TY - JOUR A1 - Thomsen-Schmidt, P. A1 - Hasche, K. A1 - Ulm, G. A1 - Herrmann, K. A1 - Krumrey, M. A1 - Ade, G. A1 - Stümpel, J. A1 - Busch, I. A1 - Schädlich, S. A1 - Schindler, A. A1 - Frank, W. A1 - Hirsch, D. A1 - Procop, Mathias A1 - Beck, Uwe T1 - Realisation and metrological characterisation of thickness standards below 100 nm T2 - Applied physics A N2 - High-accuracy film thickness measurements in the range below 100 nm can be made by various complex methods like spectral ellipsometry (SE), scanning force microscopy (SFM), grazing incidence X-ray reflectometry (GIXR), or X-ray fluorescence analysis (XRF). The measurement results achieved with these methods are based on different interactions between the film and the probe. A key question in nanotechnology is how to achieve consistent results on a level of uncertainty below one nanometre with different techniques. Two different types of thickness standards are realised. Metal film standards for X-ray techniques in the thickness range 10 to 50 nm are calibrated by GIXR with monochromatised synchrotron radiation of 8048 eV. The results obtained at four different facilities show excellent agreement. SiO2 on Si standards for SE and SFM in the thickness range 6 to 1000 nm are calibrated by GIXR with monochromatised synchrotron radiation of 1841 eV and with a metrological SFM. Consistent results within the combined uncertainties are obtained with the two methods. Surfaces and interfaces of both types of standards are additionally investigated by transmission electron microscopy (TEM). PB - Springer CY - Berlin PY - 2004 UR - https://opus4.kobv.de/opus4-bam/frontdoor/index/index/docId/17195 AN - OPUS4-17195 SN - 0947-8396 VL - 78 IS - 5 SP - 645 EP - 649 AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany