TY - JOUR A1 - Höhse, Marek A1 - Gornushkin, Igor B. A1 - Merk, Sven A1 - Panne, Ulrich T1 - Assessment of suitability of diode pumped solid state lasers for laser induced breakdown and Raman spectroscopy JF - Journal of analytical atomic spectrometry N2 - The potential of a diode pumped solid state (DPSS) laser for laser induced breakdown spectroscopy (LIBS) and Raman spectroscopy was investigated. The DPSS laser operating at repetition rates of 1 Hz–200 kHz, pulse energy of ~1 mJ, and pulse duration of ~20 ns was added to a conventional LIBS system equipped with the flashlamp-pumped Nd:YAG laser (10 Hz, 400 mJ, 6 ns) to directly compare the two laser systems. Despite the lower mass ablated per laser pulse (~several nanograms) and a weaker plasma, the total mass ablated per measurement interval (~1 s) and total emission signal were significantly higher than those obtained with the flashlamp pumped Nd:YAG laser. This resulted in the improved signal-to-noise ratio and better limits of detection (LOD). The best LODs were obtained for Cu, Cr and Ni in ultrapure standard iron samples, 0.7 ppm, 1 ppm and 5 ppm, correspondingly. The plasma from the DPSS laser was characterized in terms of its radiative properties, temperature, and shockwave propagation speed. The suitability of this laser for LIBS-Raman was demonstrated by mapping the elemental and molecular composition of a mineral sample. The results strongly imply that DPSS lasers are the promising source for LIBS, Raman, or combined LIBS-Raman spectroscopy. PY - 2011 DO - https://doi.org/10.1039/c0ja00038h SN - 0267-9477 SN - 1364-5544 VL - 26 IS - 2 SP - 414 EP - 424 PB - Royal Society of Chemistry CY - London AN - OPUS4-22943 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -