TY - JOUR A1 - Hübert, Thomas A1 - Tiebe, Carlo A1 - Detjens, Marc A1 - Majewski, J. T1 - On-site calibration system for trace humidity sensors JF - Measurement N2 - A portable device for calibration of trace humidity sensors and an adopted calibration procedure have been developed. The calibration device is based on humidity generation by permeating water through polymeric membrane tubes. Water vapour transmission rates for various polymers were experimentally determined in order to select the most suitable polymeric material. The developed trace humidity generator consists of a gas-flow polymeric hose immersed in a water reservoir thermostated by a sensorcontrolled heater. Mole fractions of water vapour between 1 µmol/ molˉ¹ and 350 µmol/molˉ¹ (equivalent to frost-point temperatures from -76 °C to -31 °C) were generated by varying either the operating temperature or gas flow. The operating temperature can be varied from 20 °C to 60 °C and kept stable within 0.1 K. Uncertainty analysis indicated that the trace humidity generator produces gas flows of constant humidity amounts with a relative expanded uncertainty less than 3.4% (k = 2) of the generated value. KW - Polymeric tubes KW - Measurements KW - Trace Humidity KW - Sensor calibration KW - Water vapour permeation PY - 2016 DO - https://doi.org/10.1016/j.measurement.2016.05.013 SN - 0263-2241 VL - 91 IS - 9 SP - 251 EP - 257 PB - Elsevier Ltd. AN - OPUS4-36762 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -