TY - JOUR A1 - Dorgerloh, Ute A1 - Becker, Roland A1 - Freeling, Finnian A1 - Scheurer, Marco A1 - Gökçe, Emine A1 - Strack, Katrin A1 - Sommerfeld, Thomas A1 - Sauer, Andreas A1 - Schneider, Jonas A1 - Härtel, Christoph A1 - Valkov, Vassil A1 - Moser, Lukas A1 - Geiß, Sabine A1 - Junginger, Sabine A1 - Schlittenbauer, Linda A1 - Suess, Elke A1 - Ruth, Katinka T1 - Standardising the quantification of trifluoroacetic acid in water: Interlaboratory validation trial using liquid chromatography-mass spectrometric detection (LC–MS/MS) N2 - The concentration of trifluoroacetic acid in surface and groundwaters and drinking water is subject of increasing concern. Since there is no standardised procedure available, German standardisation body DIN has initiated the first draft standard using liquid chromatography with tandem mass spectrometry (LC–MS/MS) after direct injection for the concentration range between ≥ 0.1 and 3 µg/L. Herein, the interlaboratory comparison as final validation step involving 12 expert laboratories from Germany and Switzerland is described. Two groundwaters, two surface waters, a drinking water, a rainwater, and two fortified groundwater samples displayed relative repeatability standard deviations between 3 and 7% and relative reproducibility standard deviation between 6 and 20% and were included in the draft standard DIN 38407-53. KW - TFA Trifluoroacetic acid KW - LC/MS-MS KW - Validation trial PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-632376 DO - https://doi.org/10.1007/s00769-025-01640-2 SN - 0949-1775 SP - 1 EP - 4 PB - Springer CY - Berlin AN - OPUS4-63237 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Dorgerloh, Ute A1 - Becker, Roland A1 - Riedel, Juliane A1 - Hofmann, Andrea T1 - Comparison of gas- and liquid chromatography-mass spectrometry for trace analysis of anilines in groundwater N2 - Three chromatographic procedures were investigated regarding their potential for the quantification of aniline and 19 of its methylated and chlorinated derivatives in groundwater. These methods were based on liquid-liquid-extraction in combination with gas chromatography and single quadrupole mass spectrometry (GC/MS) according to German standard DIN 38407-16:1999 and its extension using tandem mass spectrometry (GC/MS-MS), both following liquid-liquid extraction, and as third alternative the direct injection of the water sample into a liquid chromatograph coupled to tandem mass spectrometry (LC/MS-MS). Results were compared using fortified water and real-world contaminated groundwater used in an interlaboratory comparison. It could be shown that GC/MS and GC/MS-MS yielded results deviating less than 10% from each other while all three procedure displayed quantification results deviating less than 15% from the intercomparison reference values in case of each analyte in the concentration range between 1 and 45 µg L-1. Though GC/MS-MS displays a ten-fold higher sensitivity than single quadrupole GC/MS, the precision of both methods in the concentration range was similar. LC/MS-MS has the advantage of no further sample preparation due to the direct injection and leads for methylanilines and meta-, para- substituted chloroanilines to results sufficiently equivalent to the standardised GC/MS method. However, LC/MS-MS is not suitable for ortho-chloroaniline derivates due to significantly lower ion yields than meta- and para-substituted chloroanilines. KW - Interlaboratory comparison KW - Aniline KW - Chloroanilines KW - Methylanilines KW - Groundwater KW - GC/MS KW - GC/MS-MS KW - LC/MS-MS PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-535793 DO - https://doi.org/10.1080/03067319.2021.1987423 VL - 103 IS - 19 SP - 8465 EP - 8477 PB - Taylor & Francis CY - London AN - OPUS4-53579 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -