TY - CONF A1 - Detjens, Marc A1 - Tiebe, Carlo A1 - Banach, Ulrich A1 - Ritter, U. T1 - Ionic Liquid Based Coulometric Trace Humidity Sensors T2 - ALLSENSORS 2018 N2 - This work presents a first attempt to use ionic liquids as a new coating for planar coulometric sensors. These sensors are used for the measurement of trace humidity in various gases. Usually, the coating of the sensors is tetraphosphorus decaoxide and its hydrolysis products. Instead, a hygroscopic ionic liquid was used as sensor coating in this work. Generated frost point temperatures tf in the gas ranged from -80 °C to -30 °C, which is equivalent to vapour mole fractions xv from 0.5 μmol·mol-1 to 376 μmol·mol-1. In addition to the coulometric sensors, the generated humidity is determined by a precision dew point hygrometer as reference. First results show that it is possible to measure humidity with ionic liquid based coulometric sensors. T2 - ALLSENSOR 2018 CY - Rom, Italy DA - 25.03.2018 KW - Ionic liquids. KW - Coulometric sensors KW - Sensor coating KW - Trace humidity measurement PY - 2018 SN - 978-1-61208-621-7 SP - 13 EP - 14 PB - IARIA AN - OPUS4-44609 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -