TY - JOUR A1 - Vogl, Jochen T1 - Advances in isotope ratio mass spectrometry and required isotope reference materials JF - Mass spectrometry N2 - The article gives a condensed version of the keynote lecture held at the International Mass Spectrometry Conference 2012 in Kyoto. Starting with some examples for isotope research the key requirements for metrologically valid procedures enabling traceable and comparable isotope data are discussed. Of course multi-collector mass spectrometers are required which offer sufficiently high isotope ratio precision for the intended research work. Following this, corrections for mass fractionation/discrimination, validation of the analytical procedure including chemical sample preparation and complete uncertainty budgets are the most important issues for obtaining a metrologically valid procedure for isotope ratio determination. Only the application of such metrologically valid procedures enables the generation of traceable and comparable isotope data. To realize this suitable isotope and/or δ-reference materials are required, which currently are not sufficiently available for most isotope systems. Boron is given as an example, for which the situation regarding isotope and δ-reference materials is excellent. Boron may therefore serve as prototype for other isotope systems. KW - Isotope reference materials KW - Delta reference materials KW - Traceability KW - Comparability KW - Measurement uncertainty KW - Synthetic isotope mixtures KW - Mass fractionation KW - Mass discrimination PY - 2013 DO - https://doi.org/10.5702/massspectrometry.S0020 SN - 2187-137X SN - 2186-5116 VL - 2 IS - 50020 SP - 1 EP - 7 AN - OPUS4-28562 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -