are required. With respect to surface water, the presence of anthropogenic small organic molecules such as prescription and over‐the‐counter pharmaceuticals as well as illicit drugs like cannabinoids, heroin, or cocaine, has become a challenge for scientists to develop new analytical tools for screening and on‐site analysis because many of them serve as markers for anthropogenic input and consumer behavior. Here, a modular approach for the detection of cocaine is presented, integrating an electrochemical enzyme‐linked immunosorbent assay (ELISA) performed on antibody‐grafted magnetic beads in a hybrid microfluidic sensor utilizing flexible tubing, static chip and screen‐printed electrode (SPE) elements for incubation, recognition, and cyclic voltammetry measurements. A linear response of the sensor vs. the logarithm of cocaine concentration was obtained with a limit of detection of 0.15 ng/L. Within an overall assay time of 25 minutes, concentrations down to 1 ng/L could be reliably determined in water, oral fluids, and urine, the system possessing a dynamic working range up to 1 mg/L. KW - Cocaine PY - 2018 DO - https://doi.org/10.1002/dta.2515 VL - 11 SP - 492 EP - 500 PB - Wiley CY - Weinheim AN - OPUS4-48319 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -