TY - JOUR A1 - Giesen, Charlotte A1 - Mairinger, T. A1 - Khoury, L. A1 - Müller, Larissa A1 - Jakubowski, Norbert A1 - Panne, Ulrich T1 - Multiplexed immunohistochemical detection of tumor markers in breast cancer tissue using laser ablation inductively coupled plasma mass spectrometry T2 - Analytical chemistry N2 - We optimized multiplexed immunohistochemistry (IHC) on breast cancer tissue. Up to 20 tumor markers are routinely evaluated for one patient, and thus, a common analysis results in a series of time consuming staining procedures. As an alternative, we used lanthanides for labeling of primary antibodies, which are applied in IHC. Laser ablation (LA) ICPMS was elaborated as a detection tool for multiplexed IHC of tissue sections. In this study, we optimized sample preparation steps and LA ICPMS parameters to achieve a sufficient signal-to-background ratio. The results prove the high selectivity of applied antibodies, which was sustained after labeling. Up to three tumor markers (Her 2, CK 7, and MUC 1) were detected simultaneously in a single multiplex analysis of a 5 µm thin breast cancer tissue at a laser spot size of 200 µm. Furthermore, the LA ICPMS results indicate a significantly higher expression level of MUC 1 compared to Her 2 and CK 7, which was not obvious from the conventionally stained tissue sections. PB - American Chemical Society CY - Washington, DC PY - 2011 UR - https://opus4.kobv.de/opus4-bam/frontdoor/index/index/docId/24966 AN - OPUS4-24966 SN - 0003-2700 SN - 1520-6882 N1 - Geburtsname von Müller, Larissa: Wäntig, L. - Birth name of Müller, Larissa: Wäntig, L. VL - 83 IS - 21 SP - 8177 EP - 8183 AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany