TY - CONF A1 - Abbas, Ioana M. A1 - Hoffmann, Holger A1 - Weller, Michael G. T1 - Troubleshooting for the clinical measurement of iron biomarkers using LC-MS/MS: Suppression of sample losses in autosampler vials N2 - We developed a rapid and robust HPLC-MS/MS (QqQ) method for the quantification of hepcidin-25, a promising new biomarker in iron metabolism, in human samples. The novelty of the method is the use of special HPLC vials to avoid adsorptive losses due to the basic character of the peptide that causes interaction with the silanol groups of the vial’s glass surface. Up to 90% decrease in the MS/MS signal was observed, when commercial HPLC vials were used, while vials treated with 3-(2-aminoethylamino)propylmethyldimethoxysilane or 1H,1H,2H,2H-perfluorooctyltriethoxysilane, leading to no losses in the range of physiological hepcidin-25 mean serum levels (10-20 µg/L). T2 - MSACL 2017 EU CY - Salzburg, Austria DA - 10.09.2017 KW - Peptide losses KW - Adsorption KW - Silanization PY - 2017 AN - OPUS4-44613 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -