Self-referencing SPR-biosensors based on penetration difference of evanescent waves

  • SPR based biosensors register binding of analytes to the surface with immobilized receptors by measuring changes of the refractive index near this surface. An important task in the improvement of this measurement technology is a separation of signals, corresponding to the changes in the chemosensitive layer, from undesired contributions of bulk phase, for example, due to fluctuations of temperature, concentrations of solutes, pressure. The wavelength of the incident light influences strongly the penetration depth of the corresponding evanescent wave. This dependence was exploited here for compensation of the contribution of the bulk refractive index. It was performed using differential SPR measurements at two wavelengths with differing penetration depths. Theoretical analysis and numerical optimization of the suggested approach, named a Penetration Difference Self-Referencing SPR (PDSR-SPR), were performed. Experimental test was performed using 658 and 980 nm laser diodes. Over 20 times suppression of variations of bulk refractiveSPR based biosensors register binding of analytes to the surface with immobilized receptors by measuring changes of the refractive index near this surface. An important task in the improvement of this measurement technology is a separation of signals, corresponding to the changes in the chemosensitive layer, from undesired contributions of bulk phase, for example, due to fluctuations of temperature, concentrations of solutes, pressure. The wavelength of the incident light influences strongly the penetration depth of the corresponding evanescent wave. This dependence was exploited here for compensation of the contribution of the bulk refractive index. It was performed using differential SPR measurements at two wavelengths with differing penetration depths. Theoretical analysis and numerical optimization of the suggested approach, named a Penetration Difference Self-Referencing SPR (PDSR-SPR), were performed. Experimental test was performed using 658 and 980 nm laser diodes. Over 20 times suppression of variations of bulk refractive index with magnitude up to 1000 μRIU was observed. Finally, PDSR-SPR approach was applied for monitoring of antibodies binding to the immobilized antigens.show moreshow less

Export metadata

Additional Services

Search Google Scholar
Metadaten
Author: Shavkat NizamovORCiD, Vladimir M. MirskyORCiDGND
DOI:https://doi.org/10.1016/j.bios.2011.07.030
ISSN:0956-5663
Title of the source (English):Biosensors and Bioelectronics
Document Type:Scientific journal article peer-reviewed
Language:English
Year of publication:2011
Tag:Affinity senso; Evanescent wave; Immunosensor; SPR-biosenso; Surface plasmon resonance
Volume/Year:28
Issue number:1
First Page:263
Last Page:269
Faculty/Chair:Fakultät 2 Umwelt und Naturwissenschaften / FG Nanobiotechnologie